Literature DB >> 22701793

High-pitch computed tomography coronary angiography-a new dose-saving algorithm: estimation of radiation exposure.

Dominik Ketelsen1, Markus Buchgeister, Andreas Korn, Michael Fenchel, Bernhard Schmidt, Thomas G Flohr, Christoph Thomas, Christoph Schabel, Ilias Tsiflikas, Roland Syha, Claus D Claussen, Martin Heuschmid.   

Abstract

Purpose. To estimate effective dose and organ equivalent doses of prospective ECG-triggered high-pitch CTCA. Materials and Methods. For dose measurements, an Alderson-Rando phantom equipped with thermoluminescent dosimeters was used. The effective dose was calculated according to ICRP 103. Exposure was performed on a second-generation dual-source scanner (SOMATOM Definition Flash, Siemens Medical Solutions, Germany). The following scan parameters were used: 320 mAs per rotation, 100 and 120 kV, pitch 3.4 for prospectively ECG-triggered high-pitch CTCA, scan range of 13.5 cm, collimation 64 × 2 × 0.6 mm with z-flying focal spot, gantry rotation time 280 ms, and simulated heart rate of 60 beats per minute. Results. Depending on the applied tube potential, the effective whole-body dose of the cardiac scan ranged from 1.1 mSv to 1.6 mSv and from 1.2 to 1.8 mSv for males and females, respectively. The radiosensitive breast tissue in the range of the primary beam caused an increased female-specific effective dose of 8.6%±0.3% compared to males. Decreasing the tube potential, a significant reduction of the effective dose of 35.8% and 36.0% can be achieved for males and females, respectively (P < 0.001). Conclusion. The radiologist and the CT technician should be aware of this new dose-saving strategy to keep the radiation exposure as low as reasonablly achievable.

Entities:  

Year:  2012        PMID: 22701793      PMCID: PMC3371345          DOI: 10.1155/2012/724129

Source DB:  PubMed          Journal:  Radiol Res Pract        ISSN: 2090-195X


1. Introduction

At present, computed tomography coronary angiography (CTCA) is an important, widely accepted diagnostic tool for the assessment of coronary artery disease. Several studies have shown the potential of different dose-saving strategies to keep the radiation exposure as low as reasonablly achievable. Hausleiter et al. reported in an international multicenter trial (PROTECTION I) a mean effective dose of 12 mSv in CTCA, ranging from 5 to 30 mSv [1]. Radiation exposure can be reduced substantially by currently available strategies, but these possibilities are used infrequently [1]. Since the introduction of a second-generation dual-source scanner system, a new scanning mode with increased table feed is available. Compared to retrospective ECG-gated and prospective ECG-triggered CTCA, this high-pitch, prospective triggered scanning mode has the potential for drastic dose reduction due to a gapless imaging of the heart within one heartbeat with no overlapping data acquisition. The aim of the study was to estimate effective whole-body dose and organ equivalent doses of prospective ECG-triggered high-pitch CTCA.

2. Material and Methods

2.1. Dosimetry

The experiments were performed by using an anthropomorphic, hermaphrodite male phantom with breast phantom attachments (Alderson-Rando phantom; Alderson Research Laboratories Inc., Stanford, CT, USA). The phantom was equipped with 117 thermoluminescent dosimeters (TLDs) with dimension of 1 × 1 × 6 mm (TLD-100H, Bicon-Harshow, Radiation Measurement Products, Cleveland, OH, USA) to perform radiation exposure measurements [2-4]. By using a Philips Optimus 65 (Philips Medical Systems, PC Best, The Netherlands), a calibration of the 117 was calculated, which was defined by means of parallel exposure of 33 TLDs with a known radiation dose using 102 kV and 10 mA for 100 ms at a source-skin distance (SSD) of 100 cm. To minimize the Heel effect, wire markers in the field were avoided and all expositions were done in the same position with respect to the orientation of the X-ray tube. Crosschecked by an ionization dosimeter positioned in the same phantom depths, the reference TLDs were exposed with a dose of 1.081 mGy. No further correction factors were used as the calibration voltage is close to the CT tube voltage. The evaluation of the irradiated TLDs was performed using a TLD reader (Model 5500 TLD Reader, Bicron Radiation Measurement Products, Solon, OH, USA) within 24 h after radiation exposure. The readout TLD values in nanocoulombs were multiplied by an individual calibration factor. Dependent on the anatomical position of each organ 39 different positions in the Alderson-Rando Phantom were used to assess the organ doses. Due to measurement deviation, three TLDs were placed at each point of dose measurement to minimize bias. The number of TLDs allocated to different organ positions were as follows: 3 at the brain, thyroid gland, esophagus, thymus, heart, breast, stomach, upper colon, spleen, kidneys, adrenal glands, pancreas, small intestine, lower colon, urinary bladder, muscle tissue, red bone marrow, skin, ovaries, and testicles, 42 at the lung, and 15 at the liver. The effective dose was calculated by summarizing the weighted organ doses according to the guidelines of ICRP 103 [5]. Radiation doses of simulated small organs (i.e., thyroid gland) were directly rated into the calculation. Doses of larger organs (i.e., lung) were determined by assessing the mean of measured TLDs from the entire organ. To assess gender-specific differences, the testicles were used to measure the male-specific gonadal dose while radiation dose of the breast and the ovaries accounted to the female-specific radiation exposure. The breast phantom attachments could influence the calculation of the male-specific effective dose by increased soft tissue in the scan range. Nevertheless, evaluated values of effective dose using a hermaphrodite phantom are sufficiently precise for application in radiological radiation protection [5, 6].

