| Literature DB >> 22639296 |
Z Jamil1, G Tearney, N Bruining, K Sihan, G van Soest, J Ligthart, R van Domburg, B Bouma, E Regar.
Abstract
Recently, Fourier domain OCT (FD-OCT) has been introduced for clinical use. This approach allows in vivo, high resolution (15 micron) imaging with very fast data acquisition, however, it requires brief flushing of the lumen during imaging. The reproducibility of such fast data acquisition under intracoronary flush application is poorly understood. To assess the inter-study variability of FD-OCT and to compare lumen morphometry to the established invasive imaging method, IVUS. 18 consecutive patients with coronary artery disease scheduled for PCI were included. In each target vessel a FD-OCT pullback (MGH system, light source 1,310 nm, 105 fps, pullback speed 20 mm/s) was acquired during brief (3 s) injection of X-ray contrast (flow 3 ml/s) through the guiding catheter. A second pullback was repeated under the same conditions after re-introduction of the FD OCT catheter into the coronary artery. IVUS and OCT imaging was performed in random order. FD-OCT and IVUS pullback data were analyzed using a recently developed software employing semi automated lumen contour and stent strut detection algorithms. Corresponding ROI were matched based on anatomical landmarks such as side branches and/or stent edges. Inter-study variability is presented as the absolute difference between the two pullbacks. FD-OCT showed remarkably good reproducibility. Inter-study variability in native vessels (cohort A) was very low for mean and minimal luminal area (0.10 ± 0.38, 0.19 ± 0.57 mm(2), respectively). Likewise inter-study variability was very low in stented coronary segments (cohort B) for mean lumen, mean stent, minimal luminal and minimal stent area (0.06 ± 0.08, 0.07 ± 0.10, 0.04 ± 0.09, 0.04 ± 0.10 mm(2), respectively). Comparison to IVUS morphometry revealed no significant differences. The differences between both imaging methods, OCT and IVUS, were very low for mean lumen, mean stent, minimal luminal and minimal stent area (0.10 ± 0.45, 0.10 ± 0.36, 0.26 ± 0.54, 0.05 ± 0.47 mm(2), respectively). FD-OCT shows excellent reproducibility and very low inter-study variability in both, native and stented coronary segments. No significant differences in quantitative lumen morphometry were observed between FD-OCT and IVUS. Evaluating these results suggest that FD-OCT is a reliable imaging tool to apply in longitudinal coronary artery disease studies.Entities:
Mesh:
Substances:
Year: 2012 PMID: 22639296 PMCID: PMC3550705 DOI: 10.1007/s10554-012-0067-8
Source DB: PubMed Journal: Int J Cardiovasc Imaging ISSN: 1569-5794 Impact factor: 2.357
Fig. 1Corresponding cross sectional (1) longitudinal (2) FD-OCT (A, B) and IVUS (C) images. The yellow lines on 2 indicate the region of interest. The blue lines indicate stent area and the red lines indicate lumen area
Fig. 2A1 and A2 show cross sectional images of two different pullbacks where the z-offset is slightly different. A1 has a correct z-offset and A2 shows a image of a pullback where the z-offset needs to be corrected. B Strut malapposition was defined as presence of at least one strut separated from the vessel wall (not being a side branch), if the distance between the endoluminal reflection of the strut and the vessel wall was larger than the thickness of the stent strut and polymer. The arrow in B shows a side branch. The blue line indicate stent area and the red line indicate lumen area. C Example of tissue protrusion: defined as convex shaped tissue between the stent struts without disruption of the continuity of the endoluminal vessel surface strut. The arrow in C shows two areas of tissue protrusion
Definitions
| Lumen area (mm2) | The surface of the lumen clearly visualized during flush administration as dark, signal poor region delineated by the most inner, signal intense endoluminal leading edge |
| Mean lumen area (mm2) | The mean of lumen areas of all frames in the selected ROI |
| Minimal lumen area (mm2) (MLA) | The smallest lumen area in the selected ROI |
| Relocation of minimal luminal area (MLA) | In those cases which the longitudinal position of MLA of first, test pullback or OCT was located in a different position in the second, retest pullback or IVUS. For the comparison of the position of the MLA between the two repeated pullbacks, the distance of the frame with the MLA from the distal starting point of the ROI was noted in the test series. Then, the position of the frame with the MLA was expressed as length percentage of the total ROI. The same procedure was performed in the re-test series and IVUS pullbacks |
