Intracoronary ultrasound imaging: intraobserver and interobserver variability of morphometric measurements.

D Hausmann1, A J Lundkvist, G J Friedrich, W L Mullen, P J Fitzgerald, P G Yock.   

Abstract

Measurements of lumen and plaque dimensions by intracoronary ultrasound imaging are useful in assessing effects of intracoronary interventions and in quantifying plaque burden in transplant patients or during regression trials. However, these measurements are affected by inter- and intraobserver variability. In 87 patients, 120 intracoronary ultrasound images were obtained with a 4.3F, 30 MHz catheter. Morphometric measurements were performed two times by three independent observers using computerized planimetry. Intraobserver and interobserver agreement for qualitative parameters (presence of atherosclerotic plaque, calcified plaque, and residual nondiseased wall) was high (> 87%). For quantitative parameters measured directly in the images (lumen area, minimal and maximal lumen diameters, area within the internal elastic lamina, arc of calcium plaque) interobserver and intraobserver correlation between measurements was high (correlation coefficient r > 0.90) and differences between measurements were low (mean differences < 10%; SD < 20%). Measurement of the arc of nondiseased wall showed less interobserver correlation (r = 0.76 to 0.91), but percentages of difference between the measurements were low. Parameters derived from directly measured variables (plaque area, area stenosis, thickness, and eccentricity) showed slightly higher variability (correlations between measurements r = 0.78 to 0.91). SD for percentages of difference ranged between 20% and 30% (plaque area, area stenosis, and thickness) and systematic deviation between measurements (mean differences > 10%) occurred for plaque area. Thus most qualitative and quantitative measurements of lumen and plaque dimensions performed in intracoronary ultrasound images have low in intraobserver and interobserver variability; derived parameters may have slightly higher variability. Variability of morphometric measurements has to be considered, especially when serial ultrasound measurements are compared.

Entities:  

Mesh:

Year:  1994        PMID: 7942437     DOI: 10.1016/0002-8703(94)90263-1

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  8 in total

1.  Temporal changes of coronary artery plaque located behind the struts of the everolimus eluting bioresorbable vascular scaffold.

Authors:  Salvatore Brugaletta; Hector M Garcia-Garcia; Scot Garg; Josep Gomez-Lara; Roberto Diletti; Yoshinobu Onuma; Robert Jan van Geuns; Dougal McClean; Dariusz Dudek; Leif Thuesen; Bernard Chevalier; Stephan Windecker; Robert Whitbourn; Cecile Dorange; Karine Miquel-Hebert; Krishnankutty Sudhir; John A Ormiston; Patrick W Serruys
Journal:  Int J Cardiovasc Imaging       Date:  2010-10-13       Impact factor: 2.357

2.  Multi-laboratory inter-institute reproducibility study of IVOCT and IVUS assessments using published consensus document definitions.

Authors:  Edouard Gerbaud; Giora Weisz; Atsushi Tanaka; Manabu Kashiwagi; Takehisa Shimizu; Lin Wang; Christiano Souza; Brett E Bouma; Melissa J Suter; Milen Shishkov; Giovanni J Ughi; Elkan F Halpern; Mireille Rosenberg; Sergio Waxman; Jeffrey W Moses; Gary S Mintz; Akiko Maehara; Guillermo J Tearney
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2015-09-15       Impact factor: 6.875

3.  In vivo variability in quantitative coronary ultrasound and tissue characterization measurements with mechanical and phased-array catheters.

Authors:  Gastón A Rodriguez-Granillo; Eugène P McFadden; Jiro Aoki; Carlos A G van Mieghem; Evelyn Regar; Nico Bruining; Patrick W Serruys
Journal:  Int J Cardiovasc Imaging       Date:  2005-12-14       Impact factor: 2.357

Review 4.  [Methods for coronary functional assessment].

Authors:  M Elsner
Journal:  Herz       Date:  1998-03       Impact factor: 1.443

5.  Qualitative and quantitative accuracy of ultrasound-based virtual histology for detection of necrotic core in human coronary arteries.

Authors:  Salvatore Brugaletta; Clarissa Cola; Victoria Martin-Yuste; Gemma Vilahur; Juan Oriol; Teresa Padro; Jose M Guerra; Roger Borras; Lina Badimon; Manel Sabate
Journal:  Int J Cardiovasc Imaging       Date:  2014-01-20       Impact factor: 2.357

Review 6.  Intracoronary ultrasound: current state of the art.

Authors:  P P Kearney; I R Starkey; G R Sutherland
Journal:  Br Heart J       Date:  1995-05

7.  Analysis of 1 year virtual histology changes in coronary plaque located behind the struts of the everolimus eluting bioresorbable vascular scaffold.

Authors:  Salvatore Brugaletta; Josep Gomez-Lara; Hector M Garcia-Garcia; Jung Ho Heo; Vasim Farooq; Robert J van Geuns; Bernard Chevalier; Stephan Windecker; Dougal McClean; Leif Thuesen; Robert Whitbourn; Ian Meredith; Cecile Dorange; Susan Veldhof; Richard Rapoza; John A Ormiston; Patrick W Serruys
Journal:  Int J Cardiovasc Imaging       Date:  2011-11-23       Impact factor: 2.357

8.  Interstudy reproducibility of the second generation, Fourier domain optical coherence tomography in patients with coronary artery disease and comparison with intravascular ultrasound: a study applying automated contour detection.

Authors:  Z Jamil; G Tearney; N Bruining; K Sihan; G van Soest; J Ligthart; R van Domburg; B Bouma; E Regar
Journal:  Int J Cardiovasc Imaging       Date:  2012-05-26       Impact factor: 2.357

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.