Improved B1 homogeneity of 3 Tesla breast MRI using dual-source parallel radiofrequency excitation.

Habib Rahbar1, Savannah C Partridge, Wendy B Demartini, Robert L Gutierrez, Sana Parsian, Constance D Lehman.   

Abstract

PURPOSE: To compare breast MRI B(1) homogeneity at 3 Tesla (T) with and without dual-source parallel radiofrequency (RF) excitation.
MATERIALS AND METHODS: After institutional review board approval, we evaluated 14 consecutive breast MR examinations performed at 3T that included three-dimensional B(1) maps created separately with conventional single-source and dual-source parallel RF excitation techniques. We measured B(1) values (expressed as % of intended B(1) ) on each B(1) map at nipple level in multiple bilateral locations: anterior, lateral, central, medial, and posterior. Mean whole breast and location specific B(1) values were calculated and compared between right and left breasts using paired t-test.
RESULTS: Mean whole breast B(1) values differed significantly between right and left breasts with standard single-source RF excitation (difference L-R, Δ = 9.2%; P < 0.001) but not with dual-source parallel RF excitation (Δ = 2.3%; P = 0.085). Location specific B(1) values differed significantly between right and left on single-source in the lateral (P = 0.014), central (P = 0.0001), medial (P = 0.0013), and posterior (P < 0.0001) locations. Conversely, mean B(1) values differed significantly on dual-source parallel RF excitation for only the anterior (P = 0.030) and lateral (P = 0.0003) locations.
CONCLUSION: B(1) homogeneity is improved with dual-source parallel RF excitation on 3T breast MRI when compared with standard single-source RF excitation technique.
Copyright © 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2012        PMID: 22282269      PMCID: PMC3326217          DOI: 10.1002/jmri.23571

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  8 in total

1.  Dual-source parallel radiofrequency excitation body MR imaging compared with standard MR imaging at 3.0 T: initial clinical experience.

Authors:  Winfried A Willinek; Jürgen Gieseke; Guido M Kukuk; Michael Nelles; Roy König; Nuschin Morakkabati-Spitz; Frank Träber; Daniel Thomas; Christiane K Kuhl; Hans H Schild
Journal:  Radiology       Date:  2010-09       Impact factor: 11.105

2.  Experimental analysis of parallel excitation using dedicated coil setups and simultaneous RF transmission on multiple channels.

Authors:  Peter Ullmann; Sven Junge; Markus Wick; Frank Seifert; Wolfgang Ruhm; Jürgen Hennig
Journal:  Magn Reson Med       Date:  2005-10       Impact factor: 4.668

3.  Actual flip-angle imaging in the pulsed steady state: a method for rapid three-dimensional mapping of the transmitted radiofrequency field.

Authors:  Vasily L Yarnykh
Journal:  Magn Reson Med       Date:  2007-01       Impact factor: 4.668

Review 4.  Parallel RF transmission in MRI.

Authors:  Ulrich Katscher; Peter Börnert
Journal:  NMR Biomed       Date:  2006-05       Impact factor: 4.044

5.  Eight-channel transmit/receive body MRI coil at 3T.

Authors:  P Vernickel; P Röschmann; C Findeklee; K-M Lüdeke; Ch Leussler; J Overweg; U Katscher; I Grässlin; K Schünemann
Journal:  Magn Reson Med       Date:  2007-08       Impact factor: 4.668

6.  Effect of B1 inhomogeneity on breast MR imaging at 3.0 T.

Authors:  Christiane K Kuhl; Hendrik Kooijman; Juergen Gieseke; Hans H Schild
Journal:  Radiology       Date:  2007-09       Impact factor: 11.105

7.  B1 transmission-field inhomogeneity and enhancement ratio errors in dynamic contrast-enhanced MRI (DCE-MRI) of the breast at 3T.

Authors:  Che A Azlan; Pierluigi Di Giovanni; Trevor S Ahearn; Scott I K Semple; Fiona J Gilbert; Thomas W Redpath
Journal:  J Magn Reson Imaging       Date:  2010-01       Impact factor: 4.813

8.  Dual-source parallel RF transmission for clinical MR imaging of the spine at 3.0 T: intraindividual comparison with conventional single-source transmission.

