PURPOSE: Long-term studies are required to support the use of metal-on-metal (MoM) bearings in total hip arthroplasty (THA) given the concern about systemic metal ion release and reports of adverse local soft tissue reactions. The purpose of this study was to report the seven to 13-year clinical, radiographic, and metal ion results in patients following MoM THA.
METHODS: We studied 163 prostheses after second-generation MoM THA between July 1997 and November 2003. Cobalt and chromium metal ions were collected using whole and analysed by inductively-coupled plasma-mass spectrometry.
RESULTS: The mean follow-up was 8.87 years (range, 7-13 years). Four hips (2.5 %) were revised. The Kaplan-Meier survivorship was 91.3 % for revision for all causes, and 97.5 % when excluding the hips revised for a manufacturer's defect. Median whole blood cobalt levels peaked at a value of 2.87 μg/L at four years (p < 0.0001 vs. pre-operative) and subsequently decreased to 2.0 μg/L after nine years (p = 0.002 vs. four years). Median chromium levels maximally increased up to 0.75 μg/L after five years (p < 0.0001 vs. pre-operative) and tended to decrease thereafter to values of 0.56 μg/L after seven years.
CONCLUSIONS: This seven to 13-year follow-up study indicates that the clinical and radiological results following MoM THA are satisfactory with low revision rates. Cobalt and chromium ion levels peaked at four and five years, respectively, and gradually decreased thereafter.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22678122      PMCID: PMC3427452          DOI: 10.1007/s00264-012-1570-1

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  31 in total

1.  The Otto Aufranc Award. Wear and lubrication of metal-on-metal hip implants.

Authors:  F W Chan; J D Bobyn; J B Medley; J J Krygier; M Tanzer
Journal:  Clin Orthop Relat Res       Date:  1999-12       Impact factor: 4.176

2.  Induction of apoptosis and necrosis by metal ions in vitro.

Authors:  Olga L Huk; Isabelle Catelas; Fackson Mwale; John Antoniou; David J Zukor; Alain Petit
Journal:  J Arthroplasty       Date:  2004-12       Impact factor: 4.757

3.  A consensus paper on metal ions in metal-on-metal hip arthroplasties.

Authors:  Steven J MacDonald; Wolfram Brodner; Joshua J Jacobs
Journal:  J Arthroplasty       Date:  2004-12       Impact factor: 4.757

4.  Cup inclination and serum concentration of cobalt and chromium after metal-on-metal total hip arthroplasty.

Authors:  Wolfram Brodner; Alexander Grübl; Rafael Jankovsky; Vanee Meisinger; Stephan Lehr; Florian Gottsauner-Wolf
Journal:  J Arthroplasty       Date:  2004-12       Impact factor: 4.757

5.  Engineering issues and wear performance of metal on metal hip implants.

Authors:  F W Chan; J D Bobyn; J B Medley; J J Krygier; S Yue; M Tanzer
Journal:  Clin Orthop Relat Res       Date:  1996-12       Impact factor: 4.176

6.  Development of a routine method for the determination of trace metals in whole blood by magnetic sector inductively coupled plasma mass spectrometry with particular relevance to patients with total hip and knee arthroplasty.

Authors:  C P Case; L Ellis; J C Turner; B Fairman
Journal:  Clin Chem       Date:  2001-02       Impact factor: 8.327

7.  Metal-specific differences in levels of DNA damage caused by synovial fluid recovered at revision arthroplasty.

Authors:  A P Davies; A Sood; A C Lewis; R Newson; I D Learmonth; C P Case
Journal:  J Bone Joint Surg Br       Date:  2005-10

8.  EBRA: a method to measure migration of acetabular components.

Authors:  M Krismer; R Bauer; J Tschupik; P Mayrhofer
Journal:  J Biomech       Date:  1995-10       Impact factor: 2.712

9.  Metal-on-metal bearings and hypersensitivity in patients with artificial hip joints. A clinical and histomorphological study.

