The dose delivery verification for a head and neck static intensity modulated radiation therapy (IMRT) case using a scanning liquid ionization chamber electronic portal imaging device (SLIC-EPID) was investigated. Acquired electronic portal images were firstly converted into transmitted dose maps using an in-house developed method. The dose distributions were then compared with those calculated in a virtual EPID using the Pinnacle(3) treatment planning system (TPS). Using gamma evaluation with the ΔD(max) and DTA criteria of 3%/2.54 mm, an excellent agreement was observed between transmitted dose measured using SLIC-EPID and that calculated by TPS (gamma score approximately 95%) for large MLC fields. In contrast, for several small subfields, due to SLIC-EPID image blurring, significant disagreement was found in the gamma results. Differences between EPID and TPS dose maps were also observed for several parts of the radiation subfields, when the radiation beam passed through air on the outside of tissue. The transmitted dose distributions measured using portal imagers such as SLIC-EPID can be used to verify the dose delivery to a patient. However, several aspects such as accurate calibration procedure and imager response under different conditions should be taken into the consideration. In addition, SLIC-EPID image blurring is another important issue, which should be considered if the SLIC-EPID is used for clinical dosimetry verification.
The dose delivery verification for a head and neck static intensity modulated radiation therapy (IMRT) case using a scanning liquid ionization chamber electronic portal imaging device (SLIC-EPID) was investigated. Acquired electronic portal images were firstly converted into transmitted dose maps using an in-house developed method. The dose distributions were then compared with those calculated in a virtual EPID using the Pinnacle(3) treatment planning system (TPS). Using gamma evaluation with the ΔD(max) and DTA criteria of 3%/2.54 mm, an excellent agreement was observed between transmitted dose measured using SLIC-EPID and that calculated by TPS (gamma score approximately 95%) for large MLC fields. In contrast, for several small subfields, due to SLIC-EPID image blurring, significant disagreement was found in the gamma results. Differences between EPID and TPS dose maps were also observed for several parts of the radiation subfields, when the radiation beam passed through air on the outside of tissue. The transmitted dose distributions measured using portal imagers such as SLIC-EPID can be used to verify the dose delivery to a patient. However, several aspects such as accurate calibration procedure and imager response under different conditions should be taken into the consideration. In addition, SLIC-EPID image blurring is another important issue, which should be considered if the SLIC-EPID is used for clinical dosimetry verification.
The term “transmitted dose” as used in in vivo dosimetry, is defined as the dose distribution behind a patient/phantom, can be used for two purposes: verification of dose delivery using dose distribution behind the patient[1-6] and the calculation of dose distribution in the Volume of Interest (VOI) or in the mid-plane of a patient, applying back-projection algorithms to the measured transmitted doses.[7-12]Electronic Portal Imaging Devices (EPIDs) are standard devices incorporated in modern medical linear accelerators. They can be used for a range of tasks in radiation therapy such as patient set-up verification,[13-18] dosimetric verification,[19-27] and Quality Assurance programs.[28-37] Although EPID calibration for two-dimensional dosimetric purposes is a labor-intensive task, after calibration the use of EPID offers significant time efficiency due to the lack of film processing, frequent film calibration and scanning procedure.Although today, amorphous silicon-based EPIDs (aSi-EPIDs) are the most popular EPIDs incorporated in newly installed linacs, as a result of rapid EPID/linac installation growth in the past decade, many clinical linacs around the world still utilize scanning liquid ionization chamber EPIDs (SLIC-EPIDs), especially in developing countries. Furthermore, there are several advantages of SLIC-EPID response. For example, it has been shown that this type of EPIDs is able to detect 0.1 mm Multileaf Collimator (MLC) leaf displacement in Intensity Modulated Radiation Therapy (IMRT) applications[37] and it is a suitable device for radiation dose rate evaluation. In contrast, the image blurring for small radiation fields observed in the current study shows there is still need for studies demonstrating the use of SLIC-EPIDs in clinical procedures, e.g. the IMRT.This study is an EPID dosimetry based assessment of a clinical head and neck IMRT delivery. In a previous study, we have shown that a scanning liquid chamber EPID can be used to verify dose delivery for a segmented intensity modulated radiation therapy (sIMRT) for prostate.[38] In order to investigate the SLIC-EPID response in a treatment scenario with more inhomogeneous regions compared to the pelvic area, the current study focuses on a more complex situation in the head and neck. As the first comment, it should be pointed out that throughout the current study, a blurring of electronic portal images (EPIs) was observed for small radiation field sizes. It appears that the blurring can be categorized as two horizontal and vertical linear blurring combined during image acquisition. The impact of additional build-up layer and the conventional jaws positioning on the EPI blurring which occurred for small radiation fields has also been investigated for a range of regular and irregular radiation fields.
