| Literature DB >> 25929558 |
Ajay Kumar Pajiyar1, Zhiting Wen2, Haiyun Wang3, Lin Ma4, Lumin Miao5, Guolin Wang6.
Abstract
BACKGROUND: The Guardian Laryngeal Mask Airway (G-LMA) is a new silicone-based single-use extraglottic device with the drainage port and a cuff pilot valve with pressure indicator. The aim of this study is to compare the clinical performance of this laryngeal mask airway with ProSeal laryngeal mask airway (P-LMA).Entities:
Mesh:
Year: 2015 PMID: 25929558 PMCID: PMC4429672 DOI: 10.1186/s12871-015-0039-3
Source DB: PubMed Journal: BMC Anesthesiol ISSN: 1471-2253 Impact factor: 2.217
Figure 1Guardian laryngeal mask airway with pilot balloon valve indicating pressure logo (Ultimate Medical Pty Ltd, Richmond, Vic, Australia).
Figure 2Patients flow chart (G-LMA vs. P-LMA).
Demographic characteristics and type of surgery performed
| Variables | Group P (P-LMA)(n = 40) | Group G (G-LMA)(n = 40) |
|---|---|---|
|
| 45.20 ± 14.78 | 45.75 ± 13.58 |
|
| 8/32 | 12/28 |
|
| 59.75 ± 4.72 | 61.65 ± 8.95 |
|
| 160.30 ± 4.11 | 162.20 ± 4.86 |
|
| 23.27 ± 1.85 | 23.38 ± 2.71 |
|
| ||
|
| 20 | 18 |
|
| 8 | 10 |
|
| 8 | 6 |
|
| 4 | 6 |
P-LMA: ProSeal laryngeal mask airway; G-LMA: Guardian laryngeal mask airway.
Airway sealing pressure, cuff volume, intracuff pressure and pharyngolaryngeal morbidity
| Variables | Group P (P-LMA)(n = 40) | Group G (G-LMA)(n = 40) | |
|---|---|---|---|
|
| 13.90 ± 1.19** | 22.60 ± 3.27** | <0.0001** |
|
| 29.41 ± 4.33* | 32.36 ± 4.93* | 0.04* |
|
| |||
|
| 58.83 ± 1.14 | 58.90 ± 1.15 | 0.823 |
|
| 58.67 ± 1.09 | 58.27 ± 1.20 | 0.182 |
|
| 58.40 ± 1.23 | 58.33 ± 1.35 | 0.870 |
|
| 58.27 ± 1.09 | 58.44 ± 1.03 | 0.658 |
|
| 9/40 (22.5%) | 8/40 (20%) | 1.000 |
|
| |||
|
| |||
|
| 7/40 (17.5%) | 6/40 (15%) | 1.000 |
|
| 7/40 | 6/40 | |
|
| 0 | 0 | |
|
| |||
|
| 5/40 (12.5%) | 4/40 (10%) | 1.000 |
|
| 5/40 | 4/40 | |
|
| 0 | 0 | |
|
| |||
|
| 0 | 0 | |
|
| 0 | 0 | |
|
| 0 | 0 | |
**P < 0.01, *p < 0.05, P-LMA: ProSeal laryngeal mask airway; G-LMA: Guardian laryngeal mask airway; h: hours, min: minutes, ml: milliliter.
Device insertion characteristics, gastric tube insertion characteristics and fiber-optic view
| Variables | Group P (P-LMA) | Group G (G-LMA) | |
|---|---|---|---|
| (n = 40) | (n = 40) | ||
|
| |||
|
| 38 (95%) | 39 (97.5%) | 1.000 |
|
| 2 (5%) | 1 (2.5%) | |
|
| 100% | 100% | |
|
| 15.95 ± 1.84** | 13.20 ± 1.82** | <0.0001** |
|
| |||
|
| 28 | 30 | |
|
| 10 | 9 | |
|
| 2 | 1 | |
|
| 0 | 0 | |
|
| 6/40 | 4/40 | 0.373 |
|
| 6/40 | 4/40 | 0.373 |
|
| |||
|
| 35 (87.5%) | 38 (95%) | 0.431 |
|
| 5 (12.5%) | 2 (5%) | |
|
| 40 (100%) | 40 (100%) | |
|
| 13.75 ± 1.74** | 10.05 ± 1.35** | <0.0001** |
|
| |||
|
| 14/21/3/2 | 17/20/2/1 | 0.836 |
**P < 0.01, *p < 0.05, P-LMA: ProSeal laryngeal mask airway; G-LMA: Guardia laryngeal mask airway.
Figure 3Comparison of airway sealing pressure in two groups (G-LMA vs. P-LMA). Results are given as means ± SD,*p < 0.05.