Authors:  Hans-Georg Willert; Gottfried H Buchhorn; Afshin Fayyazi; Renata Flury; Markus Windler; Georg Köster; Christoph H Lohmann
Journal:  J Bone Joint Surg Am       Date:  2005-01       Impact factor: 5.284

10.  Metal release in patients who have had a primary total hip arthroplasty. A prospective, controlled, longitudinal study.

Authors:  J J Jacobs; A K Skipor; L M Patterson; N J Hallab; W G Paprosky; J Black; J O Galante
Journal:  J Bone Joint Surg Am       Date:  1998-10       Impact factor: 5.284

View more
  19 in total

1.  Changes in serum chromium levels over 12 years after Metasul metal-on-metal total hip arthroplasty.

Authors:  Katsuhiko Maezawa; Masahiko Nozawa; Takahito Yuasa; Munehiko Sugimoto; Kazuo Kaneko
Journal:  J Orthop       Date:  2018-02-28

2.  Metal-on-metal hip prostheses: correlation between debris in the synovial fluid and levels of cobalt and chromium ions in the bloodstream.

Authors:  Dalila De Pasquale; Susanna Stea; Stefano Squarzoni; Barbara Bordini; Marilina Amabile; Simona Catalani; Pietro Apostoli; Aldo Toni
Journal:  Int Orthop       Date:  2014-03       Impact factor: 3.075

3.  Repeated metal ion measurements in patients with high risk metal-on-metal hip replacement.

Authors:  Aleksi Reito; Teemu Moilanen; Timo Puolakka; Jorma Pajamäki; Antti Eskelinen
Journal:  Int Orthop       Date:  2014-03-18       Impact factor: 3.075

Review 4.  [Total hip arthroplasty in young patients : Bearings and custom-made prostheses].

Authors:  C Benignus; M Morlock; J Beckmann
Journal:  Orthopade       Date:  2019-04       Impact factor: 1.087

5.  Minimum ten-year results of a 28-mm metal-on-metal bearing in cementless total hip arthroplasty in patients fifty years of age and younger.

Authors:  Moritz M Innmann; Tobias Gotterbarm; Jan Philippe Kretzer; Christian Merle; Volker Ewerbeck; Stefan Weiss; Peter R Aldinger; Marcus R Streit
Journal:  Int Orthop       Date:  2013-12-20       Impact factor: 3.075

6.  The Role of Metal-on-Metal Bearings in Total Hip Arthroplasty and Hip Resurfacing: Review Article.

Authors:  David Sands; Emil H Schemitsch
Journal:  HSS J       Date:  2016-08-25

7.  Adverse Reactions to Metal on Metal Are Not Exclusive to Large Heads in Total Hip Arthroplasty.

Authors:  Adolph V Lombardi; Keith R Berend; Joanne B Adams; Keri L Satterwhite
Journal:  Clin Orthop Relat Res       Date:  2016-02       Impact factor: 4.176

8.  What are the predictors and prevalence of pseudotumor and elevated metal ions after large-diameter metal-on-metal THA?

Authors:  Nick Bayley; Habeeb Khan; Paul Grosso; Thomas Hupel; David Stevens; Matthew Snider; Emil Schemitsch; Paul Kuzyk
Journal:  Clin Orthop Relat Res       Date:  2015-02       Impact factor: 4.176

9.  The effect of a manufacturer recall on the threshold to revise a metal-on-metal hip.

Authors:  Saket Tibrewal; Shiraz Sabah; Johann Henckel; Alister Hart
Journal:  Int Orthop       Date:  2014-05-15       Impact factor: 3.075

Review 10.  How have alternative bearings and modularity affected revision rates in total hip arthroplasty?

Authors:  William M Mihalko; Markus A Wimmer; Carol A Pacione; Michel P Laurent; Robert F Murphy; Carson Rider
Journal:  Clin Orthop Relat Res       Date:  2014-12       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.