Quality evaluation of image-based iterative reconstruction for CT: Comparison with hybrid iterative reconstruction.

Hiroki Kawashima1, Katsuhiro Ichikawa1, Kosuke Matsubara1, Hiroji Nagata2, Tadanori Takata3, Satoshi Kobayashi1.   

Abstract

The purpose of this study is to evaluate the physical image quality of a commercially available image-based iterative reconstruction (IIR) system for two object contrasts to resemble a soft tissue (60 HU) and an enhanced vessel (270 HU), and compare the results with those of filtered back projection (FBP) and iterative reconstruction (IR). A 192-slice computed tomography (CT) scanner was used for data acquisitions. IIR images were processed from the FBP images. Task-based in-plane transfer function (TTF) and slice sensitivity profile (SSPtask ) were measured from rod objects inside of a 25-cm diameter water phantom at four dose levels (2.5, 5, 10, and 20 mGy). Noise power spectrum (NPS) was measured from the water-only part. System performance (SP) function was calculated as TTF2 /NPS over FBP, IR, and IIR for comparison. In addition, an image subtraction was performed using images of rod objects, a bar-pattern phantom, and a clinical abdomen case to observe the noise reduction performance of IIR. As a results, IIR mostly preserved TTF and SSPtask of FBP, whereas IR exhibited enhanced TTF at 10 and 20 mGy for 60 HU contrast and at all doses for 270 HU contrast. SP of IIR at 2.5, 5, 10 mGy (half doses) were similar to those of FBP at 5, 10, 20 mGy, respectively. IR exhibited enhanced SP at medium to high frequencies. The subtracted images showed weak remained edge signals in the bar-pattern and abdominal images. In conclusion, IIR uniformly improved the task-based image quality of FBP over the entire frequency range, whereas IR improved the characteristics over medium to high frequencies. The dose reduction potential of IIR estimated from SP is approximately 50%, when allowing the slight signal reductions.
© 2019 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  computed tomography; dose reduction; image quality; iterative reconstruction; noise reduction

Mesh:

Year:  2019        PMID: 31050148      PMCID: PMC6560231          DOI: 10.1002/acm2.12597

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  14 in total

1.  Accuracy of a simple method for deriving the presampled modulation transfer function of a digital radiographic system from an edge image.

Authors:  Egbert Buhr; Susanne Günther-Kohfahl; Ulrich Neitzel
Journal:  Med Phys       Date:  2003-09       Impact factor: 4.071

2.  Dose reduction in pediatric CT: a rational approach.

Authors:  John M Boone; Estella M Geraghty; J Anthony Seibert; Sandra L Wootton-Gorges
Journal:  Radiology       Date:  2003-08       Impact factor: 11.105

Review 3.  CT dose reduction and dose management tools: overview of available options.

Authors:  Cynthia H McCollough; Michael R Bruesewitz; James M Kofler
Journal:  Radiographics       Date:  2006 Mar-Apr       Impact factor: 5.333

4.  Assessment of the dose reduction potential of a model-based iterative reconstruction algorithm using a task-based performance metrology.

Authors:  Ehsan Samei; Samuel Richard
Journal:  Med Phys       Date:  2015-01       Impact factor: 4.071

5.  An Improved Index of Image Quality for Task-based Performance of CT Iterative Reconstruction across Three Commercial Implementations.

Authors:  Olav Christianson; Joseph J S Chen; Zhitong Yang; Ganesh Saiprasad; Alden Dima; James J Filliben; Adele Peskin; Christopher Trimble; Eliot L Siegel; Ehsan Samei
Journal:  Radiology       Date:  2015-02-13       Impact factor: 11.105

6.  Towards task-based assessment of CT performance: system and object MTF across different reconstruction algorithms.

Authors:  Samuel Richard; Daniela B Husarik; Girijesh Yadava; Simon N Murphy; Ehsan Samei
Journal:  Med Phys       Date:  2012-07       Impact factor: 4.071

7.  Preliminary results: prospective clinical study to assess image-based iterative reconstruction for abdominal computed tomography acquired at 2 radiation dose levels.

Authors:  Sarvenaz Pourjabbar; Sarabjeet Singh; Anand K Singh; Rocio P Johnston; Anuradha S Shenoy-Bhangle; Synho Do; Atul Padole; Michael A Blake; Anders Persson; Mannudeep K Kalra
Journal:  J Comput Assist Tomogr       Date:  2014 Jan-Feb       Impact factor: 1.826

8.  Submillisievert chest CT with filtered back projection and iterative reconstruction techniques.

Authors:  Atul Padole; Sarabjeet Singh; Jeanne B Ackman; Carol Wu; Synho Do; Sarvenaz Pourjabbar; Ranish Deedar Ali Khawaja; Alexi Otrakji; Subba Digumarthy; Jo-Anne Shepard; Mannudeep Kalra
Journal:  AJR Am J Roentgenol       Date:  2014-10       Impact factor: 3.959

9.  Physical characterization of a new CT iterative reconstruction method operating in sinogram space.

Authors:  Caterina Ghetti; Francesca Palleri; Giulio Serreli; Ornella Ortenzia; Livia Ruffini
Journal:  J Appl Clin Med Phys       Date:  2013-07-08       Impact factor: 2.102

10.  A phantom study investigating the relationship between ground-glass opacity visibility and physical detectability index in low-dose chest computed tomography.

Authors:  Katsuhiro Ichikawa; Takeshi Kobayashi; Motoyasu Sagawa; Ayako Katagiri; Yukiko Uno; Ryo Nishioka; Jun Matsuyama
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

View more
  1 in total

1.  Performance of clinically available deep learning image reconstruction in computed tomography: a phantom study.

Authors:  Hiroki Kawashima; Katsuhiro Ichikawa; Tadanori Takata; Wataru Mitsui; Hiroshi Ueta; Norihide Yoneda; Satoshi Kobayashi
Journal:  J Med Imaging (Bellingham)       Date:  2020-12-16
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.