2.2. Scan Protocols

Scans were performed on a second-generation dual-source CT scanner (SOMATOM Definition Flash, Siemens Medical Solutions, Forchheim, Germany). The following parameters were used: 320 mAs per rotation, 100 and 120 kV, pitch 3.4 for prospectively ECG-triggered CTCA (table feed 450 mm/s), scan range of 13.5 cm (four blocks of sequential detector coverage), collimation 64 × 2 × 0.6 mm with z-flying focal spot, and gantry rotation time 280 ms. The scanner software simulated a heart rate of 60 beats per minute. 60% of the RR-interval was set to trigger the scan. In high-pitch CTCA, the heart is scanned within one heartbeat. Data acquisition with two X-ray tubes and detectors allows a gapless imaging of the heart despite a high pitch of 3.4. Gaps in the data from the first measurement system resulting from the high pitch are filled completely with the data from the second measurement system that acquires the data a quarter rotation later [7].

2.3. Statistical Analysis

Normal distribution of measured organ-specific dose values of each scan protocol was assessed using the Kolmogorov-Smirnov test. Statistical significance between different scan protocols was evaluated using the two-tailed paired Student's t-test comparing the measured organ-specific dose values. A P value <0.05 was considered to be significant.

3. Results

Effective, gender-specific doses of scan protocols with 100 kV and 120 kV and reported CT dose index and dose length product are shown in Table 1.
Table 1

Effective radiation exposure of high-pitch CTCA using 100 kV and 120 kV.

ProtocolEffective dose male [mSv]Effective dose female [mSv]Total mAsCTDIVol (mGy)DLP (mGy cm)
1 (100 kV)1.11.26883.156
2 (120 kV)1.61.86895.295

CTDIVol: volume CT dose index; DLP: dose-length product.

Depending on the applied tube potential, the effective whole-body dose of the cardiac scan ranged from 1.1 mSv to 1.6 mSv and from 1.2 to 1.8 mSv for males and females, respectively. Directly irradiated organs within the scan range received organ equivalent doses of up to 0.33 mSv to the lung and up to 0.13 mSv to the breast. The radiosensitive breast tissue in the range of the primary beam caused an increased female-specific effective dose of 8.6% ± 0.3% compared to males. The gonadal doses were <0.02 mSv in both scan protocols. Details about organ equivalent doses are displayed in Table 2.
Table 2

Measured organ equivalent doses of high-pitch CTCA using different tube voltages (mSv).

Organ/tissue100 kV120 kV
Thyroid gland0.030.04
Esophagus0.050.05
Lung0.200.33
Breast0.070.13
Stomach0.450.67
Liver0.060.12
Colon0.030.03
Urinary bladder0.000.00
Red bone marrow0.010.02
Skeleton0.000.00
Skin0.000.01
Male gonads0.000.00
Female gonads0.010.01
Remaining organs0.220.37
A reduction of the tube potential from 120 kV to 100 kV significantly reduces the radiation exposure by 35.8% for males and 36.0% for females, respectively (P < 0.001).