Materials and Methods
A Varian 600CD linear accelerator equipped with a standard 80-leaf MLC and an SLIC-EPID was used. A typical sIMRT MLC sequence file for a head and neck treatment and an anthropomorphic Rando phantom were used. The head and neck sIMRT treatment consisted of five fields applied at gantry angles of 0°, 20°, 230°, 270° and 300°. The minimum and maximum numbers of subfields used in the current study were 10 and 18 for gantry angles of 300° and 230°, respectively. In order to control the undesired beam attenuation effects on the transmitted dose maps, caused by the treatment couch, reported previously as one of the main obstacles in transmitted dose verification,[38] the treatment couch was removed from under the head and neck area of the phantom. A sagittal view of the CT dataset including treatment couch, a water equivalent modeled EPID, the Rando phantom and the dose grid matrix used is shown in [Figure 1].
Figure 1
A sagittal view of a CT dataset of the Rando phantom and the modeled EPID. The position of treatment couch removed is also shown outside the dose grid region
A sagittal view of a CT dataset of the Rando phantom and the modeled EPID. The position of treatment couch removed is also shown outside the dose grid region
Extension of CT images for portal dose calculations
The anthropomorphic phantom was scanned using a CT simulator (AcQSim CT, Philips Medical System, Cleveland, OH, USA) with a 3-mm slice thickness and a matrix size of 512 × 512 pixels. In order to model the EPID at 140 cm from the radiation source, the images were then extended to 1024 × 1024 pixels by adding rows of pixels with pixel values corresponding to air CT numbers around the acquired CT image.[38-40] . For non-zero gantry angles, in order to extract the calculated transmitted dose distributions easily through the calculated dose grid matrix, the original CT images were rotated in the opposite direction to the gantry rotation. The routinely available syntax command for image rotation in MATLAB software (MathWorks, Natick, MA, USA) rotates the image around the central point of this image. This command would generally be used to rotate patient CT images for oblique beams. In the current work, due to the non-central position of the tumor site, the extended image was rotated around the center of Region Of Interest (ROI) The results of image rotations around the central point of image were then compared to that rotated around the center of ROI.
Transmitted dose maps measured using a SLIC-EPID and calculated using Pinnacle3 Treatment Planning System
Transmitted dose maps calculated using Pinnacle3 TPS
In order to calculate the predicted transmitted dose maps using the available version of Pinnacle3 Treatment Planning System (TPS;version 6.2b) (ADAC Inc., PHILIPS Medical System, Milpitas, CA, USA), the extended phantom CT data were imported. For both measured and calculated cases, in order to control the extra-transmission through and between MLC leaves, the conventional collimators were placed 1 cm behind the MLCs’ position. MLC leaf offset with a value of 0.6 mm was applied manually to the TPS MLC leaf settings for all subfields to account for rounded leaf ends.[38-40] . As Figure 1 shows, the EPID was then modeled as a 5 × 30 × 30 cm3 slab behind the phantom. The distance between dmax of a thin slab of water-equivalent material representing the EPID and the isocenter was set up to be 40 cm [source to EPID distance (SED = 140 cm)]. A corresponding fraction of 2 Gy total dose was prescribed for each subfield to the isocenter point located inside the anthropomorphic phantom. A three-dimensional dose grid with the voxel size of 0.175 × 0.175 × 0.175 mm3 was defined to calculate the dose distributions. Due to the limited amount of available computer memory in the TPS computer, the dose grid voxel sizes were defined as 0.25 × 0.25 × 0.25 mm3 for a gantry angle of 270°. The dose delivered to the phantom and the transmitted dose were calculated using the collapsed cone convolution superposition algorithm. The EPID was modeled with a water-equivalent material. From this model, the dose distribution in the build-up layer was found for all subfields and the transmitted dose extracted.
