The accuracy of non-invasively continuous total hemoglobin measurement by pulse CO-Oximetry undergoing acute normovolemic hemodilution and reinfusion of autologous blood.

Junichi Saito1, Masato Kitayama, Masafumi Oishi, Tomoyuki Kudo, Masahiro Sawada, Hiroshi Hashimoto, Kazuyoshi Hirota.   

Abstract

BACKGROUND: Non-invasively continuous total hemoglobin (SpHb) measurement has not been assessed adequately in acute bleeding and rapid blood transfusion during surgery. Thus, we have assessed the efficacy of SpHb during both acute normovolemic hemodilution (ANH) and autologous blood transfusion (ABT).
METHODS: Twenty-four patients undergoing urological and gynecological surgery were enrolled. ANH was induced by withdrawing blood of 800 g with simultaneous fluid administration. When surgical hemostasis was completed, collected blood was reinfused. Measurement of SpHb, perfusion index (PI) and real total Hb (tHb) were done before and after each 400 ml blood removal (-0, -400, -800 ml) and reinfusion (+0, +400, +800 ml).
RESULTS: A Bland-Altman analysis for repeated measurements showed a bias (precision) g/dl of 1.12 (1.25), 1.43 (1.24) and 1.10 (1.23) for all data, during ANH and during ABT, respectively. Additionally, a bias (precision) increased with a reduction in tHb (g/dl): ≥10.0; 0.74 (1.30), 8.0-10.0; 1.15 (1.12) and <8.0; 1.60 (1.28). Although the difference between SpHb and tHb was almost zero before anesthesia induction, it became significant just before ANH and did not change further by ANH and ABT. Significant correlations between SpHb and tHb for all data (r = 0.75, n = 228, p < 0.001) were observed. PI slightly correlated with the difference between SpHb and tHb (r = 0.38, n = 216, p < 0.001). Furthermore, before and after induction of anesthesia, PI also correlated with the difference between SpHb and tHb (r = 0.42, n = 23, p = 0.048 and r = 0.51, n = 22, p = 0.016, respectively).
CONCLUSIONS: The present data suggest that SpHb may overestimate tHb during ANH and ABT. In addition, PI and tHb levels had an impact on the accuracy of SpHb measurements.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 24972855     DOI: 10.1007/s00540-014-1863-1

Source DB:  PubMed          Journal:  J Anesth        ISSN: 0913-8668            Impact factor:   2.078


  18 in total

Review 1.  Rheology of the microcirculation.

Authors:  A R Pries; T W Secomb
Journal:  Clin Hemorheol Microcirc       Date:  2003       Impact factor: 2.375

2.  Error in noninvasive spectrophotometric measurement of blood hemoglobin concentration under conditions of blood loss.

Authors:  Rotem Naftalovich; Daniel Naftalovich
Journal:  Med Hypotheses       Date:  2011-08-05       Impact factor: 1.538

3.  The accuracy of noninvasive hemoglobin measurement by multiwavelength pulse oximetry after cardiac surgery.

Authors:  Ba-Vinh Nguyen; Jean-Louis Vincent; Emmanuel Nowak; Michelle Coat; Nicolas Paleiron; Pierre Gouny; Mehdi Ould-Ahmed; Maité Guillouet; Charles Christian Arvieux; Gildas Gueret
Journal:  Anesth Analg       Date:  2011-09-14       Impact factor: 5.108

4.  Perfusion index and plethysmographic variability index in patients with interscalene nerve catheters.

Authors:  Anne Sebastiani; Larissa Philippi; Stefan Boehme; Dorothea Closhen; Irene Schmidtmann; Anton Scherhag; Klaus Markstaller; Kristin Engelhard; Gunther Pestel
Journal:  Can J Anaesth       Date:  2012-10-02       Impact factor: 5.063

5.  Continuous noninvasive hemoglobin monitoring during complex spine surgery.

Authors:  Lauren Berkow; Stephanie Rotolo; Erin Mirski
Journal:  Anesth Analg       Date:  2011-09-29       Impact factor: 5.108

6.  Comparison of the accuracy of noninvasive hemoglobin monitoring by spectrophotometry (SpHb) and HemoCue® with automated laboratory hemoglobin measurement.

