Literature DB >> 22559603

A multichannel, real-time MRI RF power monitor for independent SAR determination.

Abdel-Monem M El-Sharkawy1, Di Qian, Paul A Bottomley, William A Edelstein.   

Abstract

PURPOSE: Accurate measurements of the RF power delivered during clinical MRI are essential for safety and regulatory compliance, avoiding inappropriate restrictions on clinical MRI sequences, and for testing the MRI safety of peripheral and interventional devices at known RF exposure levels. The goal is to make independent RF power measurements to test the accuracy of scanner-reported specific absorption rate (SAR) over the extraordinary range of operating conditions routinely encountered in MRI.
METHODS: A six channel, high dynamic range, real-time power profiling system was designed and built for monitoring power delivery during MRI up to 440 MHz. The system was calibrated and used in two 3 T scanners to measure power applied to human subjects during MRI scans. The results were compared with the scanner-reported SAR.
RESULTS: The new power measurement system has highly linear performance over a 90 dB dynamic range and a wide range of MRI duty cycles. It has about 0.1 dB insertion loss that does not interfere with scanner operation. The measurements of whole-body SAR in volunteers showed that scanner-reported SAR was significantly overestimated by up to about 2.2 fold.
CONCLUSIONS: The new power monitor system can accurately and independently measure RF power deposition over the wide range of conditions routinely encountered during MRI. Scanner-reported SAR values are not appropriate for setting exposure limits during device or pulse sequence testing.

Entities:  

Mesh:

Year:  2012        PMID: 22559603      PMCID: PMC3338589          DOI: 10.1118/1.3700169

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  33 in total

1.  Specific absorption rate as a poor indicator of magnetic resonance-related implant heating.

Authors:  Wolfgang R Nitz; Gerd Brinker; Dirk Diehl; Georg Frese
Journal:  Invest Radiol       Date:  2005-12       Impact factor: 6.016

2.  Homogeneous tissue model estimates of RF power deposition in human NMR studies. Local elevations predicted in surface coil decoupling.

Authors:  P A Bottomley; P B Roemer
Journal:  Ann N Y Acad Sci       Date:  1992-03-31       Impact factor: 5.691

3.  Comments on MR heating tests of critical implants.

Authors:  Frank G Shellock
Journal:  J Magn Reson Imaging       Date:  2007-11       Impact factor: 4.813

4.  SAR and temperature: simulations and comparison to regulatory limits for MRI.

Authors:  Zhangwei Wang; James C Lin; Weihua Mao; Wanzhan Liu; Michael B Smith; Christopher M Collins
Journal:  J Magn Reson Imaging       Date:  2007-08       Impact factor: 4.813

5.  Dependence of RF heating on SAR and implant position in a 1.5T MR system.

Authors:  Hiroyuki Muranaka; Takayoshi Horiguchi; Shuji Usui; Yoshitake Ueda; Osamu Nakamura; Fumiaki Ikeda
Journal:  Magn Reson Med Sci       Date:  2007       Impact factor: 2.471

6.  Sampling and evaluation of specific absorption rates during patient examinations performed on 1.5-Tesla MR systems.

Authors:  G Brix; M Reinl; G Brinker
Journal:  Magn Reson Imaging       Date:  2001-07       Impact factor: 2.546

Review 7.  Turning up the heat on MRI.

Authors:  Paul A Bottomley
Journal:  J Am Coll Radiol       Date:  2008-07       Impact factor: 5.532

8.  Experimental and numerical assessment of MRI-induced temperature change and SAR distributions in phantoms and in vivo.

Authors:  Sukhoon Oh; Andrew G Webb; Thomas Neuberger; BuSik Park; Christopher M Collins
Journal:  Magn Reson Med       Date:  2010-01       Impact factor: 4.668

9.  Simultaneous B(0)- and B(1)+-map acquisition for fast localized shim, frequency, and RF power determination in the heart at 3 T.

Authors:  Michael Schär; Evert-Jan Vonken; Matthias Stuber
Journal:  Magn Reson Med       Date:  2010-02       Impact factor: 4.668

10.  Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating.

Authors:  Kenneth B Baker; Jean A Tkach; John A Nyenhuis; Michael Phillips; Frank G Shellock; Jorge Gonzalez-Martinez; Ali R Rezai
Journal:  J Magn Reson Imaging       Date:  2004-08       Impact factor: 4.813

View more
  16 in total

1.  A whole-body RF dosimeter for independent SAR measurement in MR scanners.

Authors:  Di Qian; AbdEl-Monem M El-Sharkawy; Paul A Bottomley; William A Edelstein
Journal:  Proc Int Soc Magn Reson Med Sci Meet Exhib Int Soc Magn Reson Med Sci Meet Exhib       Date:  2013

2.  Parallel transmit excitation at 1.5 T based on the minimization of a driving function for device heating.

Authors:  N Gudino; M Sonmez; Z Yao; T Baig; S Nielles-Vallespin; A Z Faranesh; R J Lederman; M Martens; R S Balaban; M S Hansen; M A Griswold
Journal:  Med Phys       Date:  2015-01       Impact factor: 4.071

Review 3.  Sodium MRI in human heart: a review.

Authors:  Paul A Bottomley
Journal:  NMR Biomed       Date:  2015-02-12       Impact factor: 4.044

4.  Measurements of RF power reflected and radiated by multichannel transmit MR coils at 7T.

Authors:  Gerd Weidemann; Frank Seifert; Werner Hoffmann; Harald Pfeiffer; Reiner Seemann; Bernd Ittermann
Journal:  MAGMA       Date:  2016-04-02       Impact factor: 2.310

5.  Evaluation of epidural and peripheral nerve catheter heating during magnetic resonance imaging.

Authors:  Sean Owens; M Arcan Erturk; Jean-Pierre P Ouanes; Jamie D Murphy; Christopher L Wu; Paul A Bottomley
Journal:  Reg Anesth Pain Med       Date:  2014 Nov-Dec       Impact factor: 6.288

6.  An RF dosimeter for independent SAR measurement in MRI scanners.

Authors:  Di Qian; Abdel-Monem M El-Sharkawy; Paul A Bottomley; William A Edelstein
Journal:  Med Phys       Date:  2013-12       Impact factor: 4.071

7.  Optically controlled on-coil amplifier with RF monitoring feedback.

Authors:  Natalia Gudino; Jacco A de Zwart; Qi Duan; Stephen J Dodd; Joe Murphy-Boesch; Peter van Gelderen; Jeff H Duyn
Journal:  Magn Reson Med       Date:  2017-09-14       Impact factor: 4.668

Review 8.  Magnetic resonance safety.

Authors:  Steffen Sammet
Journal:  Abdom Radiol (NY)       Date:  2016-03

9.  Radiofrequency Ablation, MR Thermometry, and High-Spatial-Resolution MR Parametric Imaging with a Single, Minimally Invasive Device.

Authors:  M Arcan Ertürk; Shashank Sathyanarayana Hegde; Paul A Bottomley
Journal:  Radiology       Date:  2016-05-26       Impact factor: 11.105

10.  Minimum acquisition methods for simultaneously imaging T(1), T(2), and proton density with B(1) correction and no spin-echoes.

Authors:  Guan Wang; AbdEl-Monem M El-Sharkawy; Paul A Bottomley
Journal:  J Magn Reson       Date:  2014-03-01       Impact factor: 2.229

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.