4. Discussion

Several studies described the correlation between the pitch value and the radiation exposure in computed tomography coronary artery (CTCA). Typically, the pitch is less than 1, usually between 0.2 and 0.5 in first-generation dual-source CTCA resulting in relevant overlaps by advancing the table much less than one detector width during one scanner rotation [3, 7]. Thus, repeated exposure of the same heart region during several consecutive rotations causes an increase of radiation exposure. The evaluated, prospective ECG-triggered high-pitch CTCA uses a pitch value of 3.4 resulting in a table feed of 450 mm/s with no overlapping data acquisition by filling the gaps in the data of the spiral acquisition of the first measurement system with the data from the second measurement system one-quarter rotation later. The whole heart is completely scanned within one heart cycle, that is, the topmost image will be acquired at the set trigger point and subsequent images display the heart in later phases of the heart cycle [8-11]. In this study, we determined the effective radiation exposure of prospective ECG-triggered high-pitch CTCA depending on the gender and the used tube voltage. The results are concordant to recently reported dose values from Goetti et al. [8]. The radiation exposure is substantially lower than other reported dose values of retrospective ECG-gated or prospective ECG-triggered CTCA. A comparative study of different dose-saving techniques reports effective dose values of 5.8 mSv to 16.0 mSv for retrospective ECG-gated protocols without ECG-pulsing, and 4.2 mSv to 9.8 mSv for protocols with different ECG-pulsing techniques and 2.8 mSv to 4.3 mSv for prospective ECG-triggered, the so-called step-and-shoot, CTCA [12]. A further radiation protective effect comprises the acquisition of less overlapping data. This results in reduction of the organ equivalent dose of the extremely radiosensitive female breast tissue, which is always in the scan range, but rarely organ of interest. In high-pitch CTCA, the female-specific effective dose is higher compared to male effective dose measurements by an average of up to 8.6%. In conventional overlapping CTCA data acquisition, however, females received increased effective dose values of up to 70% [6, 12–14]. By decreasing the tube voltage, the effective dose can be further reduced by up to 36.0% in high-pitch CTCA. However, changes in tube voltage are complex and affect image noise as well as tissue contrast. Nevertheless, several studies confirmed a sufficient image quality of 100 kV CTCA protocols in slender patients with a body mass index lower than 30 kg/m2 [15]. A possible limitation of the study is the lack to determine image quality as an additional parameter of the technique. However, initial feasibility studies reported an excellent image quality in a selected patient population [9, 15]. High-pitch CTCA shows equivalent image quality in patients with low and stable heart rates up to 60 bpm compared to prospective ECG-triggered and retrospective ECG-gated CTCA. A diagnostic image quality can be achieved in 97% of coronary segments with high-pitch CTCA [16-18]. These studies recommend this prospective, R-wave triggered, high-pitch scanning mode for patients with low and regular heart rates (<55–60 beats per minute) to acquire the complete dataset before the onset of atrial contraction [7, 9, 10].

5. Conclusion

In conclusion, the prospective ECG-triggered high-pitch CTCA has the potential to scan the heart with an effective dose between 1 mSv and 2 mSv, which is substantially lower than previously described CTCA scanning modes. The radiologist and the CT technician should be aware of this new dose-saving strategy to keep the radiation exposure as low as reasonablly achievable.
  18 in total

1.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

2.  High-pitch spiral acquisition: a new scan mode for coronary CT angiography.

Authors:  Stephan Achenbach; Mohamed Marwan; Tiziano Schepis; Tobias Pflederer; Herbert Bruder; Thomas Allmendinger; Martin Petersilka; Katharina Anders; Michael Lell; Axel Kuettner; Dieter Ropers; Werner G Daniel; Thomas Flohr
Journal:  J Cardiovasc Comput Tomogr       Date:  2009-02-26

3.  Dual-source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: image reconstruction and assessment of image quality.

Authors:  Thomas G Flohr; Shuai Leng; Lifeng Yu; Thomas Aiimendinger; Herbert Bruder; Martin Petersilka; Christian D Eusemann; Karl Stierstorfer; Bernhard Schmidt; Cynthia H McCollough
Journal:  Med Phys       Date:  2009-12       Impact factor: 4.071

4.  Image quality and radiation dose comparison of prospectively triggered low-dose CCTA: 128-slice dual-source high-pitch spiral versus 64-slice single-source sequential acquisition.

Authors:  Jelena R Ghadri; Silke M Küest; Robert Goetti; Michael Fiechter; Aju P Pazhenkottil; Rene N Nkoulou; Felix P Kuhn; Carsten Pietsch; Patrick von Schulthess; Oliver Gaemperli; Christian Templin; Philipp A Kaufmann
Journal:  Int J Cardiovasc Imaging       Date:  2011-07-10       Impact factor: 2.357

5.  Predictors of image quality in high-pitch coronary CT angiography.

Authors:  Paul Stolzmann; Robert P Goetti; Pal Maurovich-Horvat; Udo Hoffmann; Thomas G Flohr; Sebastian Leschka; Hatem Alkadhi
Journal:  AJR Am J Roentgenol       Date:  2011-10       Impact factor: 3.959

6.  Image quality and radiation exposure using different low-dose scan protocols in dual-source CT coronary angiography: randomized study.

