Advantage of 3D volumetric dosemeter in delivery quality assurance of dynamic arc therapy: comparison of pencil beam and Monte Carlo calculations.

H-J Shin1, J H Song, J-Y Jung, Y-K Kwak, C S Kay, Y-N Kang, B O Choi, H S Jang, S H Son.   

Abstract

OBJECTIVE: To evaluate the accuracy of pencil beam calculation (PBC) and Monte Carlo calculation (MCC) for dynamic arc therapy (DAT) in a cylindrically shaped homogenous phantom, by comparing the two plans with an ion chamber, a film and a three-dimensional (3D) volumetric dosemeter.
METHODS: For this study, an in-house phantom was constructed, and the PBC and MCC plans for DAT were performed using iPlan® RT (BrainLAB®, Heimstetten, Germany). The A16 micro ion chamber (Standard Imaging, Middleton, WI), Gafchromic® EBT2 film (International Specialty Products, Wayne, NJ) and ArcCHECK™ (Sun Nuclear, Melbourne, FL) were used for measurements. For comparison with each plan, two-dimensional (2D) and 3D gamma analyses were performed using 3%/3 mm and 2%/2 mm criteria.
RESULTS: The difference between the PBC and MCC plans using 2D and 3D gamma analyses was found to be 7.85% and 28.8%, respectively. The ion chamber and 2D dose distribution measurements did not exhibit this difference revealed by the comparison between the PBC and MCC plans. However, the 3D assessment showed a significant difference between the PBC and MCC (62.7% for PBC vs 93.4% for MCC, p = 0.034).
CONCLUSION: Evaluation using a 3D volumetric dosemeter can be clinically useful for delivery quality assurance (QA), and the MCC should be used to achieve the most reliable dose calculation for DAT. ADVANCES IN KNOWLEDGE: (1) The DAT plan calculated using the PBC has a limitation in the calculation methods, and a 3D volumetric dosemeter was found to be an adequate tool for delivery QA of DAT. (2) The MCC was superior to PBC in terms of the accuracy in dose calculation for DAT even in the homogenous condition.

Entities:  

Mesh:

Year:  2013        PMID: 24234583      PMCID: PMC3856546          DOI: 10.1259/bjr.20130353

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  37 in total

1.  A 2-D diode array and analysis software for verification of intensity modulated radiation therapy delivery.

Authors:  Paul A Jursinic; Ben E Nelms
Journal:  Med Phys       Date:  2003-05       Impact factor: 4.071

2.  Comparison of dose calculation algorithms with Monte Carlo methods for photon arcs.

Authors:  James C L Chow; Eugene Wong; Jeff Z Chen; Jake Van Dyk
Journal:  Med Phys       Date:  2003-10       Impact factor: 4.071

3.  Evaluation of a 2D detector array for patient-specific VMAT QA with different setups.

Authors:  Ramesh Boggula; Mattias Birkner; Frank Lohr; Volker Steil; Frederik Wenz; Hansjoerg Wertz
Journal:  Phys Med Biol       Date:  2011-10-25       Impact factor: 3.609

4.  Optimizing the accuracy of a helical diode array dosimeter: a comprehensive calibration methodology coupled with a novel virtual inclinometer.

Authors:  Jakub Kozelka; Joshua Robinson; Benjamin Nelms; Geoffrey Zhang; Dennis Savitskij; Vladimir Feygelman
Journal:  Med Phys       Date:  2011-09       Impact factor: 4.071

5.  An arc-sequencing algorithm for intensity modulated arc therapy.

Authors:  D M Shepard; D Cao; M K N Afghan; M A Earl
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

6.  Volumetric modulated arc therapy: IMRT in a single gantry arc.

Authors:  Karl Otto
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

7.  Variable dose rate single-arc IMAT delivered with a constant dose rate and variable angular spacing.

Authors:  Grace Tang; Matthew A Earl; Cedric X Yu
Journal:  Phys Med Biol       Date:  2009-10-09       Impact factor: 3.609

8.  Calibration of a novel four-dimensional diode array.

Authors:  Guanghua Yan; Bo Lu; Jakub Kozelka; Chihray Liu; Jonathan G Li
Journal:  Med Phys       Date:  2010-01       Impact factor: 4.071

9.  Evaluation of the trade-offs encountered in planning and treating locally advanced head and neck cancer: intensity-modulated radiation therapy vs dual-arc volumetric-modulated arc therapy.

Authors:  M Oliver; D McConnell; M Romani; A McAllister; A Pearce; A Andronowski; X Wang; K Leszczynski
Journal:  Br J Radiol       Date:  2012-07-17       Impact factor: 3.039

10.  Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array.

Authors:  Daniela Wagner; Hilke Vorwerk
Journal:  Radiat Oncol       Date:  2011-02-22       Impact factor: 3.481

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.