| Stent struts | (1) Highly reflective surfaces (metal) that cast dorsal, radial shadows (2) Highly reflective surfaces without dorsal shadowing (3) Sector shaped shadows with sharp defined borders radial to the lumen |
| Start and end of stent | The first and the last frame with circumferentially visible struts |
| Stent area (mm2) | The surface of the stent by tracing individual stent struts assuming a circular shaped lumen |
| Mean stent area (mm2) | The mean of stent areas of all frames in the selected ROI |
| Minimal stent area (mm2) (MSA) | The smallest stent area in the selected ROI |
| Relocation of minimal stent area (MSA) | Was assessed as described for relocation of minimal lumen area |
| Tissue protrusion | Convex shaped tissue between the stent struts without disruption of the continuity of the endoluminal vessel surface strut |
| Strut malapposition | Presence of at least one strut separated from the vessel wall (not being a side branch), if the distance between the endoluminal reflection of the strut and the vessel wall was larger than the thickness of the stent strut and polymer [ |
Fig. 3A Per frame (n = 1,472) analyses in native coronary arteries. Bland–Altman plot (A′) showing the differences in lumen areas between two corresponding pullbacks. Regression analyses line (A″) showing correlation between corresponding lumen areas per frame. B Per frame (n = 3,520) analyses in stented coronary arteries. Bland–Altman plots showing the differences in lumen areas (B′) and stent areas (C′) between two corresponding pullbacks in stented coronary arteries. Regression analyses lines showing correlation between corresponding lumen areas (B″) and stent areas (C″) per frame
Inter-study variability in native coronary segments (per segment analysis)
| Mean lumen area | Minimal luminal area | |
|---|---|---|
| OCT pb1 | 6.74 ± (3.99) mm2 | 3.24 ± (3.85) mm2 |
| OCT pb2 | 6.64 ± (3.73) mm2 | 3.05 ± (3.30) mm2 |
| Pullback 1 versus pullback 2 | ||
| Absolute difference | 0.10 ± (0.38) mm2 | 0.19 ± (0.57) mm2 |
| Relative difference | 1.55 % | 6.11 % |
| Linear regression | ||
| Slope | 1.07 | 1.16 |
| Intercept | −0.35 | −0.31 |
| R2 | 0.99 | 0.99 |
| | <0.001 | <0.001 |
Fig. 4Comparison of the longitudinal position of minimal luminal area (MLA) in native coronary segments between two corresponding OCT pullbacks. X-axis: length percentage of the total ROI/longitudinal position. Y-axis: corresponding pullbacks
mean lumen area, mean stent area, minimal luminal area and minimal stent area in stented coronary segments in mm2
| Mean lumen area (mm2) | Mean stent area (mm2) | Minimal luminal area (mm2) | Minimal stent area (mm2) | |
|---|---|---|---|---|
| OCT pullback 1 | 7.18 ± (1.37) | 7.73 ± (1.09) | 5.32 ± (1.30) | 6.12 ± (0.85) |
| OCT pullback 2 | 7.12 ± (1.40) | 7.66 ± (1.07) | 5.36 ± (1.35) | 6.08 ± (0.88) |
| Pullback 1 versus pullback 2 | ||||
| Absolute difference | 0.06 ± (0.08) | 0.07 ± (0.10) | 0.04 ± (0.09) | 0.04 ± (0.10) |
| Relative difference | 0.81 % | 0.93 % | 0.68 % | 0.66 % |
| Linear regression | ||||
| Slope | 0.98 | 1.01 | 0.96 | 0.97 |
| Intercept | 0.20 | −0.03 | 0.16 | 0.24 |
| R2 | 0.99 | 0.99 | 0.99 | 0.99 |
| | <0.001 | <0.001 | <0.001 | <0.001 |
Fig. 5Comparison of the position of minimal luminal area (MLA) (A′, B′) and minimal stent are (MSA) (A″, B″) in stented coronary segments between two corresponding OCT pullbacks (A′, A″) and between OCT and IVUS (B′, B″). X-axis: length percentage of the total ROI/longitudinal position. Y-axis: corresponding pullbacks
OCT comparison to IVUS (per segment analysis)
| Mean lumen area (mm2) | Mean stent area (mm2) | Minimal lumen area (mm2) | Minimal stent area (mm2) | |
|---|---|---|---|---|
| OCT pullback | 6.24 ± (1.04) | 6.84 ± (1.06) | 4.59 ± (1.05) | 5.31 ± (0.78) |
| IVUS pullback | 6.34 ± (1.18) | 6.74 ± (1.30) | 4.84 ± (1.05) | 5.35 ± (1.04) |
| Pullback 1 versus pullback 2 | ||||
| Absolute difference | 0.10 ± (0.45) | 0.10 ± (0.36) | 0.26 ± (0.54) | 0.05 ± (0.47) |
| Relative difference | 1.6 % | 1.5 % | 5.4 % | 0.9 % |
| Linear regression | ||||
| Slope | 0.82 | 0.80 | 0.86 | 0.68 |
| Intercept | 1.05 | 1.47 | 0.40 | 1.68 |
| R2 | 0.86 | 0.94 | 0.75 | 0.82 |
| | <0.001 | <0.001 | =0.001 | <0.001 |
Fig. 6Bland Altman plots showing the differences in mean lumen areas (A′) and mean stent areas (B′) between OCT and IVUS in stented coronary arteries. Regression analyses lines showing correlation of mean lumen areas (A″) and mean stent areas (B″) between OCT and IVUS in stented coronary arteries