Authors:  Lisan A Neefjes; Anoeshka S Dharampal; Alexia Rossi; Koen Nieman; Annick C Weustink; Marcel L Dijkshoorn; Gert-Jan R Ten Kate; Admir Dedic; Stella L Papadopoulou; Marcel van Straten; Filippo Cademartiri; Gabriël P Krestin; Pim J de Feyter; Nico R Mollet
Journal:  Radiology       Date:  2011-10-03       Impact factor: 11.105

Review 7.  Pushing the envelope: new computed tomography techniques for cardiothoracic imaging.

Authors:  Thomas G Flohr; Ernst Klotz; Thomas Allmendinger; Rainer Raupach; Herbert Bruder; Bernhard Schmidt
Journal:  J Thorac Imaging       Date:  2010-05       Impact factor: 3.000

8.  Diagnostic accuracy of high-pitch dual-source CT for the assessment of coronary stenoses: first experience.

Authors:  Sebastian Leschka; Paul Stolzmann; Lotus Desbiolles; Stephan Baumueller; Robert Goetti; Thomas Schertler; Hans Scheffel; Andre Plass; Volkmar Falk; Gudrun Feuchtner; Borut Marincek; Hatem Alkadhi
Journal:  Eur Radiol       Date:  2009-09-16       Impact factor: 5.315

9.  Dual-source computed tomography: estimation of radiation exposure of ECG-gated and ECG-triggered coronary angiography.

Authors:  Dominik Ketelsen; Christoph Thomas; Matthias Werner; Marie H Luetkhoff; Markus Buchgeister; Ilias Tsiflikas; Anja Reimann; Christof Burgstahler; Harald Brodoefel; Andreas F Kopp; Claus D Claussen; Martin Heuschmid
Journal:  Eur J Radiol       Date:  2008-12-18       Impact factor: 3.528

10.  2007 recommendations of the ICRP change basis for estimation of the effective dose: what is the impact on radiation dose assessment of patient and personnel?

Authors:  H von Boetticher; J Lachmund; H K Looe; W Hoffmann; B Poppe
Journal:  Rofo       Date:  2008-05
View more
  6 in total

1.  Effects of a high-pitch protocol and a hybrid iterative reconstruction algorithm on image quality of cerebral subtracted 3D CT angiography.

Authors:  Yuji Iyama; Takeshi Nakaura; Masafumi Kidoh; Naoto Kiyota; Shouzaburou Uemura; Kazunori Harada; Yasuyuki Yamashita
Journal:  Jpn J Radiol       Date:  2015-09-19       Impact factor: 2.374

Review 2.  Stress echocardiography: what is new and how does it compare with myocardial perfusion imaging and other modalities?

Authors:  Marysia S Tweet; Adelaide M Arruda-Olson; Nandan S Anavekar; Patricia A Pellikka
Journal:  Curr Cardiol Rep       Date:  2015-06       Impact factor: 2.931

3.  Utility of cardiac CT for evaluating delayed contrast enhancement in dilated cardiomyopathy.

Authors:  V Cerny; P Kuchynka; J Marek; L Lambert; M Masek; T Palecek; D Ambroz; A Linhart; J Danes
Journal:  Herz       Date:  2016-12-15       Impact factor: 1.443

Review 4.  Role of CT in the Pre- and Postoperative Assessment of Conotruncal Anomalies.

Authors:  Parveen Kumar; Mona Bhatia
Journal:  Radiol Cardiothorac Imaging       Date:  2022-06-30

5.  Assessment of radiation safety in cardiac CT angiography.

Authors:  Yuan-Hao Lee; Po-Yen Chang; Shee Yen Tay; Cheng-Yu Tsai; Po-Hsin Cheng; Wilson T Lao; Wing P Chan
Journal:  Medicine (Baltimore)       Date:  2017-12       Impact factor: 1.817

6.  Image Quality and Dose Reduction by Dual Source Computed Tomography Coronary Angiography: Protocol Comparison.

Authors:  Ernesto Forte; Serena Monti; Chiara Anna Parente; Lukas Beyer; Roberto De Rosa; Teresa Infante; Carlo Cavaliere; Filippo Cademartiri; Marco Salvatore; Christian Stroszczynski; Carlo Tedeschi
Journal:  Dose Response       Date:  2018-10-16       Impact factor: 2.658