Literature DB >> 7852125

The use of an electronic portal imaging device for exit dosimetry and quality control measurements.

M C Kirby1, P C Williams.   

Abstract

PURPOSE: To determine ways in which electronic portal imaging devices (EPIDs) could be used to (a) measure exit doses for external beam radiotherapy and (b) perform quality control checks on linear accelerators. METHODS AND MATERIALS: When imaging, our fluoroscopic EPID adjusts the gain, offset, and frame acquisition time of the charge coupled device (CCD) camera automatically, to allow for the range of photon transmissions through the patient, and to optimize the signal-to-noise ratio. However, our EPID can be programmed to act as an integrating dosemeter. EPID dosemeter measurements were made for 20 MV photons, for different field sizes and thicknesses of unit density phantom material placed at varying exit surface to detector distances. These were compared with simultaneous Silicon diode exit dose measurements. Our exit dosimetry technique was verified using an anthropomorphic type phantom, and some initial measurements have been made for patients treated with irregularly shaped 20 MV x-ray fields. In this dosimetry mode, our EPID was also used to measure certain quality control parameters, x-ray field flatness, and the verification of segmented intensity modulated field prescriptions.
RESULTS: Configured for dosimetry, our EPID exhibited a highly linear response, capable of resolving individual monitor units. Exit doses could be measured to within about 3% of that measured using Silicon diodes. Field flatness was determined to within 1.5% of Farmer dosemeter measurements. Segmented intensity modulated fields can be easily verified.
CONCLUSIONS: Our EPID has the versatility to assess a range of parameters pertinent to the delivery of high quality, high precision radiotherapy. When configured appropriately, it can measure exit doses in vivo, with reasonable accuracy, perform certain quick quality control checks, and analyze segmented intensity modulated treatment fields.

Entities:  

Mesh:

Year:  1995        PMID: 7852125     DOI: 10.1016/0360-3016(94)00388-2

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  7 in total

1.  Evaluation of relative transmitted dose for a step and shoot head and neck intensity modulated radiation therapy using a scanning liquid ionization chamber electronic portal imaging device.

Authors:  Mohammad Mohammadi; Eva Bezak
Journal:  J Med Phys       Date:  2012-01

2.  Guide to clinical use of electronic portal imaging.

Authors:  M G Herman; J J Kruse; C R Hagness
Journal:  J Appl Clin Med Phys       Date:  2000       Impact factor: 2.102

3.  Feasibility of using two-dimensional array dosimeter for in vivo dose reconstruction via transit dosimetry.

Authors:  Heeteak Chung; Jonathan Li; Sanjiv Samant
Journal:  J Appl Clin Med Phys       Date:  2011-04-08       Impact factor: 2.102

4.  Multi-parametric Improvements in the CCD Camera-based EPID for Portal Dosimetry.

Authors:  Vahideh Nazari; Seied R Mahdavi; Ahmad Mostaar; Hassan Nedaei; Mohammad A Mosleh Shirazi; Alireza Nikoofar; Golbarg Esmailie
Journal:  J Med Signals Sens       Date:  2017 Jan-Mar

5.  Dosimetric Validation of Digital Megavolt Imager for Flattening Filter Free Beams in the Pre-Treatment Quality Assurance of Stereotactic Body Radiation Therapy for Liver Metastases.

Authors:  Vendhan Subramani; Murali Rathakrishnan; Arunai Nambiraj N; Saraswathi Chitra S; Murali Venkatraman
Journal:  Asian Pac J Cancer Prev       Date:  2020-06-01

6.  The use of extended dose range film for dosimetric calibration of a scanning liquid-filled ionization chamber electronic portal imaging device.

Authors:  Mohammad Mohammadi; Eva Bezak; Paul Reich
Journal:  J Appl Clin Med Phys       Date:  2006-05-15       Impact factor: 2.102

7.  Evaluating the sensitivity of Halcyon's automatic transit image acquisition for treatment error detection: A phantom study using static IMRT.

Authors:  Xenia Ray; Casey Bojechko; Kevin L Moore
Journal:  J Appl Clin Med Phys       Date:  2019-10-06       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.