| Literature DB >> 24892330 |
Arjunan Manikandan1, Biplab Sarkar, Maitreyee Nandy, Chandra Sekaran Sureka, Michael S Gossman, Nadendla Sujatha, Vivek Thirupathur Rajendran.
Abstract
The aim of this study was to assess the performance of a gantry-mounted detector system and a couch set detector system using a systematic multileaf collimator positional error manually introduced for volumetric-modulated arc therapy. Four head and neck and esophagus VMAT plans were evaluated by measurement using an electronic portal imaging device and an ion chamber array. Each plan was copied and duplicated with a 1 mm systematic MLC positional error in the left leaf bank. Direct comparison of measurements for plans with and without the error permitted observational characteristics for quality assurance performance between detectors. A total of 48 different plans were evaluated for this testing. The mean percentage planar dose differences required to satisfy a 95% match between plans with and without the MLCPE were 5.2% ± 0.5% for the chamber array with gantry motion, 8.12% ± 1.04% for the chamber array with a static gantry at 0°, and 10.9%± 1.4% for the EPID with gantry motion. It was observed that the EPID was less accurate due to overresponse of the MLCPE in the left leaf bank. The EPID always images bank-A on the ipsilateral side of the detector, whereas for a chamber array or for a patient, that bank changes as it crosses the -90° or +90° position. A couch set detector system can reproduce the TPS calculated values most consistently. We recommend it as the most reliable patient specific QA system for MLC position error testing. This research is highlighted by the finding of up to 12.7% dose variation for H/N and esophagus cases for VMAT delivery, where the mere source of error was the stated clinically acceptability of 1 mm MLC position deviation of TG-142.Entities:
Mesh:
Year: 2014 PMID: 24892330 PMCID: PMC5711059 DOI: 10.1120/jacmp.v15i3.4495
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Ion chamber array and Octavius II phantom setup for VMAT dose delivery measurement.
Figure 2Isodose distribution for head and neck case using a single full arc VMAT plan. Treatment plan generated for 66 Gy in 33 fractions and 59.4 Gy in 33 fractions
Figure 3Plan testing results for eight VMAT plans: four esophagus plans (P1‐P4) and four H/N plans (P5‐P8) with and without a 1 mm MLC positioning error; measurements were obtained by the CS detector for arc therapy, as well as collapsed static gantry delivery; measurements were obtained by the GS detector for arc therapy only.
Cumulative plan testing results and statistics for eight VMAT plans: four esophagus plans (P1‐P4) and four H/N plans (P5‐P8) with and without a 1 mm MLC positioning error. Relative dose difference between MLC0PE vs. MLC1PE to achieve 95% matching rate. indicates measurements with a static gantry delivery; indicates arc therapy delivery
|
|
|
|
| |||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
|
|
|
|
| |||||
| Patient 1 | ‐ | 6.8 | ‐ | 5.1 | – | 10.0 | ‐ | 6.5 | ||||
| Patient 2 | ‐ | 7.8 | ‐ | 5.4 | ‐ | 10.5 | ‐ | 6.9 | ||||
| Patient 3 | ‐ | 8.4 | ‐ | 5.7 | ‐ | 9.8 | ‐ | 7.4 | ||||
| Patient 4 | ‐ | 7.6 | ‐ | 5.0 | ‐ | 9.8 | ‐ | 6.4 | ||||
| Patient 5 | 7.5 | ‐ | 5.6 | ‐ | 9.5 | ‐ | 5.1 | ‐ | ||||
| Patient 6 | 9.6 | ‐ | 5.4 | ‐ | 12.5 | ‐ | 5.2 | ‐ | ||||
| Patient 7 | 9.7 | ‐ | 4.8 | ‐ | 12.7 | ‐ | 5.2 | ‐ | ||||
| Patient 8 | 7.5 | ‐ | 4.3 | ‐ | 12.4 | ‐ | 4.6 | ‐ | ||||
| Mean(4P) | 8.6 | 7.7 | 5.0 | 5.3 | 11.8 | 10.0 | 5.1 | 6.8 | ||||
| SD(4P) | 1.2 | 0.7 | 0.6 | 0.3 | 1.5 | 0.3 | 0.3 | 0.5 | ||||
| Mean(8P) | 8.1 | 5.2 | 10.9 | 5.9 | ||||||||
| SD (8P) | 1.0 | 0.5 | 1.38 | 1.0 | ||||||||
.
