In vitro validation of new Fourier-domain optical coherence tomography.

Satoko Tahara1, Hiram G Bezerra, Motaz Baibars, Hiroyuki Kyono, Wei Wang, Stan Pokras, Emile Mehanna, Christopher L Petersen, Marco A Costa.   

Abstract

AIMS: To validate the accuracy and evaluate the intra- and inter-observer variability of vascular measurements using novel Fourier-domain optical coherence tomography (FD-OCT) versus intravascular ultrasound (IVUS) in a coronary phantom model. METHODS AND
RESULTS: A dedicated cylindrical phantom comprised of four sections with differing dimensions was used to compare images generated by five FD-OCT and five IVUS catheters. Each FD-OCT catheter was tested using three different consoles, generating 15 imaging pullbacks. Two independent experts, blinded to the phantom data, performed measurements. There were 180 FD-OCT and 60 IVUS cross-sectional measurements of mean lumen diameter (MeanLD), minimal lumen diameter (MLD), maximum lumen diameter (MaxLD) and lumen area (LA) at three points in each section. FD-OCT measurements had excellent correlations with IVUS (concordance correlation coefficient [CCC]: ≥0.9769 for MeanLD, LA or longitudinal length [LL]) and phantom dimensions (CCCs for FD-OCT: ≥0.9958 for MeanLD, LA or LL). FD-OCT measurements were larger than IVUS (p<0.0001), but showed less measurement errors compared to IVUS (p<0.0001). Moreover, FD-OCT caused less discrepancy between MaxLD and MLD versus IVUS (p<0.0001). Intra- and inter-observer variability was low for both FD-OCT (CCCs for MeanLD, LA and LL ≥0.9996) and IVUS (≥0.9935). Image catheter position did not influence FD-OCT measurements.
CONCLUSIONS: FD-OCT was more accurate than IVUS and had similar high reproducibility to determine vascular dimensions in vitro. These results support the use of FD-OCT in the clinical setting.

Entities:  

Mesh:

Year:  2011        PMID: 21252023     DOI: 10.4244/EIJV6I7A149

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  16 in total

1.  Fully automated side branch detection in intravascular optical coherence tomography pullback runs.

Authors:  Ancong Wang; Jeroen Eggermont; Johan H C Reiber; Jouke Dijkstra
Journal:  Biomed Opt Express       Date:  2014-08-25       Impact factor: 3.732

2.  New insight to estimate under-expansion after stent implantation on bifurcation lesions using optical coherence tomography.

Authors:  Daisuke Nakamura; Guilherme F Attizzani; Setsu Nishino; Kentaro Tanaka; Mohamad Soud; Gabriel T Pereira; Milana Leygerman; Anas Fares; Audrey Schnell; Marco A Costa; Andrejs Erglis; Hiram G Bezerra
Journal:  Int J Cardiovasc Imaging       Date:  2017-06-08       Impact factor: 2.357

3.  Intravascular frequency-domain optical coherence tomography assessment of atherosclerosis and stent-vessel interactions in human carotid arteries.

Authors:  M R Jones; G F Attizzani; C A Given; W H Brooks; M A Costa; H G Bezerra
Journal:  AJNR Am J Neuroradiol       Date:  2012-03-15       Impact factor: 3.825

4.  Variations in optical coherence tomography resolution and uniformity: a multi-system performance comparison.

Authors:  Anthony Fouad; T Joshua Pfefer; Chao-Wei Chen; Wei Gong; Anant Agrawal; Peter H Tomlins; Peter D Woolliams; Rebekah A Drezek; Yu Chen
Journal:  Biomed Opt Express       Date:  2014-06-09       Impact factor: 3.732

5.  Compliance charts to guide non-complex small artery stenting: validation by quantitative coronary angiography.

Authors:  Yohan Chacko; J Kimberly Haladyn; Debbie M Smith; Richard Lim
Journal:  Heart Asia       Date:  2013-05-31

Review 6.  Bioresorbable Scaffolds for Coronary Stenosis: When and How Based Upon Current Studies.

Authors:  Alexandre Abizaid; J Ribamar Costa
Journal:  Curr Cardiol Rep       Date:  2017-03       Impact factor: 2.931

7.  Utilization of frequency domain optical coherence tomography and fractional flow reserve to assess intermediate coronary artery stenoses: conciliating anatomic and physiologic information.

Authors:  Gregory T Stefano; Hiram G Bezerra; Guilherme Attizzani; Daniel Chamié; Emile Mehanna; Hirosada Yamamoto; Marco A Costa
Journal:  Int J Cardiovasc Imaging       Date:  2011-03-17       Impact factor: 2.357

Review 8.  Assessment of coronary stent by optical coherence tomography, methodology and definitions.

Authors:  Emile Aziz Mehanna; Guilherme Ferragut Attizzani; Hiroyuki Kyono; Michael Hake; Hiram Grando Bezerra
Journal:  Int J Cardiovasc Imaging       Date:  2011-02-19       Impact factor: 2.357

9.  Developing a New Dimension for Fourier Domain Optical Coherence Tomography Images by Simultaneous Measurement of the Refractive Index and Thickness.

Authors:  Ahmad Amjadi; Hamed Ghodsi; Salile Khandani; Bahareh Khishkhah; Payman Rajai; Mohammadreza Razzaghi
Journal:  J Lasers Med Sci       Date:  2021-12-29

10.  Comparison of longitudinal geometric measurement in human coronary arteries between frequency-domain optical coherence tomography and intravascular ultrasound.

Authors:  Yong Liu; Kunihiro Shimamura; Takashi Kubo; Atsuhi Tanaka; Hironori Kitabata; Yasushi Ino; Takashi Tanimoto; Yasutsugu Shiono; Makoto Orii; Takashi Yamano; Tomoyuki Yamaguchi; Kumiko Hirata; Toshio Imanishi; Takashi Akasaka
Journal:  Int J Cardiovasc Imaging       Date:  2013-11-24       Impact factor: 2.357

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.