Literature DB >> 22457404

Main clinical, therapeutic and technical factors related to patient's maximum skin dose in interventional cardiology procedures.

N Journy1, S Sinno-Tellier, C Maccia, A Le Tertre, P Pirard, P Pagès, D Eilstein, J Donadieu, O Bar.   

Abstract

OBJECTIVE: The study aimed to characterise the factors related to the X-ray dose delivered to the patient's skin during interventional cardiology procedures.
METHODS: We studied 177 coronary angiographies (CAs) and/or percutaneous transluminal coronary angioplasties (PTCAs) carried out in a French clinic on the same radiography table. The clinical and therapeutic characteristics, and the technical parameters of the procedures, were collected. The dose area product (DAP) and the maximum skin dose (MSD) were measured by an ionisation chamber (Diamentor; Philips, Amsterdam, The Netherlands) and radiosensitive film (Gafchromic; International Specialty Products Advanced Materials Group, Wayne, NJ). Multivariate analyses were used to assess the effects of the factors of interest on dose.
RESULTS: The mean MSD and DAP were respectively 389 mGy and 65 Gy cm(-2) for CAs, and 916 mGy and 69 Gy cm(-2) for PTCAs. For 8% of the procedures, the MSD exceeded 2 Gy. Although a linear relationship between the MSD and the DAP was observed for CAs (r=0.93), a simple extrapolation of such a model to PTCAs would lead to an inadequate assessment of the risk, especially for the highest dose values. For PTCAs, the body mass index, the therapeutic complexity, the fluoroscopy time and the number of cine frames were independent explanatory factors of the MSD, whoever the practitioner was. Moreover, the effect of technical factors such as collimation, cinematography settings and X-ray tube orientations on the DAP was shown.
CONCLUSION: Optimising the technical options for interventional procedures and training staff on radiation protection might notably reduce the dose and ultimately avoid patient skin lesions.

Entities:  

Mesh:

Year:  2012        PMID: 22457404      PMCID: PMC3486680          DOI: 10.1259/bjr/30010948

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  23 in total

1.  Skin dose and dose-area product values for interventional cardiology procedures.

Authors:  E Vano; L Gonzalez; J I Ten; J M Fernandez; E Guibelalde; C Macaya
Journal:  Br J Radiol       Date:  2001-01       Impact factor: 3.039

2.  Clinical and technical determinants of the complexity of percutaneous transluminal coronary angioplasty procedures: analysis in relation to radiation exposure parameters.

Authors:  G Bernardi; R Padovani; G Morocutti; E Vaño; M R Malisan; M Rinuncini; L Spedicato; P M Fioretti
Journal:  Catheter Cardiovasc Interv       Date:  2000-09       Impact factor: 2.692

3.  Radiation exposure to patient's skin during percutaneous coronary intervention for various lesions, including chronic total occlusion.

Authors:  Shigeru Suzuki; Shigeru Furui; Hiroshi Kohtake; Naoyuki Yokoyama; Ken Kozuma; Yoshito Yamamoto; Takaaki Isshiki
Journal:  Circ J       Date:  2006-01       Impact factor: 2.993

4.  Patient dose related to the complexity of interventional cardiology procedures.

Authors:  R Padovani; G Bernardi; M R Malisan; E Vañó; G Morocutti; P M Fioretti
Journal:  Radiat Prot Dosimetry       Date:  2001       Impact factor: 0.972

5.  The dependence of patient dose on factors relating to the technique and complexity of Interventional Cardiology procedures.

Authors:  Miltiadis G Delichas; Kyriakos Psarrakos; Konstantinos Hatziioannou; Georgios Giannoglou; Elisabeth Molyvda-Athanasopoulou; Emmanouil Papanastassiou; Anastasios Sioundas
Journal:  Phys Med       Date:  2005 October - December       Impact factor: 2.685

6.  Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures.

Authors:  Koichi Chida; Haruo Saito; Hiroki Otani; Masahiro Kohzuki; Shoki Takahashi; Shogo Yamada; Kunio Shirato; Masayuki Zuguchi
Journal:  AJR Am J Roentgenol       Date:  2006-03       Impact factor: 3.959

7.  Patient doses and dosimetric evaluations in interventional cardiology.

Authors:  Dogan Bor; Turan Olğar; Türkay Toklu; Ayça Cağlan; Elif Onal; Renato Padovani
Journal:  Phys Med       Date:  2008-04-25       Impact factor: 2.685

8.  Effective techniques for reduction of radiation dosage to patients undergoing invasive cardiac procedures.

Authors:  E Kuon; C Glaser; J B Dahm
Journal:  Br J Radiol       Date:  2003-06       Impact factor: 3.039

9.  A large-scale multicentre study of patient skin doses in interventional cardiology: dose-area product action levels and dose reference levels.

Authors:  E Bogaert; K Bacher; K Lemmens; M Carlier; W Desmet; X De Wagter; D Djian; C Hanet; G Heyndrickx; V Legrand; Y Taeymans; H Thierens
Journal:  Br J Radiol       Date:  2009-01-05       Impact factor: 3.039

10.  Patient dose in cardiac radiology.

Authors:  Andreas I Stratis; Prodromos L Anthopoulos; Isidoros P Gavaliatsis; Georghios P Ifantis; Anastasios I Salahas; Ioannis P Antonellis; Antonios G Tavernarakis; Michael I Molfetas
Journal:  Hellenic J Cardiol       Date:  2009 Jan-Feb
View more
  4 in total

1.  Understanding and Minimising Occupational Radiation in the Catheterisation Laboratory with PISAX and the ACIST CVi® Contrast Delivery System.

Authors:  Olivier Bar
Journal:  Interv Cardiol       Date:  2013-03

2.  Radiation doses during cardiac catheterisation procedures in India: a multicentre study: Radiation dose study.

Authors:  Vijayakumar Subban; Sophie Amelot; Suma M Victor; Anil Potdar; Vishawanath Yadav; Tejas Patel; Sanjay Shah; Thomas Alexander; Balakumaran Jeyakumaran; Juno Angel; Mullasari S Ajit
Journal:  AsiaIntervention       Date:  2020-07-20

3.  Patient's Radiation Exposure in Coronary Angiography and Angioplasty: The Impact of Different Projections.

Authors:  Alireza Farajollahi; Atena Rahimi; Ebrahim Khayati Shal; Samad Ghaffari; Morteza Ghojazadeh; Arezou Tajlil; Naser Aslanabadi
Journal:  J Cardiovasc Thorac Res       Date:  2014-12-30

4.  Novel X-ray image noise reduction technology reduces patient radiation dose while maintaining image quality in coronary angiography.

Authors:  T Ten Cate; M van Wely; H Gehlmann; M Mauti; C Camaro; N Reifart; H Suryapranata; M J de Boer
Journal:  Neth Heart J       Date:  2015-11       Impact factor: 2.380

  4 in total