Authors:  Lionel Lamhaut; Roxana Apriotesei; Xavier Combes; Marc Lejay; Pierre Carli; Benoît Vivien
Journal:  Anesthesiology       Date:  2011-09       Impact factor: 7.892

7.  Validation of noninvasive hemoglobin measurements using the Masimo Radical-7 SpHb Station.

Authors:  Marlin Wayne Causey; Seth Miller; Andrew Foster; Alec Beekley; David Zenger; Matthew Martin
Journal:  Am J Surg       Date:  2011-05       Impact factor: 2.565

8.  Use of acute normovolemic hemodilution in patients undergoing radical prostatectomy.

Authors:  Akito Terai; Naoki Terada; Koji Yoshimura; Kentaro Ichioka; Nobufumi Ueda; Noriaki Utsunomiya; Naoki Kohei; Yoichi Arai; Akitomo Yonei
Journal:  Urology       Date:  2005-06       Impact factor: 2.649

9.  Intraoperative moderate acute normovolemic hemodilution associated with a comprehensive blood-sparing protocol in off-pump coronary surgery.

Authors:  Valter Casati; Stefano Benussi; Luca Sandrelli; Maria Antonietta Grasso; Salvatore Spagnolo; Armando D'Angelo
Journal:  Anesth Analg       Date:  2004-05       Impact factor: 5.108

10.  Pulse oximeter perfusion index as an early indicator of sympathectomy after epidural anesthesia.

Authors:  Y Ginosar; C F Weiniger; Y Meroz; V Kurz; T Bdolah-Abram; A Babchenko; M Nitzan; E M Davidson
Journal:  Acta Anaesthesiol Scand       Date:  2009-04-24       Impact factor: 2.105

View more
  6 in total

1.  Comparison of invasive and noninvasive blood hemoglobin measurement in the operating room: a systematic review and meta-analysis.

Authors:  Hosein Shabaninejad; Nashmil Ghadimi; Kourosh Sayehmiri; Hossein Hosseinifard; Rasoul Azarfarin; Hasan Abolghasem Gorji
Journal:  J Anesth       Date:  2019-03-20       Impact factor: 2.078

2.  Impact of acute changes in perfusion index and blood pressure on the accuracy of non-invasive continuous hemoglobin concentration measurements during induction of anesthesia.

Authors:  Junichi Saito; Masato Kitayama; Erika Amanai; Kentaro Toyooka; Kazuyoshi Hirota
Journal:  J Anesth       Date:  2017-01-09       Impact factor: 2.078

3.  Continuous hemoglobin measurement during frontal advancement operations can improve patient outcomes.

Authors:  Ayten Saracoglu; Ruslan Abdullayev; Mustafa Sakar; Bulent Sacak; Feyza Girgin Incekoy; Zuhal Aykac
Journal:  J Clin Monit Comput       Date:  2022-03-07       Impact factor: 2.502

4.  Meta-analytic estimation of measurement variability and assessment of its impact on decision-making: the case of perioperative haemoglobin concentration monitoring.

Authors:  Emmanuel Charpentier; Vincent Looten; Björn Fahlgren; Alexandre Barna; Loïc Guillevin
Journal:  BMC Med Res Methodol       Date:  2016-01-19       Impact factor: 4.615

5.  Intraoperative continuous noninvasive hemoglobin monitoring in patients with placenta previa undergoing cesarean section: a prospective observational study.

Authors:  Hyunsung Kim; Sang-Hwan Do; Jung-Won Hwang; Hyo-Seok Na
Journal:  Anesth Pain Med (Seoul)       Date:  2019-10-31

6.  Accuracy of Continuous and Noninvasive Hemoglobin Monitoring in the Presence of CO2 Insufflation: An Observational Pilot Study.

Authors:  Hye Sun Lee; Ji Won Yoo; Ha Yeon Kim; Na Young Kim; Ji Eun Kim
Journal:  Med Sci Monit       Date:  2021-10-27
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.