Two-dimensional transmitted dose measurements using a scanning liquid ionization chamber EPID.

Mohammad Mohammadi1, Eva Bezak.   

Abstract

The use of a scanning liquid ionization chamber electronic portal imaging device (SLIC-EPID) for two-dimensional transmitted dosimetry was investigated and a calibration method was developed using extended dose range (EDR2) film. In order to convert pixel value to dose, the acquired SLIC-EPID pixel values were calibrated using an ionization chamber on the central axis. The relationship between pixel values, dose rate and absorbed dose was identified for various linac output repetition rates. To correct EPIs for dosimetric purposes, the off-axis ratio of dose profiles measured by EPIDs and EDR2 film was used to derive correction factor matrices (CFMs) for a range of source-to-EPID distances (SEDs). The corrected relative dose maps acquired for different conditions, including open and wedged fields, measured using a SLIC-EPID were compared with EDR2 film images using a gamma function algorithm with distance to agreement (DTA) = 2.5 mm and dose difference (DeltaDmax) = 1% criteria. The results showed that (a) for two-dimensional dosimetric purposes, EPIDs must be calibrated using appropriate two-dimensional correction factors and (b) SLIC-EPIDs can be used to measure the transmitted dose with good accuracy.

Entities:  

Mesh:

Year:  2006        PMID: 16723778     DOI: 10.1088/0031-9155/51/11/019

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

1.  Modification of the gamma function for the recognition of over- and under-dose regions in three dimensions.

Authors:  Mohammad Mohammadi; Nima Rostampour; Thomas P Rutten
Journal:  J Med Phys       Date:  2012-10

2.  Evaluation of relative transmitted dose for a step and shoot head and neck intensity modulated radiation therapy using a scanning liquid ionization chamber electronic portal imaging device.

Authors:  Mohammad Mohammadi; Eva Bezak
Journal:  J Med Phys       Date:  2012-01

3.  Feasibility study on the verification of actual beam delivery in a treatment room using EPID transit dosimetry.

Authors:  Tae Seong Baek; Eun Ji Chung; Jaeman Son; Myonggeun Yoon
Journal:  Radiat Oncol       Date:  2014-12-04       Impact factor: 3.481

4.  A new approach for the pixel map sensitivity (PMS) evaluation of an electronic portal imaging device (EPID).

Authors:  Alberto Boriano; Francesco Lucio; Elisa Calamia; Elvio Russi; Flavio Marchetto
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

5.  The use of extended dose range film for dosimetric calibration of a scanning liquid-filled ionization chamber electronic portal imaging device.

Authors:  Mohammad Mohammadi; Eva Bezak; Paul Reich
Journal:  J Appl Clin Med Phys       Date:  2006-05-15       Impact factor: 2.102

6.  An Empirical Transmitted EPID Dosimetry Method using a Back-Projection Algorithm.

Authors:  Hashemi S M; Bahreyni M H; Mohammadi M; Nasseri S; Bayani S; Gholamhosseinian H; Salek R; Shahedi F; Momennezhad M
Journal:  J Biomed Phys Eng       Date:  2019-10-01
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.