Laryngeal mask airway size selection in males and females: ease of insertion, oropharyngeal leak pressure, pharyngeal mucosal pressures and anatomical position.

J Brimacombe1, C Keller.   

Abstract

We have compared ease of insertion, oropharyngeal leak pressure, directly measured pharyngeal mucosal pressure and anatomical position (assessed fibreoptically) for the size 4 and size 5 laryngeal mask airway (LMA) in 20 male and 20 female patients. Microchip pressure sensors were attached to the LMA at locations corresponding to the piriform fossa, hypopharynx, base of the tongue, lateral and posterior pharynx, and the oropharynx. Oropharyngeal leak pressure, mucosal pressure and fibreoptic position were recorded during inflation of the cuff from 0 to 30 ml in 10-ml increments. In males, oropharyngeal leak pressure over the inflation range was higher for size 5 (21 vs 17 cm H2O; P = 0.01); mucosal pressure over the inflation range was higher in the posterior pharynx for size 4 (7 vs 2 cm H2O; P = 0.007), and higher in the piriform fossa (8 vs 5 cm H2O; P = 0.003) and hypopharynx (9 vs 5 cm H2O; P = 0.003) for size 5. In females, oropharyngeal leak pressure over the inflation range was the same (21 vs 21 cm H2O), but mucosal pressure over the inflation range was higher in the piriform fossa (21 vs 8 cm H2O; P = 0.003) and posterior pharynx (4 vs 2 cm H2O; P = 0.004) for size 4, and higher in the lateral pharynx (5 vs 1 cm H2O; P = 0.01) and oropharynx (11 vs 5 cm H2O; P = 0.009) for size 5. The distribution of mucosal pressure was different for size 4 between males and females, but not for size 5. For both males and females, fibreoptic position was similar. We conclude that the size 5 LMA is optimal in males, but either size is suitable for females. The shape of the pharynx may be different between males and females.

Entities:  

Mesh:

Year:  1999        PMID: 10536546     DOI: 10.1093/bja/82.5.703

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  8 in total

1.  Comparison of weight-based and pinna size-based selection of ProSeal laryngeal mask airway in paediatric population - A prospective exploratory trial.

Authors:  Ramya Ravi; V K Mohan; Ashok Shankar Badhe; Sandeep Kumar Mishra; Prasanna Udupi Bidkar
Journal:  Indian J Anaesth       Date:  2019-01

2.  Comparison of the size 3 and size 4 ProSeal™ laryngeal mask airway in anesthetized, non-paralyzed women: a randomized controlled trial.

Authors:  Mi-Hyun Kim; Jung-Won Hwang; Eun-Sung Kim; Sung-Hee Han; Young-Tae Jeon; Sun-Mi Lee
Journal:  J Anesth       Date:  2014-09-24       Impact factor: 2.078

3.  Randomised Comparison of the AMBU AuraOnce Laryngeal Mask and the LMA Unique Laryngeal Mask Airway in Spontaneously Breathing Adults.

Authors:  Daryl Lindsay Williams; James M Zeng; Karl D Alexander; David T Andrews
Journal:  Anesthesiol Res Pract       Date:  2012-02-29

4.  Laryngeal mask placement in a teaching institution: analysis of difficult placements.

Authors:  Anastasia D Katsiampoura; Peter V Killoran; Ruggero M Corso; Chunyan Cai; Carin A Hagberg; Davide Cattano
Journal:  F1000Res       Date:  2015-04-29

5.  Techniques for the insertion of the ProSeal laryngeal mask airway: comparison of the Foley airway stylet tool with the introducer tool in a prospective, randomized study.

Authors:  Mao-Kai Chen; Hung-Te Hsu; I-Cheng Lu; Chih-Kai Shih; Ya-Chun Shen; Kuang-Yi Tseng; Kuang-I Cheng
Journal:  BMC Anesthesiol       Date:  2014-11-18       Impact factor: 2.217

6.  Comparison of ProSeal laryngeal mask airway size 2 and 2½ in anesthetized and paralyzed pediatric patients with same weight group: A prospective randomized clinical study.

Authors:  Reena Mahajan; Susheela Taxak
Journal:  Anesth Essays Res       Date:  2014 Sep-Dec

7.  Size selection of the Ambu AuraOnce laryngeal mask in Chinese men weighing >70 kg: a pilot study.

Authors:  Jiahui Chen; Chunhuan Chen; Wei Xu; Xiaoguang Zhang
Journal:  J Int Med Res       Date:  2021-05       Impact factor: 1.671

8.  Randomized Comparison of Actual and Ideal Body Weight for Size Selection of the Laryngeal Mask Airway Classic in Overweight Patients.

Authors:  Min-Soo Kim; Jong Seok Lee; Sang Beom Nam; Hyo Jong Kang; Ji Eun Kim
Journal:  J Korean Med Sci       Date:  2015-07-15       Impact factor: 2.153