Independent calculation of monitor units for VMAT and SPORT.

Xin Chen1, Karl Bush1, Aiping Ding1, Lei Xing1.   

Abstract

PURPOSE: Dose and monitor units (MUs) represent two important facets of a radiation therapy treatment. In current practice, verification of a treatment plan is commonly done in dose domain, in which a phantom measurement or forward dose calculation is performed to examine the dosimetric accuracy and the MU settings of a given treatment plan. While it is desirable to verify directly the MU settings, a computational framework for obtaining the MU values from a known dose distribution has yet to be developed. This work presents a strategy to calculate independently the MUs from a given dose distribution of volumetric modulated arc therapy (VMAT) and station parameter optimized radiation therapy (SPORT).
METHODS: The dose at a point can be expressed as a sum of contributions from all the station points (or control points). This relationship forms the basis of the proposed MU verification technique. To proceed, the authors first obtain the matrix elements which characterize the dosimetric contribution of the involved station points by computing the doses at a series of voxels, typically on the prescription surface of the VMAT/SPORT treatment plan, with unit MU setting for all the station points. An in-house Monte Carlo (MC) software is used for the dose matrix calculation. The MUs of the station points are then derived by minimizing the least-squares difference between doses computed by the treatment planning system (TPS) and that of the MC for the selected set of voxels on the prescription surface. The technique is applied to 16 clinical cases with a variety of energies, disease sites, and TPS dose calculation algorithms.
RESULTS: For all plans except the lung cases with large tissue density inhomogeneity, the independently computed MUs agree with that of TPS to within 2.7% for all the station points. In the dose domain, no significant difference between the MC and Eclipse Anisotropic Analytical Algorithm (AAA) dose distribution is found in terms of isodose contours, dose profiles, gamma index, and dose volume histogram (DVH) for these cases. For the lung cases, the MC-calculated MUs differ significantly from that of the treatment plan computed using AAA. However, the discrepancies are reduced to within 3% when the TPS dose calculation algorithm is switched to a transport equation-based technique (Acuros™). Comparison in the dose domain between the MC and Eclipse AAA/Acuros calculation yields conclusion consistent with the MU calculation.
CONCLUSIONS: A computational framework relating the MU and dose domains has been established. The framework does not only enable them to verify the MU values of the involved station points of a VMAT plan directly in the MU domain but also provide a much needed mechanism to adaptively modify the MU values of the station points in accordance to a specific change in the dose domain.

Mesh:

Year:  2015        PMID: 25652504      PMCID: PMC4312348          DOI: 10.1118/1.4906185

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  31 in total

1.  Monitor unit calculation for an intensity modulated photon field by a simple scatter-summation algorithm.

Authors:  L Xing; Y Chen; G Luxton; J G Li; A L Boyer
Journal:  Phys Med Biol       Date:  2000-03       Impact factor: 3.609

2.  Monte Carlo simulation of photon migration in a cloud computing environment with MapReduce.

Authors:  Guillem Pratx; Lei Xing
Journal:  J Biomed Opt       Date:  2011-12       Impact factor: 3.170

3.  Quality assurance of dynamic parameters in volumetric modulated arc therapy.

Authors:  A Manikandan; B Sarkar; R Holla; T R Vivek; N Sujatha
Journal:  Br J Radiol       Date:  2012-07       Impact factor: 3.039

4.  Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee.

Authors:  Gary A Ezzell; James M Galvin; Daniel Low; Jatinder R Palta; Isaac Rosen; Michael B Sharpe; Ping Xia; Ying Xiao; Lei Xing; Cedric X Yu
Journal:  Med Phys       Date:  2003-08       Impact factor: 4.071

5.  A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems.

Authors:  Mohammad Hussein; Pejman Rowshanfarzad; Martin A Ebert; Andrew Nisbet; Catharine H Clark
Journal:  Radiother Oncol       Date:  2013-10-04       Impact factor: 6.280

6.  Point/counterpoint. DASSIM-RT is likely to become the method of choice over conventional IMRT and VMAT for delivery of highly conformal radiotherapy.

Authors:  Lei Xing; Mark H Phillips; Colin G Orton
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

7.  An adaptive planning strategy for station parameter optimized radiation therapy (SPORT): Segmentally boosted VMAT.

Authors:  Ruijiang Li; Lei Xing
Journal:  Med Phys       Date:  2013-05       Impact factor: 4.071

8.  Toward real-time Monte Carlo simulation using a commercial cloud computing infrastructure.

Authors:  Henry Wang; Yunzhi Ma; Guillem Pratx; Lei Xing
Journal:  Phys Med Biol       Date:  2011-08-12       Impact factor: 3.609

9.  Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA.

Authors:  Heming Zhen; Benjamin E Nelms; Wolfgang A Tome
Journal:  Med Phys       Date:  2011-10       Impact factor: 4.071

10.  An end-to-end examination of geometric accuracy of IGRT using a new digital accelerator equipped with onboard imaging system.

Authors:  Lei Wang; Kayla N Kielar; Ed Mok; Annie Hsu; Sonja Dieterich; Lei Xing
Journal:  Phys Med Biol       Date:  2012-01-18       Impact factor: 3.609

View more
  2 in total

1.  Monte Carlo modeling of ultrasound probes for image guided radiotherapy.

Authors:  Magdalena Bazalova-Carter; Jeffrey Schlosser; Josephine Chen; Dimitre Hristov
Journal:  Med Phys       Date:  2015-10       Impact factor: 4.071

2.  A Homogeneous Water-Equivalent Anthropomorphic Phantom for Dosimetric Verification of Radiotherapy Plans.

Authors:  Manikandan Arjunan; Sureka Chandra Sekaran; Biplab Sarkar; Sujatha Manikandan
Journal:  J Med Phys       Date:  2018 Apr-Jun
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.