Transmitted dose maps measured using an SLIC-EPID
The EPI acquisition and EPID dose calibration procedure are fully explained in literature.[41-44] To reduce the effects of fluctuation in EPI pixel values, three consecutive EPIs were acquired for each subfield and averaged. Other procedures were the same as described previously for sIMRT prostate case.[38] Details of EPID and treatment planning operating characteristics are shown in [Table 1].
Table 1
SLIC-EPID and Pinnacle3 TPS set-ups for portal dosimetry
SLIC-EPID and Pinnacle3 TPS set-ups for portal dosimetry
The comparison of the measured and calculated transmitted dose distributions
The transmitted dose maps measured using an SLIC-EPID and calculated using a Pinnacle3 TPS were normalized to a point located in a large homogenous area of the irradiated field. In cases where the normalization point is positioned near significant inhomogeneities, especially for small subfields, the uncertainty in the normalization factor is greater. Using an in-house code written in MATLAB and due to the SLIC-EPID pixel size (1.27 mm × 1.27 mm), both sets of transmitted dose maps were then compared using the gamma function algorithm for each subfield and the corresponding accumulated fields with the ΔDmax and Distance To Agreement (DTA) criteria of 3%/2.54 mm.[45]After evaluating individual subfields, the corresponding total field transmitted dose maps were then created by multiplying the relative weighting factor contribution with the dose for each subfield. The corresponding calculated transmitted dose maps for the total fields were also prepared by adding up all of the subfields. The measured and calculated total transmitted dose maps were normalized to a point located in an approximately homogeneous area located inside the MLC field.
SLIC-EPID response for small radiation fields
Since blurring was observed in some small radiation fields during SLIC-EPI acquisition, several issues including the smallest field size with sharp edges, the impact of additional build-up layer on the EPI quality and the impact of conventional jaws were evaluated.
The impact of additional build-up layer
Several EPIs were acquired for nominal square radiation field sizes from 1 × 1 to 5 × 5 cm2, using conventional jaws with and without an additional build-up layer (5mm RW3 layer[41]).To prevent the impact of statistical fluctuations on EPIs, for each set-up, three consecutive EPIs were acquired, averaged and converted to the radiation fluence map, which is the radiation beam reaching the EPID in the absence of patient phantom or any other attenuator[42] . Two series of acquired EPIs (with/without an additional build-up layer) were acquired and the raw SLIC-EPIs were then converted to the radiation fluence map. The results were then compared with those measured using Extended Dose Range 2 (EDR2) film under the same conditions.
The impact of conventional jaws positioning
The impact of conventional jaws positioning on the blurring of EPIs acquired for small MLC fields was also explored. Setting up the conventional jaws in various positions, the impact of intra/inter leaf leakage on the EPI blurring was investigated. This was performed for two groups of MLC fields: firstly, small fields and secondly, the small fields in the presence of larger fields. The conventional jaws were positioned far from MLC leaf positioning (14 × 20 cm2) and also close to the opened MLC leaf positions (0.5 cm behind). The corresponding absolute dose difference was then evaluated.
