Relative profile and dose verification of intensity-modulated radiation therapy.

J Chang1, G S Mageras, C S Chui, C C Ling, W Lutz.   

Abstract

PURPOSE: To develop a quality assurance (QA) procedure to assess the intensity profile and dosimetry for intensity-modulated (IM) treatment fields using electronic portal imaging devices (EPIDs). METHODS AND MATERIALS: A series of rapidly acquired (approximately 1/sec) portal images are summed and converted to dose. For relative intensity QA, the intended profile is subtracted point-by-point from the measured profile forming a series of error values. The standard deviation, sigma, of the errors, a measure of the goodness of the match, is minimized by applying a normalization and uniform scatter subtraction from the measured profile. For dose verification (dose to isocenter), an empirically determined phantom-correction factor is added to incorporate the effect of patient presence on EPID readings. Seventy prostate treatment fields were used in a phantom study to verify these approaches. Sensitivity was studied by creating artificial mismatches.
RESULTS: The average sigma for relative profile verification is 3.3% (percentage of average intended intensity) whereas artificial mismatches resulted in sigma values from 5% to 27%. The average isocentric dose calculated from EPID readings is 1.001 relative to the planned dose with a standard deviation of 0.018.
CONCLUSIONS: An EPID can be used for profile verification and absolute isocentric dose measurement for IM fields.

Entities:  

Mesh:

Year:  2000        PMID: 10758329     DOI: 10.1016/s0360-3016(99)00555-6

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  7 in total

1.  The effect of influence quantities and detector orientation on small-field patient-specific IMRT QA: comparison of measurements with various ionization chambers.

Authors:  Henry Finlay Godson; Ravikumar Manickam; Sathiyan Saminathan; Kadirampatti Mani Ganesh; Retna Ponmalar
Journal:  Radiol Phys Technol       Date:  2016-12-01

2.  Evaluation of relative transmitted dose for a step and shoot head and neck intensity modulated radiation therapy using a scanning liquid ionization chamber electronic portal imaging device.

Authors:  Mohammad Mohammadi; Eva Bezak
Journal:  J Med Phys       Date:  2012-01

3.  A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction.

Authors:  Wendel Dean Renner; Kevin Norton; Timothy Holmes
Journal:  J Appl Clin Med Phys       Date:  2005-11-21       Impact factor: 2.102

4.  Feasibility of using two-dimensional array dosimeter for in vivo dose reconstruction via transit dosimetry.

Authors:  Heeteak Chung; Jonathan Li; Sanjiv Samant
Journal:  J Appl Clin Med Phys       Date:  2011-04-08       Impact factor: 2.102

5.  A new approach for the pixel map sensitivity (PMS) evaluation of an electronic portal imaging device (EPID).

Authors:  Alberto Boriano; Francesco Lucio; Elisa Calamia; Elvio Russi; Flavio Marchetto
Journal:  J Appl Clin Med Phys       Date:  2013-11-04       Impact factor: 2.102

6.  Using the frame averaging of aS500 EPID for IMRT verification.

Authors:  J Chang; C C Ling
Journal:  J Appl Clin Med Phys       Date:  2003       Impact factor: 2.102

7.  An empirical model of electronic portal imager response implemented within a commercial treatment planning system for verification of intensity-modulated radiation therapy fields.

Authors:  Rao F H Khan; Orest Z Ostapiak; Joe J Szabo
Journal:  J Appl Clin Med Phys       Date:  2008-11-11       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.