Dependence of RF heating on SAR and implant position in a 1.5T MR system.

Hiroyuki Muranaka1, Takayoshi Horiguchi, Shuji Usui, Yoshitake Ueda, Osamu Nakamura, Fumiaki Ikeda.   

Abstract

PURPOSE: We evaluated radiofrequency (RF) heating of a humerus implant embedded in a gel phantom during magnetic resonance (MR) imaging for the specific absorption rate (SAR), angle between the implant and static magnetic field (B(0)), and position of the implant in the irradiation coil.
METHODS: We embedded a stainless steel humerus implant 2 cm deep in tissue-equivalent loop and mass phantoms, placed it parallel to the static magnetic field of a 1.5T MR scanner, and recorded the temperatures of the implant surface with RF-transparent fiberoptic sensors. We measured rises in temperature at the tips of the implant by varying the SAR from 0.2 to 4.0 W/kg and evaluated RF heating of the implant for its angle to B(0) and its displacement along B(0) from the center of the RF irradiation coil.
RESULTS: RF heating was similar for the loop and mass phantoms because the eddy current flows through the periphery of both. As the SAR increased, the temperature at the implant tip increased, and there was a linear relationship between the SAR and temperature rise. The values were 6.4 degrees C at 2.0 W/kg and 12.7 degrees C at 4.0 W/kg. Rise in temperature decreased steeply as the angle between the implant and B(0) surpassed 45 degrees . In addition, as the implant was displaced from the center of the RF coil to both ends, the rise in temperature decreased.
CONCLUSION: The rise in temperature in deep tissue was estimated to be higher than 1.0 degrees C for SAR above 0.4 W/kg. RF heating was greatest when the implant was set parallel to B(0). In MR imaging of patients with implants, there is a risk of RF heating when the loop of the eddy current is formed inside the body.

Entities:  

Mesh:

Substances:

Year:  2007        PMID: 18239357     DOI: 10.2463/mrms.6.199

Source DB:  PubMed          Journal:  Magn Reson Med Sci        ISSN: 1347-3182            Impact factor:   2.471


  6 in total

1.  A multichannel, real-time MRI RF power monitor for independent SAR determination.

Authors:  Abdel-Monem M El-Sharkawy; Di Qian; Paul A Bottomley; William A Edelstein
Journal:  Med Phys       Date:  2012-05       Impact factor: 4.071

2.  3-T MRI safety assessments of magnetic dental attachments and castable magnetic alloys.

Authors:  M Hasegawa; K Miyata; Y Abe; T Ishii; T Ishigami; K Ohtani; E Nagai; T Ohyama; Y Umekawa; S Nakabayashi
Journal:  Dentomaxillofac Radiol       Date:  2015-03-18       Impact factor: 2.419

3.  Radiofrequency heating of metallic dental devices during 3.0 T MRI.

Authors:  M Hasegawa; K Miyata; Y Abe; T Ishigami
Journal:  Dentomaxillofac Radiol       Date:  2013-03-21       Impact factor: 2.419

4.  Visible Human Project® female surface based computational phantom (Nelly) for radio-frequency safety evaluation in MRI coils.

Authors:  Gregory M Noetscher; Peter Serano; William A Wartman; Kyoko Fujimoto; Sergey N Makarov
Journal:  PLoS One       Date:  2021-12-10       Impact factor: 3.240

Review 5.  Recent Advances in Transition-Metal Based Nanomaterials for Noninvasive Oncology Thermal Ablation and Imaging Diagnosis.

Authors:  Qiuxia Peng; Zhangbo Qian; Huali Gao; Kun Zhang
Journal:  Front Chem       Date:  2022-04-14       Impact factor: 5.545

6.  Detailing radio frequency heating induced by coronary stents: a 7.0 Tesla magnetic resonance study.

Authors:  Davide Santoro; Lukas Winter; Alexander Müller; Julia Vogt; Wolfgang Renz; Celal Ozerdem; Andreas Grässl; Valeriy Tkachenko; Jeanette Schulz-Menger; Thoralf Niendorf
Journal:  PLoS One       Date:  2012-11-21       Impact factor: 3.240

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.