Comparison of a new anterior segment optical coherence tomography and Oculus Pentacam for measurement of anterior chamber depth and corneal thickness.

Xuepei Li1, Yijing Zhou1, Charlotte Aimee Young2, Aiming Chen3, Guangming Jin1, Danying Zheng1.   

Abstract

BACKGROUND: Accurate measurements of anterior chamber depth (ACD) and regional corneal thickness are especially important for the diagnosis and treatment of many ocular disease. This study aimed to evaluate the repeatability and reliability of a new swept source anterior segment optical coherence tomography (AS-OCT) and its agreement with Oculus Pentacam for measurements of ACD and corneal thickness (CT).
METHODS: The central corneal thickness (CCT), superior corneal thickness (SCT), inferior corneal thickness (ICT), nasal corneal thickness (NCT), temporal corneal thickness (TCT) and ACD of the right eye from forty-nine adults aged 18 to 36 years (24.78±4.36 years old) were measured with Pentacam and AS-OCT (CASIA2). All subjects were measured twice with each device. The repeatability was determined using the coefficient of repeatability (COR), the relative COR and the limits of agreement (LOA). Bland-Altman plot was also used for evaluating the agreement between parameters from the two devices.
RESULTS: For the repeatability of CASIA2, the COR of the two measurements of ACD, CCT, SCT, ICT, NCT and TCT was 0.31 mm, 18.58, 25.83, 28.32, 26.71 and 22.09 µm respectively. There were no statistically significant differences between the CT and ACD measurements by CASIA2 (P>0.05). For measurements with Pentacam and CASIA2, the COR of ACD, CCT, SCT, NCT, ICT and TCT was 0.294 mm, 13.10, 51.57, 48.06, 56.21 and 47.69 µm respectively. No significant differences were found between the values measured by CAISA2 and Pentacam for CT and ACD (P>0.05). The Bland-Altman analysis also suggested high consistency between measurements obtained by Pentacam and CASIA2.
CONCLUSIONS: Our results suggest that Pentacam and CASIA2 have good agreement in CT and ACD measurements. The two devices can be considered interchangeable for these parameters' measurements in healthy subjects when monitoring corneal conditions or planning ocular surgery. However, subtle differences between CASIA2 and Pentacam should also be kept in mind for certain specific clinical or research purposes. 2020 Annals of Translational Medicine. All rights reserved.

Entities:  

Keywords:  Anterior chamber depth (ACD); Oculus Pentacam; corneal thickness; swept source anterior segment optical coherence tomography

Year:  2020        PMID: 32793701      PMCID: PMC7396784          DOI: 10.21037/atm-20-187

Source DB:  PubMed          Journal:  Ann Transl Med        ISSN: 2305-5839


  36 in total

1.  Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment.

Authors:  Tanuj Dada; Ramanjit Sihota; Ritu Gadia; Anand Aggarwal; Subrata Mandal; Viney Gupta
Journal:  J Cataract Refract Surg       Date:  2007-05       Impact factor: 3.351

2.  Average 3-dimensional models for the comparison of Orbscan II and Pentacam pachymetry maps in normal corneas.

Authors:  Jean-Louis Bourges; Nicolas Alfonsi; Jean-François Laliberté; Miguel Chagnon; Gilles Renard; Jean-Marc Legeais; Isabelle Brunette
Journal:  Ophthalmology       Date:  2009-09-10       Impact factor: 12.079

3.  Comparison of Oculus Pentacam and Stratus optical coherence tomography for measurement of central corneal thickness.

Authors:  Osman Melih Ceylan; Adem Turk; Cuneyt Erdurman; Tarkan Mumcuoglu; Uzeyir Erdem; Gokcen Gokce; Selim Dagli
Journal:  Cornea       Date:  2011-06       Impact factor: 2.651

4.  Anterior chamber depth measurements using Scheimpflug imaging and optical coherence tomography: repeatability, reproducibility, and agreement.