SLIC-EPID response for high dose gradient regions
In order to investigate the TPS and the SLIC-EPID response at the interface between air and phantom, the homogenous phantom (10 cm thick water-equivalent material), used to verify the proposed calibration procedure, was shifted by 2 cm in the direction of the x and y axes. Using a 17 × 17 cm2 radiation field size, the homogenous phantom was positioned so that a part of the radiation beam passed through the air. The measured and calculated transmitted dose maps were then compared using relative dose difference map and gamma function algorithm, with 3%/2.54 mm criteria.
Results and Discussion
The position of the central point of a CT image and the center of an ROI is shown in [Figure 2a]. In addition, the difference in the CT numbers between a CT image rotated to 60° around the central point of the CT image (green point) and around the Point of Interest (POI) (red point) is illustrated in [Figure 2b]. The area affected by possible fluctuation of CT number, due to the use of image processing for image rotation (approximately ±50),[38] is shown in gray scale. The variation of CT numbers affects the dose delivery plans calculated using a treatment planning system. A significant variation in CT numbers was observed in the high CT number gradient regions. A maximum difference of ±500 in CT numbers was observed in the edge of inhomogeneities. Cross-plane and inplane profiles of the image rotated around the central point of image and around the central point of POI are shown in [Figure 3]. The corresponding CT number difference is also shown as a secondary axis. These line profiles show that in the boundaries of inhomogeneities, a significant variation of CT numbers was observed.
Figure 2
(a) Raw CT image for head and neck region of the Rando phantom at AP direction;(b) the CT number difference for a CT image rotated around the central point of the image and around the central point of the ROI
Figure 3
(a) Inplane and (b) cross-plane profiles of two CT images rotated to 60° gantry angle around the central point of the image, around the POI and the corresponding CT number difference
(a) Raw CT image for head and neck region of the Rando phantom at AP direction;(b) the CT number difference for a CT image rotated around the central point of the image and around the central point of the ROI(a) Inplane and (b) cross-plane profiles of two CT images rotated to 60° gantry angle around the central point of the image, around the POI and the corresponding CT number differenceIn the transmitted dose calculation for the prostate case,[38] because of the ROI positioning in the central part of a patient (consequently in the central region of the images), the conventional image rotation does not affect the results significantly. Additionally, the prostate and surrounding regions are more homogenous than the head and neck sites. In contrast, for ROIs located outside the central region of the body, the rotation around the image center will cause artifacts. Therefore, the patient CT images must be rotated around the central point of the target volume.
Transmitted dose maps measured using an SLIC-EPID and calculated using Pinnacle3 TPS
The corresponding relative dose difference and gamma maps for a typical series of relative transmitted dose maps are shown in [Figure 4]. It refers to several radiation subfield obtained from an oblique beam (gantry angle of 230°). The x and y axes of all images represent the number image pixel number. The relative dose differences and the corresponding gamma maps show that a good agreement was observed inside the radiation fields for most MLC subfields. Several disagreeing regions were observed in the penumbral areas and more significant discrepancies were also observed for several small subfields (2 of 18 subfields) used routinely to boost the dose delivered to the central part of the tumor site. The size of subfields with significant inconsistencies was found to be less than 8 cm2.
Figure 4
Transmitted dose maps measured and calculated using SLIC-EPID and Pinnacle3 TPS, the corresponding relative dose difference and gamma maps for subfields acquired for the gantry angle of 230º. The gamma function criteria are 3%/2.54 mm for all cases
Transmitted dose maps measured and calculated using SLIC-EPID and Pinnacle3 TPS, the corresponding relative dose difference and gamma maps for subfields acquired for the gantry angle of 230º. The gamma function criteria are 3%/2.54 mm for all casesThe total field relative to measured and calculated transmitted dose maps, the relative dose difference and the corresponding gamma maps are shown in [Figure 5] for all five gantry positions used in the current work. Although inaccuracy in the positioning of the normalization points and the possible misalignment of the corresponding subfields decreases the correlation between measured and calculated dose distributions, an acceptable consistency was observed inside the ROIs for all gantry positions.
