| Literature DB >> 22844282 |
Mahnaz Sheikhi1, Sajad Ghorbanizadeh, Mehrdad Abdinian, Hossein Goroohi, Hamid Badrian.
Abstract
Introduction. The aim of this study was to determine the accuracy of linear measurements in dry human skulls in ideal position and different deviated positions of the skull. Methods. 6 dry human skulls were included in the study. Opaque markers were attached to alveolar bone. Buccolingual and mesiodistal distances and heights were measured in 5 different regions of either jaws using a digital caliper. Radiographic distances were measured in ideal, rotation, tilt, flexion, and extension positions of the skulls. The physical and radiographic measurements were compared to estimate linear measurement accuracy. Results. The mean difference between physical measurements and radiographic measurements was 0.05 ± 0.45. There was a significant difference between physical measurements and radiographic measurements in ideal, rotation, tilt, and extension positions (P value < 0.05). Conclusions. The accuracy of measurements in GALILEOUS CBCT machine varies when the position of the skull deviates from ideal; however, the differences are not clinically significant.Entities:
Year: 2012 PMID: 22844282 PMCID: PMC3403319 DOI: 10.1155/2012/214954
Source DB: PubMed Journal: Int J Dent ISSN: 1687-8728
Different physical and radiographic measured distances in molar, premolar, and anterior regions.
| Physical | Cross-sectional | Axial | Tangential | |
|---|---|---|---|---|
| Buccolingual length | Between 1st and 2nd markers | + | + | |
| Height | Between 1st and 3rd markers | + | + | |
| Mesiodistal length | Between 1st and 4th markers | + | + |
1st marker: placed on the embrasure of buccal alveolar crest.
2nd marker: placed on embrasure of lingual alveolar crest, perpendicular to the first marker.
3rd marker: placed on to the most apical region of buccal alveolar crest in the same direction with the first marker.
4th marker: placed on the adjacent tooth's embrasure of buccal alveolar crest, next to the first marker.
Figure 1Panoramic view of the skull radiograph with 10 degree left tilt. Opaque gutta-percha markers are observed in the image.
Figure 2Examples of measured distances in different views. (a) Height: the arrow shows the measured distance; (b) buccolingual and mesiodistal length.
Mean difference between physical measurements and radiographic measurements in different positions.
| Position | Mean difference ± SD (mm) |
|
|---|---|---|
| Ideal | 0.04 ± 0.44 | 0.032 |
| Rotation | 0.14 ± 0.52 | <0.001 |
| Tilt | 0.10 ± 0.53 | 0.001 |
| Flexion | 0.10 ± 0.61 | 0.107 |
| Extension | 0.04 ± 0.44 | 0.053 |
Mean difference between physical measurements and radiographic measurements in different positions, considering the measured aspect.
| Position | Aspect | Mean difference ± SD (mm) |
|
|---|---|---|---|
| Buccolingual | 0.01 ± 0.13 | 0.431 | |
| Ideal | Height | 0.15 ± 0.58 | 0.022 |
| Mesiodistal | 0.02 ± 0.49 | 0.350 | |
|
| |||
| Buccolingual | 0.17 ± 0.46 | 0.007 | |
| Rotation | Height | 0.23 ± 0.68 | 0.024 |
| Mesiodistal | 0.03 ± 0.38 | 0.041 | |
|
| |||
| Buccolingual | 0.18 ± 0.49 | 0.458 | |
| Tilt | Height | 0.12 ± 0.63 | 0.002 |
| Mesiodistal | 0.01 ± 0.44 | 0.043 | |
|
| |||
| Buccolingual | 0.13 ± 0.48 | 0.979 | |
| Flexion | Height | 0.18 ± 0.89 | 0.037 |
| Mesiodistal | 0.09 ± 0.32 | 0.317 | |
|
| |||
| Buccolingual | 0.13 ± 0.50 | 0.559 | |
| Extension | Height | 0.16 ± 0.38 | 0.004 |
| Mesiodistal | 0.01 ± 0.43 | 0.883 | |
Assessment of difference between radiographic and physical measurements: univariate ANOVA and resulting P values.
| Source | Factor |
|
|---|---|---|
| Position | 0.036 | |
| Skull | <0.001 | |
| Region | Molar, premolar, central | 0.118 |
| Jaw | Maxilla, mandible | 0.115 |
| Side | Left, right | 0.010 |
| Skull ∗ region | 0.070 | |
| Skull ∗ side | 0.006 | |
| Region ∗ position | 0.098 |
Comparison of mean differences with 0.5-mm absolute error: One-sample t-test.
| Position | Absolute value of difference ± SD (mm) |
|
|---|---|---|
| Normal | 0.05 ± 0.45 | <0.001 |
| Rotation | 0.14 ± 0.52 | <0.001 |
| Tilt | 0.35 ± 0.41 | <0.001 |
| Flexion | 0.32 ± 0.32 | 0.002 |
| Extension | 0.37 ± 0.50 | <0.001 |