Accuracy of non-invasive measurement of haemoglobin concentration by pulse co-oximetry during steady-state and dynamic conditions in liver surgery.

J J Vos1, A F Kalmar, M M R F Struys, R J Porte, J K G Wietasch, T W L Scheeren, H G D Hendriks.   

Abstract

BACKGROUND: The Masimo Radical 7 (Masimo Corp., Irvine, CA, USA) pulse co-oximeter(®) calculates haemoglobin concentration (SpHb) non-invasively using transcutaneous spectrophotometry. We compared SpHb with invasive satellite-lab haemoglobin monitoring (Hb(satlab)) during major hepatic resections both under steady-state conditions and in a dynamic phase with fluid administration of crystalloid and colloid solutions.
METHODS: Thirty patients undergoing major hepatic resection were included and randomized to receive a fluid bolus of 15 ml kg(-1) colloid (n=15) or crystalloid (n=15) solution over 30 min. SpHb was continuously measured on the index finger, and venous blood samples were analysed in both the steady-state phase (from induction until completion of parenchymal transection) and the dynamic phase (during fluid bolus).
RESULTS: Correlation was significant between SpHb and Hb(satlab) (R(2)=0.50, n=543). The modified Bland-Altman analysis for repeated measurements showed a bias (precision) of -0.27 (1.06) and -0.02 (1.07) g dl(-1) for the steady-state and dynamic phases, respectively. SpHb accuracy increased when Hb(satlab) was <10 g dl(-1), with a bias (precision) of 0.41 (0.47) vs -0.26 (1.12) g dl(-1) for values >10 g dl(-1), but accuracy decreased after colloid administration (R(2)=0.25).
CONCLUSIONS: SpHb correlated moderately with Hb(satlab) with a slight underestimation in both phases in patients undergoing major hepatic resection. Accuracy increased for lower Hb(satlab) values but decreased in the presence of colloid solution. Further improvements are necessary to improve device accuracy under these conditions, so that SpHb might become a sensitive screening device for clinically significant anaemia.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22822043     DOI: 10.1093/bja/aes234

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  17 in total

Review 1.  Accuracy of pleth variability index to predict fluid responsiveness in mechanically ventilated patients: a systematic review and meta-analysis.

Authors:  Haitao Chu; Yong Wang; Yanfei Sun; Gang Wang
Journal:  J Clin Monit Comput       Date:  2015-08-05       Impact factor: 2.502

2.  Evaluation of multiwave pulse total-hemoglobinometer during general anesthesia.

Authors:  Daisuke Toyoda; Rie Yasumura; Mitsue Fukuda; Ryoichi Ochiai; Yoshifumi Kotake
Journal:  J Anesth       Date:  2013-10-22       Impact factor: 2.078

3.  Does a non-invasive hemoglobin monitor correlate with a venous blood sample in the acutely ill?

Authors:  Benjamin A von Schweinitz; Robert A De Lorenzo; Peter J Cuenca; Richard L Anschutz; Paul B Allen
Journal:  Intern Emerg Med       Date:  2014-10-17       Impact factor: 3.397

4.  Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.

Authors:  Hosein Shabaninejad; Nashmil Ghadimi; Kourosh Sayehmiri; Hossein Hosseinifard; Rasoul Azarfarin; Hasan Abolghasem Gorji
Journal:  J Anesth       Date:  2019-03-20       Impact factor: 2.078

5.  Impact of acute changes in perfusion index and blood pressure on the accuracy of non-invasive continuous hemoglobin concentration measurements during induction of anesthesia.

Authors:  Junichi Saito; Masato Kitayama; Erika Amanai; Kentaro Toyooka; Kazuyoshi Hirota
Journal:  J Anesth       Date:  2017-01-09       Impact factor: 2.078

6.  Noninvasively Measured Hemoglobin Concentration Reflects Arterial Hemoglobin Concentration Before but Not After Cardiopulmonary Bypass in Patients Undergoing Coronary Artery or Valve Surgery.

Authors:  Matthias L Riess; Paul S Pagel
Journal:  J Cardiothorac Vasc Anesth       Date:  2016-03-24       Impact factor: 2.628

7.  Microwaving blood as a non-destructive technique for haemoglobin measurements on microlitre samples.

Authors:  Toby H Basey-Fisher; Nadia Guerra; Chiara Triulzi; Andrew Gregory; Stephen M Hanham; Molly M Stevens; Stefan A Maier; Norbert Klein
Journal:  Adv Healthc Mater       Date:  2013-09-04       Impact factor: 9.933

8.  Accuracy of non-invasive continuous total hemoglobin measurement by Pulse CO-Oximetry in severe traumatized and surgical bleeding patients.

Authors:  Werner Baulig; Burkhardt Seifert; Donat R Spahn; Oliver M Theusinger
Journal:  J Clin Monit Comput       Date:  2015-12-19       Impact factor: 2.502

9.  The accuracy of non-invasively continuous total hemoglobin measurement by pulse CO-Oximetry undergoing acute normovolemic hemodilution and reinfusion of autologous blood.

Authors:  Junichi Saito; Masato Kitayama; Masafumi Oishi; Tomoyuki Kudo; Masahiro Sawada; Hiroshi Hashimoto; Kazuyoshi Hirota
Journal:  J Anesth       Date:  2014-06-28       Impact factor: 2.078

10.  Accuracy of non-invasive haemoglobin measurements in patients undergoing transurethral resection of prostate surgery.

Authors:  Naveen Selvaraj; Gaurav Jain; Debendra Kumar Tripathy; Ankur Mittal; Haritha Indulekha
Journal:  Indian J Anaesth       Date:  2021-05-10
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.