Figure 5
(A series) Total field transmitted dose maps, (B series) calculated transmitted dose maps, (C series) relative dose differences, and (D series) the corresponding gamma maps with criteria of 3%/2.54 mm at gantry angles of 0°, 20°, 230°, 270°, and 300°
(A series) Total field transmitted dose maps, (B series) calculated transmitted dose maps, (C series) relative dose differences, and (D series) the corresponding gamma maps with criteria of 3%/2.54 mm at gantry angles of 0°, 20°, 230°, 270°, and 300°In order to provide more information on the agreement between measured and calculated transmitted dose distributions for each subfield and for the corresponding total field, the gamma scores for all evaluated subfields and total fields are given in [Table 2]. Due to the blurring of the EPIs, for subfields smaller than 8 cm2 (these are used to boost the radiation in the active tumor regions), several significant inconsistencies were observed.
Table 2
The agreement percentages between relative transmitted and the corresponding calculated dose maps (the gamma criteria were 3%/2.54 mm)
The agreement percentages between relative transmitted and the corresponding calculated dose maps (the gamma criteria were 3%/2.54 mm)The average agreement for these subfields was found to be less than 39.5%. The result for a typical small MLC field is shown in Figure 4 (see subfield 15). The relative dose difference values for regions outside the MLC fields were found to be greater than 3% of the dose delivered to the central part of MLC field, defined as the dose difference criterion for gamma assessment. Discarding the results for small fields, the maximum, minimum and average gamma score for all subfields were found to be 98.3%, 72.1% and 92.6%, respectively.As a result (the agreement between two series of consecutive EPIs is approximately 99%), so it can be said that for most fields, there is an excellent agreement between TPS and EPID transmitted dose distributions.In some subfields and consequently in the accumulated total fields, several significant differences were observed between measured and calculated transmitted dose maps inside the MLC fields and outside the patient anatomy. The relative dose differences in these regions were found to be greater than 6% for all subfields observed for two series of data sets acquired at A-P direction and at the gantry angle of 300°. The calculated dose values were found to be greater than those measured using SLIC-EPID in all these cases.The reasons for the differences between the calculated and measured transmitted dose maps were investigated. The first group of inconsistencies was for small fields (smaller than 8 cm2 at the isocenter). EPI blurring was found to be the main source of uncertainty. The second group of inconsistencies referred to several subfields for which not all parts of the radiation beam were incident on the phantom, but were passing through air. Although this type of discrepancy is irrelevant to the transmitted dose passing through the patient, the gamma score, for the MLC fields affected by this problem, decreases significantly. These two important issues are discussed in the following sections.
Evaluation of EPI blurring for small radiation fields
Typical radiation fluence maps for field sizes from 1 × 1 cm2 to 4 × 4 cm2 are shown in Figure 6 for an SED of 140 cm at Dmax. The x and y axes represent the number of pixels for all cases. A significant blurring of EPIs was observed for 1 × 1 and 2 × 2 cm2 radiation field sizes acquired both with and without the build-up layer. The absolute dose difference map between two averaged EPIs shows that the additional build-up layer increases uniformly the EPI pixel values in the blurred area, as is expected. Two horizontal and vertical strip bands, exactly the width of the radiation field size, were observed for radiation field sizes smaller than 9 cm2. The width of strip bands was found to be the same size as the radiation field size. This shows that the scanning of the row and column processing in the SLIC-EPID affects the pixel values. In addition, two effects are also discussed by Van Herk et al. for different image resolution in horizontal and vertical directions as follows: The 256-electrometer amplifiers including an appropriate filter mostly are known to be responsible for vertical blurring, and the lack of shielding between liquid ionization chambers is reported to be potentially responsible for spurious sensitivity outside the pixel area in the direction which is aligned with the electric filed lines. Figures 7 and 8 show that the blurring in the electric field direction (horizontal direction) is quite larger than in the vertical direction. It has also been shown the detective quantum efficiency (DQE) increases for ionization detectors with the decrease of the dose. If the radiation dose, delivered to SLIC-EPID ionization chambers, is originally averaged for each row or column reading, this justifies the probability of blurring that happened in SLIC-EPID during small field radiation dose deliveries.[4647]
Figure 6
