Daryl Lindsay Williams1, James M Zeng, Karl D Alexander, David T Andrews.
Abstract
We conducted a randomised single-blind controlled trial comparing the LMA-Unique (LMAU) and the AMBU AuraOnce (AMBU) disposable laryngeal mask in spontaneously breathing adult patients undergoing general anaesthesia. Eighty-two adult patients (ASA status I-IV) were randomly allocated to receive the LMAU or AMBU and were blinded to device selection. Patients received a standardized anesthetic and all airway devices were inserted by trained anaesthetists. Size selection was guided by manufacturer recommendations. All data were collected by a single, unblinded observer. When compared with the LMAU, the AMBU produced significantly higher airway sealing pressures (AMBU 20 ± 6; LMAU 15 ± 7 cm H(2)O; P = 0.001). There was no statistical difference between the two devices for overall success rate, insertion time, number of adjustments, laryngeal alignment, blood-staining, and sore throat (P ≥ 0.05). The AMBU AuraOnce disposable laryngeal mask provided a higher oropharyngeal leak pressure compared to the LMA Unique in spontaneously breathing adult patients.Entities:
Year: 2012 PMID: 22505884 PMCID: PMC3299248 DOI: 10.1155/2012/405812
Source DB: PubMed Journal: Anesthesiol Res Pract ISSN: 1687-6962
Patient demographics, surgery type, and airway sizing.
| LMAU ( | AMBU ( |
| |
|---|---|---|---|
| Age (Years) | 4620 (19–80) | 4516 (19–80) | 0.707 |
| Weight (kg) | 7213 | 7516 | 0.421 |
| Height (cm) | 17211 | 16812 | 0.114 |
| BMI (kg/m2) | 244 | 265 | 0.054 |
| Male Gender | 19 (50) | 18 (44) | 0.746 |
| ASA (I/II/III/IV) | 11/22/5/0 | 16/20/4/1 | 0.635 |
| Surgery Type | 0.875 | ||
| General | 14 (37) | 21 (51) | |
| Plastics | 12 (32) | 13 (32) | |
| Orthopaedics | 8 (21) | 5 (12) | |
| Urology | 4 (10) | 2 (5) | |
| Airway Size (3/4/5) | 14/19/5 | 19/16/6 | 0.738 |
| Duration Surgery (mins) | 3722 (15–110) | 4620 (10–130) | 0.375 |
All data are presented as mean standard deviation (range) or numbers (%). LMAU: LMA Unique laryngeal mask, AMBU: AMBU AuraOnce laryngeal mask, BMI: body mass index, ASA: American Society of Anesthesiologists physical status score.
Results.
| LMAU ( | AMBU ( |
| |
|---|---|---|---|
| Insertion success | |||
| First time | 31 (82) | 35 (85) | 0.738 |
| Overall | 38 (100) | 40 (98) | 0.519 |
| Insertion time (s) | 3216 (11–95) | 2710 (13–51) | 0.111 |
| OLP (cm H2O) | 157 (4–30) | 206 (5–30) | 0.001 |
| Laryngeal Alignment |
|
| |
| I | 21 (55) | 21 (54) | |
| II | 9 (24) | 9 (23) | |
| III | 7 (18) | 7 (18) | |
| IV | 1 (3) | 2 (5) | 0.613 |
| Blood Staining | 3 (8) | 4 (10) | 0.543 |
| Sore Throat PACU | 4 (11) | 3 (8) | 0.457 |
All data are presented as mean + standard deviation (range) or numbers (%). LMAU: LMA Unique laryngeal mask, AMBU: AMBU AuraOnce laryngeal mask, OLP: oropharyngeal leakage pressure, laryngeal alignment (I: visualise entire vocal cords, II: posterior component of cords visible, III: epiglottis but no cords visible, IV: no vocal cords or epiglottis visible).