Authors:  Michael Nelles; Roy S König; Jürgen Gieseke; Marjolijn M Guerand-van Battum; Guido M Kukuk; Hans H Schild; Winfried A Willinek
Journal:  Radiology       Date:  2010-09-21       Impact factor: 11.105

  8 in total
  16 in total

1.  Fat suppression techniques (STIR vs. SPAIR) on diffusion-weighted imaging of breast lesions at 3.0 T: preliminary experience.

Authors:  Sofia Brandão; Luísa Nogueira; Eduarda Matos; Rita Gouveia Nunes; Hugo Alexandre Ferreira; Joana Loureiro; Isabel Ramos
Journal:  Radiol Med       Date:  2015-02-10       Impact factor: 3.469

2.  Evaluation of Kinetic Entropy of Breast Masses Initially Found on MRI using Whole-lesion Curve Distribution Data: Comparison with the Standard Kinetic Analysis.

Authors:  Akiko Shimauchi; Hiroyuki Abe; David V Schacht; Jian Yulei; Federico D Pineda; Sanaz A Jansen; Rajiv Ganesh; Gillian M Newstead
Journal:  Eur Radiol       Date:  2015-02-20       Impact factor: 5.315

Review 3.  Sequential whole-body PET/MR scanner: concept, clinical use, and optimisation after two years in the clinic. The manufacturer's perspective.

Authors:  Antonis Kalemis; Bénédicte M A Delattre; Susanne Heinzer
Journal:  MAGMA       Date:  2012-08-07       Impact factor: 2.310

4.  Bilateral breast MRI by use of dual-source parallel radiofrequency excitation and image-based shimming at 3 Tesla: improvement in homogeneity on fat-suppression imaging.

Authors:  Kinya Ishizaka; Fumi Kato; Satoshi Terae; Suzuko Mito; Noriko Oyama-Manabe; Tamotsu Kamishima; Mitsuhiro Nakanishi; Hiroyuki Sugimori; Hiroyuki Hamaguchi; Hiroki Shirato
Journal:  Radiol Phys Technol       Date:  2014-06-26

5.  Quadrature transmit array design using single-feed circularly polarized patch antenna for parallel transmission in MR imaging.

Authors:  Yong Pang; Baiying Yu; Daniel B Vigneron; Xiaoliang Zhang
Journal:  Quant Imaging Med Surg       Date:  2014-02

6.  A highly decoupled transmit-receive array design with triangular elements at 7T.

Authors:  Gang Chen; Bei Zhang; Martijn A Cloos; Daniel K Sodickson; Graham C Wiggins
Journal:  Magn Reson Med       Date:  2018-03-23       Impact factor: 4.668

7.  A dual-tuned multichannel bilateral RF coil for 1 H/23 Na breast MRI at 7 T.

Authors:  Carlotta Ianniello; Guillaume Madelin; Linda Moy; Ryan Brown
Journal:  Magn Reson Med       Date:  2019-05-31       Impact factor: 4.668

8.  Transmit B1+ field inhomogeneity and T1 estimation errors in breast DCE-MRI at 3 tesla.

Authors:  Kyunghyun Sung; Bruce L Daniel; Brian A Hargreaves
Journal:  J Magn Reson Imaging       Date:  2013-01-04       Impact factor: 4.813

9.  B1 and T1 mapping of the breast with a reference tissue method.

Authors:  Federico D Pineda; Milica Medved; Xiaobing Fan; Gregory S Karczmar
Journal:  Magn Reson Med       Date:  2015-05-27       Impact factor: 4.668

10.  GRASE Revisited: breath-hold three-dimensional (3D) magnetic resonance cholangiopancreatography using a Gradient and Spin Echo (GRASE) technique at 3T.

Authors:  Ju Gang Nam; Jeong Min Lee; Hyo-Jin Kang; Sang Min Lee; Eunju Kim; Johannes M Peeters; Jeong Hee Yoon
Journal:  Eur Radiol       Date:  2018-02-01       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.