Literature DB >> 25275576

Evaluation of epidural and peripheral nerve catheter heating during magnetic resonance imaging.

Sean Owens1, M Arcan Erturk, Jean-Pierre P Ouanes, Jamie D Murphy, Christopher L Wu, Paul A Bottomley.   

Abstract

BACKGROUND: Many epidural and peripheral nerve catheters contain conducting wire that could heat during magnetic resonance imaging (MRI), requiring removal for scanning.
METHODS: We tested 2 each of 6 brands of regional analgesia catheters (from Arrow International [Reading, Pennsylvania], B. Braun Medical Inc [Bethlehem, Pennsylvania], and Smiths Medical/Portex [Keene, New Hampshire]) for exposure to clinical 1.5- and 3-T MRI. Catheters testing as nonmagnetic were placed in an epidural configuration in a standard human torso-sized phantom, and an MRI pulse sequence applied at the maximum scanner-allowed radiofrequency specific absorption rate (SAR) for 15 minutes. Temperature and SAR exposure were sampled during MRI using multiple fiberoptic temperature sensors.
RESULTS: Two catheters (the Arrow StimuCath Peripheral Nerve and B. Braun Medical Perifix FX Epidural) were found to be magnetic and not tested further. At 3 T, exposure of the remaining 3 epidural and 1 peripheral nerve catheter to the scanner's maximum RF exposure elicited anomalous heating of 4°C to 7°C in 2 Arrow Epidural (MultiPort and Flex-Tip Plus) catheters at the entry points. Temperature increases for the other catheters at 3 T, and all catheters at 1.5 T were 1.4°C or less. When normalized to the body-average US Food and Drug Administration guideline SAR of 4 W/kg, maximum projected temperature increases were 0.1°C to 2.5°C at 1.5 T and 0.7°C to 2.7°C at 3 T, except for the Arrow MultiPort Flex-Tip Plus catheter at 3 T whose increase was 14°C.
CONCLUSIONS: Most but not all catheters can be left in place during 1.5-T MRI scans. Heating of less than 3°C during MRI for most catheters is not expected to be injurious. While heating was lower at 1.5 T versus 3 T, performance differences between products underscore the need for safety testing before performing MRI.

Entities:  

Mesh:

Year:  2014        PMID: 25275576      PMCID: PMC4208947          DOI: 10.1097/AAP.0000000000000151

Source DB:  PubMed          Journal:  Reg Anesth Pain Med        ISSN: 1098-7339            Impact factor:   6.288


  17 in total

1.  MRI of acute spinal epidural hematomas.

Authors:  E M Sklar; J M Post; S Falcone
Journal:  J Comput Assist Tomogr       Date:  1999 Mar-Apr       Impact factor: 1.826

2.  Biomedical implants and devices: assessment of magnetic field interactions with a 3.0-Tesla MR system.

Authors:  Frank G Shellock
Journal:  J Magn Reson Imaging       Date:  2002-12       Impact factor: 4.813

Review 3.  Radiofrequency safety for interventional MRI procedures.

Authors:  Ergin Atalar
Journal:  Acad Radiol       Date:  2005-09       Impact factor: 3.173

Review 4.  MRI-induced heating of selected thin wire metallic implants-- laboratory and computational studies-- findings and new questions raised.

Authors:  H Bassen; W Kainz; G Mendoza; T Kellom
Journal:  Minim Invasive Ther Allied Technol       Date:  2006       Impact factor: 2.442

Review 5.  Imaging the brain. First of two parts.

Authors:  S Gilman
Journal:  N Engl J Med       Date:  1998-03-19       Impact factor: 91.245

6.  RF heating of implanted spinal fusion stimulator during magnetic resonance imaging.

Authors:  C K Chou; J A McDougall; K W Chan
Journal:  IEEE Trans Biomed Eng       Date:  1997-05       Impact factor: 4.538

7.  Epidural hematoma following epidural catheter placement in a patient with chronic renal failure.

Authors:  M Basta; P Sloan
Journal:  Can J Anaesth       Date:  1999-03       Impact factor: 5.063

Review 8.  Imaging the adult brain.

Authors:  I Moseley
Journal:  J Neurol Neurosurg Psychiatry       Date:  1995-01       Impact factor: 10.154

9.  Epidural hematoma after outpatient epidural anesthesia.

Authors:  Andre Gilbert; Brian D Owens; Michael F Mulroy
Journal:  Anesth Analg       Date:  2002-01       Impact factor: 5.108

Review 10.  Basic principles of thermal dosimetry and thermal thresholds for tissue damage from hyperthermia.

Authors:  M W Dewhirst; B L Viglianti; M Lora-Michiels; M Hanson; P J Hoopes
Journal:  Int J Hyperthermia       Date:  2003 May-Jun       Impact factor: 3.914

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.