Impact of acute changes in perfusion index and blood pressure on the accuracy of non-invasive continuous hemoglobin concentration measurements during induction of anesthesia.

Junichi Saito1, Masato Kitayama2, Erika Amanai3, Kentaro Toyooka3, Kazuyoshi Hirota3.   

Abstract

PURPOSE: Several factors affect the accuracy of non-invasive continuous hemoglobin concentration (SpHb) measurements. We had previously shown an increase in the perfusion index (PI) following induction of anesthesia which was associated with an increase in the difference between SpHb and total hemoglobin (tHb) (SpHb-tHb). We hypothesized that blunting the increase in PI by maintaining blood pressure during induction of anesthesia would improve the agreement between SpHb and tHb measurements.
METHODS: Twenty-nine adult patients were enrolled. Patients were randomly assigned by use of sequentially numbered, opaque sealed envelopes to a control (group C) or a phenylephrine group (group P). Anesthesia was induced and maintained with propofol, remifentanil, and ketamine. In group P, phenylephrine was infused at 0.5 µg/kg/min during induction of anesthesia. SpHb and PI were monitored with a Radical-7 Pulse CO-Oximeter. tHb and hematocrit were measured with the ABL800 blood gas analyzer.
RESULTS: Following induction of anesthesia, PI increased significantly in both groups (p < 0.001 and p < 0.05 in groups C and P, respectively). However, the increase in PI was significantly smaller in group P than in group C (2.6 ± 1.3 vs 0.8 ± 1.4%, p < 0.001). Similarly, the change in SpHb-tHb was significantly smaller in group P than in group C (0.40 ± 0.78 vs 0.97 ± 0.70 g/dl, p < 0.05). Changes in SpHb-tHb are correlated with changes in PI (r = 0.46, p < 0.05).
CONCLUSIONS: The findings suggest that blunting the increase in PI by maintaining arterial pressure during induction of anesthesia improves the agreement between SpHb and tHb values.

Entities:  

Keywords:  Anesthesia; Non-invasive continuous hemoglobin concentration; Perfusion index

Mesh:

Substances:

Year:  2017        PMID: 28070653     DOI: 10.1007/s00540-017-2306-6

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  10 in total

1.  Volume kinetic analysis of the distribution of 0.9% saline in conscious versus isoflurane-anesthetized sheep.

Authors:  Kirk I Brauer; Christer Svensén; Robert G Hahn; Lilian D Traber; Donald S Prough
Journal:  Anesthesiology       Date:  2002-02       Impact factor: 7.892

2.  Pressure-dependent changes in haematocrit and plasma volume during anaesthesia, a randomised clinical trial.

Authors:  T Damén; B Reinsfelt; B Redfors; A Nygren
Journal:  Acta Anaesthesiol Scand       Date:  2016-01-21       Impact factor: 2.105

3.  The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery.

Authors:  Ba-Vinh Nguyen; Jean-Louis Vincent; Emmanuel Nowak; Michelle Coat; Nicolas Paleiron; Pierre Gouny; Mehdi Ould-Ahmed; Maité Guillouet; Charles Christian Arvieux; Gildas Gueret
Journal:  Anesth Analg       Date:  2011-09-14       Impact factor: 5.108

4.  Non-invasive blood haemoglobin and plethysmographic variability index during brachial plexus block.

Authors:  C Bergek; J H Zdolsek; R G Hahn
Journal:  Br J Anaesth       Date:  2015-01-20       Impact factor: 9.166

5.  Accuracy of non-invasive measurement of haemoglobin concentration by pulse co-oximetry during steady-state and dynamic conditions in liver surgery.

Authors:  J J Vos; A F Kalmar; M M R F Struys; R J Porte; J K G Wietasch; T W L Scheeren; H G D Hendriks
Journal:  Br J Anaesth       Date:  2012-07-20       Impact factor: 9.166

6.  A comparison of three methods of hemoglobin monitoring in patients undergoing spine surgery.

Authors:  Ronald D Miller; Theresa A Ward; Stephen C Shiboski; Neal H Cohen
Journal:  Anesth Analg       Date:  2011-03-08       Impact factor: 5.108

7.  A comparison of lidocaine and bupivacaine digital nerve blocks on noninvasive continuous hemoglobin monitoring in a randomized trial in volunteers.

Authors:  Ronald D Miller; Theresa A Ward; Charles E McCulloch; Neal H Cohen
Journal:  Anesth Analg       Date:  2014-04       Impact factor: 5.108

8.  Trending and accuracy of noninvasive hemoglobin monitoring in pediatric perioperative patients.

Authors:  Mario Patino; Lindsay Schultz; Monir Hossain; Jennifer Moeller; Mohamed Mahmoud; Joel Gunter; C Dean Kurth
Journal:  Anesth Analg       Date:  2014-10       Impact factor: 5.108

9.  The accuracy of non-invasively continuous total hemoglobin measurement by pulse CO-Oximetry undergoing acute normovolemic hemodilution and reinfusion of autologous blood.

Authors:  Junichi Saito; Masato Kitayama; Masafumi Oishi; Tomoyuki Kudo; Masahiro Sawada; Hiroshi Hashimoto; Kazuyoshi Hirota
Journal:  J Anesth       Date:  2014-06-28       Impact factor: 2.078

10.  Does a digital regional nerve block improve the accuracy of noninvasive hemoglobin monitoring?

Authors:  Ronald D Miller; Theresa A Ward; Charles E McCulloch; Neal H Cohen
Journal:  J Anesth       Date:  2012-08-01       Impact factor: 2.078

  10 in total
  2 in total

1.  Accuracy of Continuous and Noninvasive Hemoglobin Monitoring in the Presence of CO2 Insufflation: An Observational Pilot Study.

Authors:  Hye Sun Lee; Ji Won Yoo; Ha Yeon Kim; Na Young Kim; Ji Eun Kim
Journal:  Med Sci Monit       Date:  2021-10-27

2.  Effects of different mean arterial pressure targets on plasma volume, ANP and glycocalyx-A randomized trial.

Authors:  Tor Damén; Sofia Saadati; Eva Forssell-Aronsson; Camilla Hesse; Peter Bentzer; Sven-Erik Ricksten; Andreas Nygren
Journal:  Acta Anaesthesiol Scand       Date:  2020-10-05       Impact factor: 2.105

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.