The primary fluence maps with build-up layer (first row), without build-up layer (second row), and corresponding dose difference maps (third row) for small radiation field sizes for 1 ×1, 2 × 2, 3 × 3 and 4 × 4 cm2
Figure 7
The cross-plane profiles measured for several small fields (a) 1 × 1 cm2, (b) 2 × 2 cm2, (c) 3 × 3 cm2 and (d) 4 × 4 cm2, achieved from transmitted dose maps measured using SLIC-EPID and EDR2 films at a distance of 140 cm
Figure 8
Absolute transmitted dose maps of MLC fields with the jaws positioned away and close to the MLC field and the corresponding absolute dose difference map for small MLC fields;several small fields in the vicinity of a large irradiated area
The primary fluence maps with build-up layer (first row), without build-up layer (second row), and corresponding dose difference maps (third row) for small radiation field sizes for 1 ×1, 2 × 2, 3 × 3 and 4 × 4 cm2The cross-plane profiles measured for several small fields (a) 1 × 1 cm2, (b) 2 × 2 cm2, (c) 3 × 3 cm2 and (d) 4 × 4 cm2, achieved from transmitted dose maps measured using SLIC-EPID and EDR2 films at a distance of 140 cmAbsolute transmitted dose maps of MLC fields with the jaws positioned away and close to the MLC field and the corresponding absolute dose difference map for small MLC fields;several small fields in the vicinity of a large irradiated areaThe corresponding cross-plane profiles for small fields (from 1× 1 cm2 to 4×4 cm2), measured using SLIC-EPID with and without extra build-up layer and using EDR Kodak Film (EDR2), are also shown in [Figure 7]. EDR2 films have been reported to have a linear dose–response region extending to 500 cGy.[48] Results show that for 1 × 1 cm2 and 2 × 2 cm2 fields, SLIC-EPIDs are not able to create a reliable dose map. In contrast, with the increase of radiation field size, the data collected by SLIC-EPID match well with those collected using EDR2 films. The uncertainties at strip lines depend on the field size and increase with the increase of radiation field size.Due to the consistency between dose maps and the corresponding dose difference maps shown in Figure 6, as well as results shown in Figure 7, it was concluded that the additional build-up layer has no significant effect on the blurring of small field images. The additional build-up layer increases the number of free electrons reaching the EPID sensitive layer and the contribution of scattered photons.[49] More scattering increases the “fog phenomenon” in EPIs, and this decreases the sharpness of EPI. The horizontal and vertical strip bands show that the scanning of the row and column processing in the SLIC-EPID affects the pixel values. Random image blurring was observed in some of the small fields. However, since small fields are only used for boost dose delivery and their field size contribution is a small fraction (1% to 3%) of the total field size, this small field blurring had a negligible effect on overall image quality for the independent gantry angles used in this study. This is the main reason (or quantitative reason to say the inconsistencies are irrelevant for dose verification); however, this is reported as a drawback of SLIC-EPID.Typical results of conventional jaws positioning affecting the amount of MLC leakage on the blurring of EPIs acquired for small MLC fields are shown in [Figure 8]. In the first row of the figure, a small MLC field with three separated areas is shown. The conventional jaws were positioned far behind MLC leaf positioning (14 × 20 cm2) (first column) and also close to the opened MLC leaf positions (0.5 cm behind) (second column). The corresponding absolute dose difference is also displayed. As the image acquisition time is limited, no saturation is observed. Due to the blurring observed in the EPIs, three small fields cannot be distinguished. The absolute dose difference map shows that the impact of jaw positioning, and therefore intra- and inter-leaf MLC leaf leakage, does not either remove or reduce the blurring effect.Although the blurring effect is observed in EPIs acquired using SLIC-EPID for small radiation fields, in the presence of large irradiated areas, the small irradiated areas can be detected clearly. As the second row in Figure 8 shows, no blurring was observed for small MLC fields (less than 9 cm2) when there is a large irradiated area in the vicinity of small fields. In this case, no significant difference was observed for retracted and closed jaw positioning.SLIC-EPIDs, like other EPIDs, are generally developed to acquire an EPI for large open fields to verify patient positioning. It can be assumed, therefore, that the row and column scanning procedure as well as the supporting software do not process correctly the EPIs acquired for small radiation fields. No significant effect of the additional build-up layer and conventional jaws positioning was found on the blurred EPIs. The main reason for this is not exactly clear. The vertical and horizontal strips show that the supporting SLIC-EPID image construction including hardware and software could be responsible for EPI blurring because there is no radiation deposition in the off-axis areas.
