Dual-source parallel RF transmission for clinical MR imaging of the spine at 3.0 T: intraindividual comparison with conventional single-source transmission.

Michael Nelles1, Roy S König, Jürgen Gieseke, Marjolijn M Guerand-van Battum, Guido M Kukuk, Hans H Schild, Winfried A Willinek.   

Abstract

PURPOSE: To prospectively and intraindividually compare single-source radiofrequency (RF) excitation and dual-source parallel RF excitation in 3.0-T magnetic resonance (MR) imaging of the spine.
MATERIALS AND METHODS: Institutional review board approval and written informed patient consent were obtained. The RF power of a 3.0-T MR imaging system was distributed to two ports of the body coil of the system by using independent RF transmit channels. The maximum B(1) field strength for dual-source parallel RF excitation was maintained, as compared with single-source RF excitation. The repetition time was reduced according to the revised RF setup with dual-source parallel RF excitation while maintaining specific energy absorption limitations. Thirty patients were examined with and without dual-source parallel RF excitation. Diagnostic quality was assessed independently by two radiologists according to a four-point grading system. Image contrast ratios (CRs) were calculated between reference tissues and vertebrae for single-source RF excitation and dual-source parallel RF excitation.
RESULTS: The mean acceleration achieved with dual-source parallel RF excitation was 36% (range, 18%-50%). The total imaging duration of a three-station total spinal examination was reduced by one-third by using dual-source parallel RF transmission. For all cases investigated, diagnostic image quality without significant differences between the two methods and with a good interobserver agreement was achieved (Kendall tau-b, 0.50-0.84). The observed image contrast changes were predominantly small (<0.10 in 15 of 24 CRs), though they were significantly different (P < .05).
CONCLUSION: While shortening examination times by approximately one-third, the dual-source parallel RF transmission mode in MR imaging of the spine yielded diagnostic image quality comparable to that with the conventional single-source RF transmission mode. © RSNA, 2010.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20858848     DOI: 10.1148/radiol.10092146

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  21 in total

1.  Improved B1 homogeneity of 3 Tesla breast MRI using dual-source parallel radiofrequency excitation.

Authors:  Habib Rahbar; Savannah C Partridge; Wendy B Demartini; Robert L Gutierrez; Sana Parsian; Constance D Lehman
Journal:  J Magn Reson Imaging       Date:  2012-01-26       Impact factor: 4.813

Review 2.  Sequential whole-body PET/MR scanner: concept, clinical use, and optimisation after two years in the clinic. The manufacturer's perspective.

Authors:  Antonis Kalemis; Bénédicte M A Delattre; Susanne Heinzer
Journal:  MAGMA       Date:  2012-08-07       Impact factor: 2.310

3.  Field shaping arrays: a means to address shading in high field breast MRI.

Authors:  Ileana Hancu; Seung-Kyun Lee; W Thomas Dixon; Laura Sacolick; Ricardo Becerra; Zhenghui Zhang; Graeme McKinnon; Vijayanand Alagappan
Journal:  J Magn Reson Imaging       Date:  2012-06-21       Impact factor: 4.813

4.  Bilateral breast MRI by use of dual-source parallel radiofrequency excitation and image-based shimming at 3 Tesla: improvement in homogeneity on fat-suppression imaging.

Authors:  Kinya Ishizaka; Fumi Kato; Satoshi Terae; Suzuko Mito; Noriko Oyama-Manabe; Tamotsu Kamishima; Mitsuhiro Nakanishi; Hiroyuki Sugimori; Hiroyuki Hamaguchi; Hiroki Shirato
Journal:  Radiol Phys Technol       Date:  2014-06-26

Review 5.  A half-century of innovation in technology-preparing MRI for the 21st century.

Authors:  Peter Börnert; David G Norris
Journal:  Br J Radiol       Date:  2020-06-15       Impact factor: 3.039

6.  Cardiac MOLLI T1 mapping at 3.0 T: comparison of patient-adaptive dual-source RF and conventional RF transmission.

Authors:  Michael Rasper; Jonathan Nadjiri; Alexandra S Sträter; Marcus Settles; Karl-Ludwig Laugwitz; Ernst J Rummeny; Armin M Huber
Journal:  Int J Cardiovasc Imaging       Date:  2017-01-30       Impact factor: 2.357

7.  Late gadolinium enhancement cardiac imaging on a 3T scanner with parallel RF transmission technique: prospective comparison of 3D-PSIR and 3D-IR.

Authors:  Anthony Schultz; Thibault Caspar; Mickaël Schaeffer; Aïssam Labani; Mi-Young Jeung; Soraya El Ghannudi; Catherine Roy; Mickaël Ohana
Journal:  Eur Radiol       Date:  2015-09-15       Impact factor: 5.315

8.  Parallel imaging of the cervical spine at 3T: optimized trade-off between speed and image quality.

Authors:  J Fruehwald-Pallamar; P Szomolanyi; N Fakhrai; A Lunzer; M Weber; M M Thurnher; M Pallamar; S Trattnig; D Prayer; I M Noebauer-Huhmann
Journal:  AJNR Am J Neuroradiol       Date:  2012-05-03       Impact factor: 3.825

9.  Zoomed EPI DWI of Acute Spinal Ischemia Using a Parallel Transmission System.

Authors:  A Seeger; U Klose; F Bischof; J Strobel; U Ernemann; T-K Hauser
Journal:  Clin Neuroradiol       Date:  2014-08-29       Impact factor: 3.649

10.  3D non-contrast-enhanced ECG-gated MR angiography of the lower extremities with dual-source radiofrequency transmission at 3.0 T: Intraindividual comparison with contrast-enhanced MR angiography in PAOD patients.

Authors:  Michael Rasper; