Arterial pressure waveform derived cardiac output FloTrac/Vigileo system (third generation software): comparison of two monitoring sites with the thermodilution cardiac output.

Sumit Vasdev1, Sandeep Chauhan, Minati Choudhury, Millind P Hote, Madhur Malik, Usha Kiran.   

Abstract

The present study was conducted to study the effect of monitoring site, radial or femoral, for arterial pressure waveform derived cardiac output using FloTrac/Vigileo system with third generation software version 3.02 during cardiac surgery. The cardiac output derived from the two sites was also compared to the pulmonary artery catheter (PAC) derived cardiac output to reevaluate the relation between them using the newer software. The effect of cardiopulmonary bypass (CPB) was also studied by doing the sub analysis before and after bypass. Forty patients undergoing coronary artery bypass surgery with cardiopulmonary bypass were enrolled in the study. Cardiac output derived from radial artery (RADCO), femoral artery (FEMCO) using FloTrac/Vigileo system with third generation software version 3.02 and cardiac output using pulmonary artery catheter (PACCO) at predefined nine time points were recorded. Three hundred and forty two cardiac output data triplets were analysed. The Bland-Altman analysis of RADCO and FEMCO revealed a mean bias of -0.28 with percentage error of 20%. The pre CPB precision of both RADCO and FEMCO was 1.25 times as that of PACCO. The post CPB precision of FEMCO was 1.2 times of PACCO while that of RADCO was 1.7 times of PACCO. The third generation of FloTrac/Vigileo system shows good correlation between the radial and femoral derived cardiac outputs in both pre and post bypass periods. The newer software correlates better to PAC derived cardiac output in the post bypass period for femoral artery than radial artery.

Entities:  

Mesh:

Year:  2012        PMID: 22350311     DOI: 10.1007/s10877-012-9341-5

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  20 in total

Review 1.  Minimally invasive hemodynamic monitoring for the intensivist: current and emerging technology.

Authors:  John C Chaney; Stephen Derdak
Journal:  Crit Care Med       Date:  2002-10       Impact factor: 7.598

2.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

Review 3.  Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision.

Authors:  Philip J Peyton; Simon W Chong
Journal:  Anesthesiology       Date:  2010-11       Impact factor: 7.892

4.  The pressure dependent dynamic elasticity of 35 thoracic and 16 abdominal human aortas in vitro described by a five component model.

Authors:  G J Langewouters; K H Wesseling; W J Goedhard
Journal:  J Biomech       Date:  1985       Impact factor: 2.712

5.  Increased risk associated with pulmonary artery catheterization in the medical intensive care unit.

Authors:  Steve G Peters; Bekele Afessa; Paul A Decker; Darrell R Schroeder; Kenneth P Offord; John P Scott
Journal:  J Crit Care       Date:  2003-09       Impact factor: 3.425

6.  Complications of right heart catheterization. A prospective autopsy study.

Authors:  A F Connors; R J Castele; N Z Farhat; J F Tomashefski
Journal:  Chest       Date:  1985-10       Impact factor: 9.410

7.  Cardiac output determination from the arterial pressure wave: clinical testing of a novel algorithm that does not require calibration.

Authors:  Gerard R Manecke; William R Auger
Journal:  J Cardiothorac Vasc Anesth       Date:  2006-10-24       Impact factor: 2.628

8.  Validity of an arterial pressure waveform analysis device: does the puncture site play a role in the agreement with intermittent pulmonary artery catheter thermodilution measurements?

Authors:  Sebastian Schramm; Eric Albrecht; Philippe Frascarolo; Pierre-Guy Chassot; Donat R Spahn
Journal:  J Cardiothorac Vasc Anesth       Date:  2009-08-22       Impact factor: 2.628

9.  Uncalibrated arterial pulse contour analysis versus continuous thermodilution technique: effects of alterations in arterial waveform.

Authors:  Suraphong Lorsomradee; Sratwadee Lorsomradee; Stefanie Cromheecke; Stefan G De Hert
Journal:  J Cardiothorac Vasc Anesth       Date:  2007-05-07       Impact factor: 2.628

10.  Evaluation of a new arterial pressure-based cardiac output device requiring no external calibration.

Authors:  Christopher Prasser; Sylvia Bele; Cornelius Keyl; Stefan Schweiger; Benedikt Trabold; Matthias Amann; Julia Welnhofer; Christoph Wiesenack
Journal:  BMC Anesthesiol       Date:  2007-11-09       Impact factor: 2.217

View more
  6 in total

1.  Comparison of cardiac output measures by transpulmonary thermodilution, pulse contour analysis, and pulmonary artery thermodilution during off-pump coronary artery bypass surgery: a subgroup analysis of the cardiovascular anaesthesia registry at a single tertiary centre.

Authors:  Youn Joung Cho; Chang-Hoon Koo; Tae Kyong Kim; Deok Man Hong; Yunseok Jeon
Journal:  J Clin Monit Comput       Date:  2015-10-01       Impact factor: 2.502

2.  Agreement in hemodynamic monitoring during orthotopic liver transplantation: a comparison of FloTrac/Vigileo at two monitoring sites with pulmonary artery catheter thermodilution.

Authors:  Matthew Lee; Laurence Weinberg; Brett Pearce; Nicholas Scurrah; David A Story; Param Pillai; Peter R McCall; Larry P McNicol; Philip J Peyton
Journal:  J Clin Monit Comput       Date:  2016-02-16       Impact factor: 2.502

Review 3.  Minimally invasive or noninvasive cardiac output measurement: an update.

Authors:  Lisa Sangkum; Geoffrey L Liu; Ling Yu; Hong Yan; Alan D Kaye; Henry Liu
Journal:  J Anesth       Date:  2016-03-09       Impact factor: 2.078

Review 4.  Accuracy of non-invasive and minimally invasive hemodynamic monitoring: where do we stand?

Authors:  Issa Pour-Ghaz; Theodore Manolukas; Nathalie Foray; Joel Raja; Aranyak Rawal; Uzoma N Ibebuogu; Rami N Khouzam
Journal:  Ann Transl Med       Date:  2019-09

5.  The value of arterial pressure waveform cardiac output measurements in the radial and femoral artery in major cardiac surgery patients.

Authors:  A van Drumpt; J van Bommel; S Hoeks; F Grüne; T Wolvetang; J Bekkers; M Ter Horst
Journal:  BMC Anesthesiol       Date:  2017-03-14       Impact factor: 2.217

6.  Performance of Third-generation FloTrac/Vigileo system during hyperdynamic therapy for delayed cerebral ischemia after subarachnoid hemorrhage.

Authors:  Tatsushi Mutoh; Tatsuya Ishikawa; Shinya Kobayashi; Akifumi Suzuki; Nobuyuki Yasui
Journal:  Surg Neurol Int       Date:  2012-08-27
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.