| Literature DB >> 23061015 |
Tatsushi Mutoh1, Tatsuya Ishikawa, Shinya Kobayashi, Akifumi Suzuki, Nobuyuki Yasui.
Abstract
BACKGROUND: Monitoring of cardiac output (CO) is important for promising safe approach to goal-directed hemodynamic therapy for delayed cerebral ischemia (DCI) after subarachnoid hemorrhage (SAH), but is often precluded by the invasiveness and complexity of ongoing monitoring modalities. We examined the clinical utility of less-invasive management using an uncalibrated arterial pressure waveform-derived cardiac output (APCO) monitor with refined algorithm (Third-generation FloTrac/Vigileo, Edwards, Irvine, CA, USA) during hyperdynamic therapy for post-SAH DCI, compared with transpulmonary thermodilution (PiCCO, Pulsion, Munich, Germany) as a reference technique.Entities:
Keywords: Cardiac output; hemodynamic monitoring; pulse contour analysis; subarachnoid hemorrhage; transpulmonary thermodilution
Year: 2012 PMID: 23061015 PMCID: PMC3463836 DOI: 10.4103/2152-7806.100195
Source DB: PubMed Journal: Surg Neurol Int ISSN: 2152-7806
Figure 1Relationship between cardiac index (CI) determined by the FloTrac/Vigileo system and reference transpulmonary thermodilution for 20 SAH patients. APCI, arterial pressure-based pulse contour CI analyzed by the FloTrac/Vigileo system; TPCI, transpulmonary thermodilution CI determined by the PiCCO system. (a) Least-squares regression line (solid line) and the line of identity (dotted line); (b) Bland–Altman plot of bias (solid line) and precision (dotted lines)
Figure 2Subgroup analyses of cardiac index (CI) determined by the FloTrac/Vigileo system and reference transpulmonary thermodilution at different doses of dobutamine during hyperdynamic therapy in 20 SAH patients. (a) Changes of CI in response to dobutamine dose increments. Values (mean ± SD) measured before (D0) and 60 min after dobutamine infusion at each increment of dose (D1, 3 μg/kg/min; D2, 6 μg/kg/min; D3, 9 μg/ kg/min; D4, 12 μg/kg/min; and D5, 15 μg/kg/min). (b) Bland–Altman plot of bias and precision analyzed between FloTrac/Vigileo system and reference transpulmonary thermodilution
Hemodynamic data for 20 subarachnoid hemorrhage patients obtained simultaneously from two hemodynamic devices during hyperdynamic therapy
Clinical data for 48 SAH patients undergoing hemodynamic management assisted with two devices