SLIC-EPID response for high dose gradient regions
Several discrepancies between EPID measurements and TPS calculations were observed in the peripheral part of the MLC radiation fields. Typical gamma maps obtained from EPID and TPS comparisons for subfields of field 1 (AP direction), field 4 (gantry angle 270°) and field 5 (gantry angle of 300°) are shown in [Figure 9]. All of the significant disagreements were found to be located in the areas where beams were passing through air (outside the RANDO phantom). As all gamma maps show (series a), there is a good agreement for gamma maps in regions where the radiation beam passed through the patient/phantom. In contrast, significant discrepancies exist for radiation field sites where the beam is passing through the air. The corresponding relative dose difference maps (series b) show that the TPS relative dose values are greater than those measured using SLIC-EPID for disagreeing areas.
Figure 9
(a) The gamma and (b) the dose difference maps for typical radiation subfields at AP direction, 270° and 300º. The PASSED area for gamma criteria and dose difference within 3% for dose difference maps are shown in gray
(a) The gamma and (b) the dose difference maps for typical radiation subfields at AP direction, 270° and 300º. The PASSED area for gamma criteria and dose difference within 3% for dose difference maps are shown in grayIn order to find the reason for this discrepancy in the particular part of the gamma maps, two possibilities were investigated. Firstly, there was a concern about the calibration method used in this work, particularly for the off-axis correction performed using EDR2 films.[42,44] The second concern was about how the EPID or the TPS responded at the interface between inhomogeneities. It has been shown that in the presence of homogenous attenuators, there is a reasonable agreement between EPID and TPS dose values for large radiation fields.[43,44] The results prove that the calibration method used in this study guaranties sufficient accuracy when using the EPID dosimetry.The measured and calculated transmitted dose maps, corresponding relative dose difference, and gamma maps are indicated in [Figure 10] and the corresponding line profiles are shown in [Figure 11]. The results showed that after normalization to the central point located in the homogeneous phantom region, a significant discrepancy was observed between the measured and calculated transmitted dose maps in the area corresponding to the air region. In addition, the relative dose difference map showed that at all points where the beam portion passes through the air, the dose calculated using TPS is always greater than that measured using SLIC-EPID. Moreover, the area irradiated through the air for EPID measurements was found to be wider than those calculated using TPS.