Authors:  Qinmei Wang; Xixia Ding; Giacomo Savini; Hao Chen; Yifan Feng; Chao Pan; Yanjun Hua; Jinhai Huang
Journal:  J Cataract Refract Surg       Date:  2014-11-18       Impact factor: 3.351

5.  Comparison of central and peripheral corneal thickness measurements with scanning-slit, Scheimpflug and Fourier-domain ocular coherence tomography.

Authors:  J Bradley Randleman; Michael J Lynn; Claudia E Perez-Straziota; Heather M Weissman; Sang Woo Kim
Journal:  Br J Ophthalmol       Date:  2015-03-30       Impact factor: 4.638

6.  Central corneal thickness determined with optical coherence tomography in various types of glaucoma.

Authors:  M Bechmann; M J Thiel; B Roesen; S Ullrich; M W Ulbig; K Ludwig
Journal:  Br J Ophthalmol       Date:  2000-11       Impact factor: 4.638

7.  Comparison of corneal thickness measurements by ultrasound and scheimpflug photography in eyes that have undergone laser in situ keratomileusis.

Authors:  Joseph B Ciolino; Steven S Khachikian; Michael W Belin
Journal:  Am J Ophthalmol       Date:  2007-11-07       Impact factor: 5.258

8.  Diurnal Variation of Optical Coherence Tomography Measurements of Static and Dynamic Anterior Segment Parameters.

Authors:  Benjamin Y Xu; Rafaella C Penteado; Robert N Weinreb
Journal:  J Glaucoma       Date:  2018-01       Impact factor: 2.503

9.  Comparison of central corneal thickness measurements by visante anterior segment optical coherence tomography with ultrasound pachymetry.

Authors:  Paul S Zhao; Tien Yin Wong; Wan-Ling Wong; Seang-Mei Saw; Tin Aung
Journal:  Am J Ophthalmol       Date:  2007-06       Impact factor: 5.258

10.  Assessment of tilt and decentration of crystalline lens and intraocular lens relative to the corneal topographic axis using anterior segment optical coherence tomography.

Authors:  Shuhei Kimura; Yuki Morizane; Yusuke Shiode; Masayuki Hirano; Shinichiro Doi; Shinji Toshima; Atsushi Fujiwara; Fumio Shiraga
Journal:  PLoS One       Date:  2017-09-01       Impact factor: 3.240

View more
  4 in total

1.  Monitoring of central corneal thickness after phacoemulsification-comparison of statical and rotating Scheimpflug pachymetry, and spectral-domain OCT.

Authors:  Daniel M Handzel; Carsten H Meyer; Alfred Wegener
Journal:  Int J Ophthalmol       Date:  2022-08-18       Impact factor: 1.645

2.  Corneal Analysis with Swept Source Optical Coherence Tomography in Patients with Coexisting Cataract and Fuchs Endothelial Corneal Dystrophy.

Authors:  Anna Nowińska; Edyta Chlasta-Twardzik; Michał Dembski; Klaudia Ulfik-Dembska; Edward Wylęgała
Journal:  Diagnostics (Basel)       Date:  2021-02-02

3.  Agreement of Anterior Segment Parameter Measurements With CASIA 2 and IOLMaster 700.

Authors:  Xiaoting Ruan; GuangYao Yang; Zhaoxia Xia; Jiaqing Zhang; Xiaoxun Gu; Yuan Tan; Zhenzhen Liu; Lixia Luo
Journal:  Front Med (Lausanne)       Date:  2022-02-11

4.  Anterior Ocular Biometrics Using Placido-scanning-slit System, Rotating Scheimpflug Tomography, and Swept-source Optical Coherence Tomography.

Authors:  Soyoung Ryu; Sook Hyun Yoon; Ikhyun Jun; Kyoung Yul Seo; Eung Kweon Kim; Tae-Im Kim
Journal:  Korean J Ophthalmol       Date:  2022-05-06
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.