Quality assurance of dynamic parameters in volumetric modulated arc therapy.

A Manikandan1, B Sarkar, R Holla, T R Vivek, N Sujatha.   

Abstract

OBJECTIVES: The purpose of this study was to demonstrate quality assurance checks for accuracy of gantry speed and position, dose rate and multileaf collimator (MLC) speed and position for a volumetric modulated arc treatment (VMAT) modality (Synergy S; Elekta, Stockholm, Sweden), and to check that all the necessary variables and parameters were synchronous.
METHODS: Three tests (for gantry position-dose delivery synchronisation, gantry speed-dose delivery synchronisation and MLC leaf speed and positions) were performed.
RESULTS: The average error in gantry position was 0.5° and the average difference was 3 MU for a linear and a parabolic relationship between gantry position and delivered dose. In the third part of this test (sawtooth variation), the maximum difference was 9.3 MU, with a gantry position difference of 1.2°. In the sweeping field method test, a linear relationship was observed between recorded doses and distance from the central axis, as expected. In the open field method, errors were encountered at the beginning and at the end of the delivery arc, termed the "beginning" and "end" errors. For MLC position verification, the maximum error was -2.46 mm and the mean error was 0.0153 ±0.4668 mm, and 3.4% of leaves analysed showed errors of >±1 mm.
CONCLUSION: This experiment demonstrates that the variables and parameters of the Synergy S are synchronous and that the system is suitable for delivering VMAT using a dynamic MLC.

Mesh:

Year:  2012        PMID: 22745206      PMCID: PMC3474081          DOI: 10.1259/bjr/19152959

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  20 in total

1.  Leaf position error during conformal dynamic arc and intensity modulated arc treatments.

Authors:  C R Ramsey; K M Spencer; R Alhakeem; A L Oliver
Journal:  Med Phys       Date:  2001-01       Impact factor: 4.071

2.  Aperture modulated arc therapy.

Authors:  S M Crooks; Xiaodong Wu; C Takita; M Watzich; Lei Xing
Journal:  Phys Med Biol       Date:  2003-05-21       Impact factor: 3.609

3.  Quantitative measurement of MLC leaf displacements using an electronic portal image device.

Authors:  Yong Yang; Lei Xing
Journal:  Phys Med Biol       Date:  2004-04-21       Impact factor: 3.609

4.  An extensive log-file analysis of step-and-shoot intensity modulated radiation therapy segment delivery errors.

Authors:  Anthony M Stell; Jonathan G Li; Omar A Zeidan; James F Dempsey
Journal:  Med Phys       Date:  2004-06       Impact factor: 4.071

5.  Sweeping-window arc therapy: an implementation of rotational IMRT with automatic beam-weight calculation.

Authors:  C Cameron
Journal:  Phys Med Biol       Date:  2005-09-07       Impact factor: 3.609

6.  Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration.

Authors:  S J K Baker; G J Budgell; R I MacKay
Journal:  Phys Med Biol       Date:  2005-03-16       Impact factor: 3.609

7.  Evaluation of two methods of predicting MLC leaf positions using EPID measurements.

Authors:  Laure Parent; Joao Seco; Phil M Evans; David R Dance; Andrew Fielding
Journal:  Med Phys       Date:  2006-09       Impact factor: 4.071

8.  Development of an optimization concept for arc-modulated cone beam therapy.

Authors:  Silke Ulrich; Simeon Nill; Uwe Oelfke
Journal:  Phys Med Biol       Date:  2007-06-13       Impact factor: 3.609

9.  Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee.

Authors:  Gary A Ezzell; James M Galvin; Daniel Low; Jatinder R Palta; Isaac Rosen; Michael B Sharpe; Ping Xia; Ying Xiao; Lei Xing; Cedric X Yu
Journal:  Med Phys       Date:  2003-08       Impact factor: 4.071

10.  Patient-specific quality assurance method for VMAT treatment delivery.

Authors:  Eduard Schreibmann; Anees Dhabaan; Eric Elder; Tim Fox
Journal:  Med Phys       Date:  2009-10       Impact factor: 4.071

View more
  12 in total

1.  Independent calculation of monitor units for VMAT and SPORT.

Authors:  Xin Chen; Karl Bush; Aiping Ding; Lei Xing
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

2.  The effect of MLC speed and acceleration on the plan delivery accuracy of VMAT.

Authors:  J M Park; H-G Wu; J H Kim; J N K Carlson; K Kim
Journal:  Br J Radiol       Date:  2015-03-03       Impact factor: 3.039

3.  Advantage of 3D volumetric dosemeter in delivery quality assurance of dynamic arc therapy: comparison of pencil beam and Monte Carlo calculations.

Authors:  H-J Shin; J H Song; J-Y Jung; Y-K Kwak; C S Kay; Y-N Kang; B O Choi; H S Jang; S H Son
Journal:  Br J Radiol       Date:  2013-11-14       Impact factor: 3.039

4.  Increasing frequency of reirradiation studies in radiation oncology: systematic review of highly cited articles.

Authors:  Carsten Nieder; Nicolaus H Andratschke; Anca L Grosu
Journal:  Am J Cancer Res       Date:  2013-04-03       Impact factor: 6.166

5.  Texture analysis on the edge-enhanced fluence of VMAT.

Authors:  So-Yeon Park; Jong Min Park; Wonmo Sung; Il Han Kim; Sung-Joon Ye
Journal:  Radiat Oncol       Date:  2015-04-01       Impact factor: 3.481

6.  3D VMAT Verification Based on Monte Carlo Log File Simulation with Experimental Feedback from Film Dosimetry.

Authors:  A R Barbeiro; A Ureba; J A Baeza; R Linares; M Perucha; E Jiménez-Ortega; S Velázquez; J C Mateos; A Leal
Journal:  PLoS One       Date:  2016-11-21       Impact factor: 3.240

7.  VMAT linear accelerator commissioning and quality assurance: dose control and gantry speed tests.

Authors:  Michael P Barnes; Pejman Rowshanfarzad; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

8.  Electronic Portal Imaging Device-Based Three-Dimensional Volumetric Dosimetry for Intensity-modulated Radiotherapy Pretreatment Quality Assurance.

Authors:  Manikandan Arjunan; Sureka Chandra Sekaran; Biplab Sarkar; Saran Kumar Manavalan
Journal:  J Med Phys       Date:  2019 Jul-Sep

9.  Prediction of VMAT delivery accuracy with textural features calculated from fluence maps.

Authors:  Jong Min Park; Jung-In Kim; So-Yeon Park
Journal:  Radiat Oncol       Date:  2019-12-23       Impact factor: 3.481

10.  Detector system dose verification comparisons for arc therapy: couch vs. gantry mount.

Authors:  Arjunan Manikandan; Biplab Sarkar; Maitreyee Nandy; Chandra Sekaran Sureka; Michael S Gossman; Nadendla Sujatha; Vivek Thirupathur Rajendran
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.