Figure 10
(a and b) Relative measured and calculated dose for a 17 × 17 cm2 field size for a 2-cm shifted 10-cm homogenous attenuator in x and y axes directions; (c and d) the corresponding relative dose difference and gamma maps with criteria of 3%/2.54 mm
Figure 11
(a) Cross-plane and (b) inplane profiles of the measured and calculated relative dose maps, the corresponding relative dose difference and gamma map with 3%/2.54 mm criteria for a 17 × 17cm2 field size for a 2-cm shifted 10-cm homogenous attenuator
(a and b) Relative measured and calculated dose for a 17 × 17 cm2 field size for a 2-cm shifted 10-cm homogenous attenuator in x and y axes directions; (c and d) the corresponding relative dose difference and gamma maps with criteria of 3%/2.54 mm(a) Cross-plane and (b) inplane profiles of the measured and calculated relative dose maps, the corresponding relative dose difference and gamma map with 3%/2.54 mm criteria for a 17 × 17cm2 field size for a 2-cm shifted 10-cm homogenous attenuatorDiscrepancies between TPS and EPID dose profiles in the air areas can arise due to EPI construction. Two factors, SLIC-EPID hardware and data processing software, allow the EPID output to change. By using a 1mm liquid film (Isooctane, spectroscopical pure, Merck) as an ionization medium,[50] the SLIC-EPID ionization chambers are not quite independent and each pixel value can be affected by the neighboring pixels located in the surrounding region. The impact of software used to construct the EPI should also be taken into consideration.Several modifications including offset, dark field and flood field corrections, that are defined in the image acquisition software (PortalVision LC250, Varian Medical System Inc, Baden Switzerland), have been used to construct EPI. Moreover, several additional corrections such as filtering, linear correction and linear fix correction algorithms can be applied following the image acquisition.[51] These corrections control the undesired noise and help to smooth EPI pixel values. Through these procedures, it is expected that the EPID response would be lower values and larger areas than that calculated using TPS for the same conditions.The output of the TPS on the interface between air and phantom should also be taken into account. The TPS, Pinnacle3 (version 6.2) calculates the dose with the collapsed cone convolution superposition algorithm.[52,53] This algorithm is based on average density scaling method rather than a local density scaling. The scaling leads to the overdose/underdose calculation results compared to the measurements and Monte Carlo simulations.[54]
Conclusion
Several issues including EPI blurring and differences observed between calculated and measured transmitted dose maps, due to inaccurate beam modelling using treatment planning system and routine measurements uncertainties, decrease the accuracy of transmitted dose measurements. However, due to the small size of these radiation fields and their contribution, the impact of blurred EPIs on the total field evaluation is negligible. Moreover, the significant inconsistencies occurring between measured and calculated dose maps were shown to be located in regions of air transmission and are thus irrelevant to patient dose delivery verification.The dose delivery verification for a head and neck sIMRT case using SLIC-EPID was investigated. For large MLC fields, an excellent agreement was observed between transmitted dose measured using SLIC-EPID and that calculated by TPS (gamma score approximately 95%). For several small subfields, however, due to EPI blurring, a significant disagreement was found in the gamma results. Additionally, a significant disagreement between EPID and TPS dose maps was found in several parts of the radiation subfields, when the radiation beam passed through air.The transmitted dose distributions measured using portal imagers such as SLIC-EPID can be used to verify the dose delivery to the patient. However, several issues such as accurate calibration procedure and imager response under different conditions should be taken into consideration. For instance, the procedure of image construction using an SLIC-EPID reduces the quality of EPIs acquired for small fields. In addition, SLIC-EPID response differs from TPS response when a large inhomogeneity, e.g. a large air pocket, is positioned in the radiation beam path.
Authors: Wouter J C van Elmpt; Sebastiaan M J J G Nijsten; Robert F H Schiffeleers; André L A J Dekker; Ben J Mijnheer; Philippe Lambin; André W H Minken Journal: Med Phys Date: 2006-07 Impact factor: 4.071
Authors: Markus Wendling; Leah N McDermott; Anton Mans; Jan-Jakob Sonke; Marcel van Herk; Ben J Mijnheer Journal: Med Phys Date: 2009-07 Impact factor: 4.071
Authors: Tina Marie Briere; Sam Beddar; Peter Balter; Ravi Murthy; Sanjay Gupta; Christopher Nelson; George Starkschall; Michael T Gillin; Sunil Krishnan Journal: Phys Med Biol Date: 2009-05-13 Impact factor: 3.609
Authors: Hashemi S M; Bahreyni M H; Mohammadi M; Nasseri S; Bayani S; Gholamhosseinian H; Salek R; Shahedi F; Momennezhad M Journal: J Biomed Phys Eng Date: 2019-10-01