Travis J McCaw1,2, Min-Sig Hwang2, Si Young Jang2, M Saiful Huq2.
Abstract
This work quantified differences between recommendations of the TG-51 and TG-51 addendum reference dosimetry protocols. Reference dosimetry was performed for flattened photon beams with nominal energies of 6, 10, 15, and 23 MV, as well as flattening-filter free (FFF) beam energies of 6 and 10 MV, following the recommendations of both the TG-51 and TG-51 addendum protocols using both a Farmer® ionization chamber and a scanning ionization chamber with calibration coefficients traceable to absorbed dose-to-water (Dw ) standards. Differences in Dw determined by the two protocols were 0.1%-0.3% for beam energies with a flattening filter, and up to 0.2% and 0.8% for FFF beams measured with the scanning and Farmer® ionization chambers, respectively, due to kQ determination, volume-averaging correction, and collimator jaw setting. Combined uncertainty was between 0.91% and 1.2% (k = 1), varying by protocol and detector.Entities:
Keywords: TG-51 addendum; photon beam reference dosimetry
Mesh:
Year: 2017 PMID: 28574211 PMCID: PMC5874962 DOI: 10.1002/acm2.12110
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
PTW N30013 Farmer® ionization chamber results
| Energy | 6X | 6X‐SRS | 10X | 15X | 23X | 6XFFF | 10XFFF |
|---|---|---|---|---|---|---|---|
|
| 1.001 | 0.999 | 1.002 | 0.999 | 1.001 | 1.005 | 1.008 |
|
| 1.000 | 1.001 | 1.001 | 1.002 | 1.002 | 0.999 | 1.000 |
|
| 1.000 | 0.998 | 1.001 | 0.996 | 0.999 | 1.004 | 1.006 |
|
| 1.001 | 1.000 | 1.001 | 1.001 | 1.000 | 1.002 | 1.002 |
| Digital readout for 10 × 10 cm2 radiation field | 9.9 × 10.2 | 9.9 × 10.0 | 10.0 × 10.2 | 9.9 × 10.2 | 9.9 × 10.0 | 10.0 × 10.3 | 10.1 × 10.4 |
Exradin A1SL scanning ionization chamber results
| Energy | 6X | 6X‐SRS | 10X | 15X | 23X | 6XFFF | 10XFFF |
|---|---|---|---|---|---|---|---|
|
| 0.999 | 0.997 | 0.998 | 0.998 | 0.999 | 1.001 | 0.998 |
|
| 0.998 | 0.998 | 0.996 | 0.998 | 0.999 | 0.998 | 0.996 |
|
| 1.000 | 1.000 | 1.000 | 0.999 | 1.000 | 1.001 | 1.000 |
|
| 1.002 | 0.999 | 1.001 | 1.001 | 1.001 | 1.002 | 1.003 |
Percentage difference in D w/MU measured with each ionization chamber
| Energy | 6X | 6X‐SRS | 10X | 15X | 23X | 6XFFF | 10XFFF |
|---|---|---|---|---|---|---|---|
| TG‐51 | 0.50 | 0.03 | 0.47 | 0.11 | 0.13 | 0.57 | 0.80 |
| TG‐51 addendum | 0.40 | 0.11 | 0.22 | 0.06 | 0.19 | 0.38 | 0.28 |
Uncertainty budget with all values given as percent standard uncertainties
| Parameter | TG‐51 | TG‐51 addendum | ||
|---|---|---|---|---|
| N30013 | A1SL | N30013 | A1SL | |
| SSD setting | 0.10 | 0.10 | 0.10 | 0.10 |
| Depth setting | 0.25 | 0.25 | 0.25 | 0.25 |
| Field‐size setting | 0.20 | 0.20 | 0.05 | 0.05 |
| Charge measurement | 0.10 | 0.30 | 0.10 | 0.30 |
|
| 0.18 | 0.18 | 0.18 | 0.18 |
| Humidity | 0.15 | 0.15 | 0.15 | 0.15 |
|
60Co | 0.65 | 0.65 | 0.65 | 0.65 |
|
| 0.50 | 0.50 | 0.40 | 0.40 |
| Assignment of | 0.50 | 0.50 | 0.20 | 0.20 |
| Stability of reference chamber | 0.10 | 0.10 | 0.10 | 0.10 |
|
| 0.10 | 0.30 | 0.10 | 0.30 |
|
| 0.10 | 0.10 | 0.10 | 0.10 |
| Pre‐irradiation history | 0.10 | 0.10 | 0.10 | 0.10 |
| Leakage current | 0.10 | 0.10 | 0.10 | 0.10 |
| Linac stability | 0.05 | 0.05 | 0.05 | 0.05 |
|
| 0.10 | 0.10 | 0.10 | 0.10 |
|
| 0.60 | 0.10 | 0.10 | 0.10 |
| Combined ( | 1.2 | 1.2 | 0.91 | 0.99 |