BACKGROUND: Joint-replacement surgery has revolutionized the treatment of osteoarthritis and is still the most effective therapy. A recent clinical trend reintroducing metal-on-metal bearing surfaces has in turn stimulated a requirement for accurate measurement of the concentrations of relevant metals in both pre- and postoperative patients. Thus, there is a need for cost-effective, multielement methods for trace metal analysis in whole blood to monitor possible increases in wear metal concentrations.
METHODS: A method was developed to allow routine analysis of whole blood samples for molybdenum, cobalt, chromium, and nickel. Sample preparation consisted of a simple 1:10 dilution of whole blood with a solution of 10 mL/L Triton X-100, 0.0002 mol/L EDTA, and 0.01 mol/L ammonium hydroxide. Final determination was performed by a double-focusing magnetic sector inductively coupled plasma mass spectrometer operated in medium-resolution mode (resolution, 3400). Online addition of rhodium was used for internal standardization.
RESULTS: Detection limits in whole blood were 0.06 microg/L for chromium, cobalt, and molybdenum and 0.30 microg/L for nickel. Base concentrations of 0.22, 0.17, 0.62, and 0.99 microg/L for chromium, cobalt, molybdenum, and nickel, respectively, in whole blood have been found. Polyatomic interferences on all four elements have been shown to be resolved from the analyte masses by use of a resolution of >3000.
CONCLUSIONS: The simple, rapid method of sample preparation is effective in minimizing potential contamination and enables 60 samples (run time, 8 h) to be analyzed before cleaning the instrument is necessary. A resolution >3000 was sufficient to separate polyatomic interferences from the masses of interest. The method was used to analyze a large number of blood samples taken from primary patients awaiting total hip arthroplasty. The method is sensitive enough to provide base concentrations for chromium, cobalt, and molybdenum in whole blood. The results for nickel were compromised by high signals for blank samples.

Entities:  

Mesh:

Substances:

Year:  2001        PMID: 11159776

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  11 in total

1.  A critical exploration of blood and environmental chromium concentration among oral cancer patients in an oral cancer prevalent area of Taiwan.

Authors:  Chi-Ting Chiang; Tsun-Kuo Chang; Yaw-Huei Hwang; Che-Chun Su; Kuo-Yang Tsai; Tzu-Hsuen Yuan; Ie-Bin Lian
Journal:  Environ Geochem Health       Date:  2010-10-27       Impact factor: 4.609

2.  Surgical variables influence metal ion levels after hip resurfacing.

Authors:  Nicholas M Desy; Stephane G Bergeron; Alain Petit; Olga L Huk; John Antoniou
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

3.  Validation of a Metallomics Analysis of Placenta Tissue by Inductively-Coupled Plasma Mass Spectrometry.

Authors:  James M Harrington; Daniel J Young; Rebecca C Fry; Frank X Weber; Susan S Sumner; Keith E Levine
Journal:  Biol Trace Elem Res       Date:  2015-07-09       Impact factor: 3.738

4.  Cadmium, chromium, lead, manganese and nickel concentrations in blood of women in non-polluted areas in Japan, as determined by inductively coupled plasma-sector field-mass spectrometry.

Authors:  Masayuki Ikeda; Fumiko Ohashi; Yoshinari Fukui; Sonoko Sakuragi; Jiro Moriguchi
Journal:  Int Arch Occup Environ Health       Date:  2010-06-03       Impact factor: 3.015

5.  Femoral head size does not affect ion values in metal-on-metal total hips.

Authors:  Mitchell Bernstein; Alan Walsh; Alain Petit; David J Zukor; Olga L Huk; John Antoniou
Journal:  Clin Orthop Relat Res       Date:  2011-06       Impact factor: 4.176

6.  Long-term follow-up and metal ion trend of patients with metal-on-metal total hip arthroplasty.

Authors:  Mitchell Bernstein; Nicholas M Desy; Alain Petit; David J Zukor; Olga L Huk; John Antoniou
Journal:  Int Orthop       Date:  2012-06-09       Impact factor: 3.075

7.  Analytical method for total chromium and nickel in urine using an inductively coupled plasma-universal cell technology-mass spectrometer (ICP-UCT-MS) in kinetic energy discrimination (KED) mode.

Authors:  C Derrick Quarles; Deanna R Jones; Jeffery M Jarrett; Gulchekhra Shakirova; Yi Pan; Kathleen L Caldwell; Robert L Jones
Journal:  J Anal At Spectrom       Date:  2014       Impact factor: 4.023

8.  Analysis of human serum and whole blood for mineral content by ICP-MS and ICP-OES: development of a mineralomics method.

Authors:  James M Harrington; Daniel J Young; Amal S Essader; Susan J Sumner; Keith E Levine
Journal:  Biol Trace Elem Res       Date:  2014-06-11       Impact factor: 3.738

Review 9.  Clinical usefulness of blood metal measurements to assess the failure of metal-on-metal hip implants.

Authors:  Barry Sampson; Alister Hart
Journal:  Ann Clin Biochem       Date:  2011-12-07       Impact factor: 2.057

10.  Molybdenum exposure and semen quality: how robust is the evidence of an effect?

Authors:  Tom Sorahan; Frank M Sullivan
Journal:  Environ Health Perspect       Date:  2009-09       Impact factor: 9.031

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.