Measured dose distributions are compared with and without leaf position error, using gamma analysis for couch stationary detector in rotational gantry measurements
|
|
| |
|---|---|---|
| Esophagus Patients | 61.6 | 78.2 |
| 87.0 | 97.7 | |
| 73.3 | 96.6 | |
| 72.5 | 94.4 | |
| Head and Neck | 53.1 | 81.8 |
| 64.4 | 92.8 | |
| 58.8 | 89.2 | |
| 53.0 | 81.5 |
DVH analysis for eight patients detailing the dose difference between plans with and without MLCPE
|
| |||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
|
|
|
| ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| PTV | Dose Coverage (D95%) | 246.0 | 251.9 | ‐2.4 | 179.2 | 182.6 | 1.9 | 203.8 | 208.4 | 5.7 | 183.9 | 189.8 | 3.2 |
| Hot Ref (D107%) | 10.3 | 42.2 | 12.4 | 4.1 | 20.7 | 8.9 | 5.5 | 30.1 | 11.4 | 0.8 | 3.8 | 10.5 | |
| RT PAROTID | MEAM (Gy) | 38.3 | 39.8 | 3.9 | 22.2 | 23.1 | 4.1 | 32.2 | 33.7 | 4.7 | 27.3 | 28.4 | 4.0 |
| LT PAROTID | MEAN (Gy) | 36.6 | 38.2 | 4.4 | 24.3 | 25.4 | 4.5 | 28.8 | 30.3 | 4.2 | 28.7 | 30.3 | 5.6 |
| SPINE | 1% volume | 42.3 | 43.9 | 3.8 | 40.4 | 42.1 | 4.2 | 40.2 | 41.7 | 3.7 | 41.4 | 43.2 | 4.4 |
| LARYNX | MEAN (Gy) | 42.9 | 44.6 | 4.0 | 43.8 | 45.3 | 3.4 | 53.6 | 58.5 | 9.1 | 42.1 | 43.6 | 3.6 |
|
| |||||||||||||
|
|
|
|
| ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| PTV | Dose Coverage (D95%) | 351.9 | 333.0 | 5.0 | 386.4 | 369.5 | 4.4 | 246.5 | 241.4 | 2.0 | 206.5 | 203.1 | 1.6 |
| Hot Ref (D107%) | 4.2 | 47.3 | 11.4 | 8.2 | 39.1 | 7.4 | 4.5 | 28.2 | 9.5 | 8.1 | 15.8 | 3.8 | |
| RT LUNG | MEAN | 27.5 | 26.8 | 2.7 | 15.6 | 15.1 | 3.5 | 12.1 | 11.8 | 2.5 | 20.8 | 20.1 | 3.8 |
| V20Gy | 66.3 | 64.3 | 3.1 | 25.6 | 24.1 | 6.1 | 8.9 | 8.0 | 10.7 | 37.7 | 35.1 | 7.5 | |
| LT LUNG | MEAN | 13.9 | 13.4 | 3.8 | 15.7 | 15.1 | 3.5 | 15.4 | 14.9 | 3.3 | 17.3 | 16.6 | 4.5 |
| V20Gy | 14.8 | 13.2 | 11.6 | 32.3 | 30.6 | 5.4 | 27.3 | 26.0 | 5.2 | 30.3 | 27.7 | 9.7 | |
| HEART | MEAN | 24.0 | 23.4 | 2.6 | 24.2 | 23.3 | 3.6 | 27.4 | 26.7 | 2.5 | 27.7 | 26.7 | 3.6 |
| V50Gy | 9.2 | 9.5 | 2.6 | 12.4 | 11.2 | 10.7 | 9.2 | 8.6 | 7.5 | 6.8 | 8.0 | 14.5 | |
Figure 4Energy fluence map results detailing the dose distribution percentage relative difference (MLC0PE‐MLC1PE) for two H/N plans and two esophagus plans with and without the 1 mm MLC error; (top) array and EPID results for arc rotation delivery; (bottom) array results for static gantry delivery.
Figure 5Schematic for geometrical relationship between the gantry, MLC, patient or CS detector, and GS detector for VMAT delivery; one MLC of left leaf bank is depicted here. At gantry angle 190° and 170° MLC from left bank projecting on the right side and left side, respectively, of the CS detector/patient. Same MLC projecting at the left side of the GS throughout the rotation. MLC direction changes for patient and CS detector when G0 crosses ( and 90°); for GS detector, this direction change is not applicable.