Use of a Victoreen 500 electrometer to determine ionization chamber collection efficiencies.

P R Almond.   

Abstract

An investigation has been made of the Victoreen 500 electrometer's capability to predict saturation corrections for ionization chamber measurements. This electrometer has been provided with a switch that reduces the polarizing voltage on the chamber to 40% of its normal value. It can be demonstrated theoretically that ion chamber measurements at two separate voltages should be sufficient to determine the saturation correction, for both pulsed beams and continuous radiation. Tests were made with the Victoreen 500 which indicated that this is the case. Saturation corrections of up to 10% have been demonstrated for both pulsed and continuous beams and these can be predicted using the Victoreen 500 electrometer to within 1%.

Entities:  

Mesh:

Substances:

Year:  1981        PMID: 6798399     DOI: 10.1118/1.595045

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Adaptation and dosimetric commissioning of a synchrotron-based proton beamline for FLASH experiments.

Authors:  Ming Yang; Xiaochun Wang; Fada Guan; Uwe Titt; Kiminori Iga; Dadi Jiang; Takeshi Takaoka; Satoshi Tootake; Tadashi Katayose; Masumi Umezawa; Emil Schüler; Steven Frank; Steven H Lin; Narayan Sahoo; Albert C Koong; Radhe Mohan; X Ronald Zhu
Journal:  Phys Med Biol       Date:  2022-08-05       Impact factor: 4.174

2.  Dosimetric Consistency of Co-60 Teletherapy Unit- a ten years Study.

Authors:  Misba H Baba; M Mohib-Ul-Haq; Aijaz A Khan
Journal:  Int J Health Sci (Qassim)       Date:  2013-01

3.  Ion recombination correction factors (P(ion)) for Varian TrueBeam high-dose-rate therapy beams.

Authors:  Stephen F Kry; Richard Popple; Andrea Molineu; David S Followill
Journal:  J Appl Clin Med Phys       Date:  2012-11-08       Impact factor: 2.102

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.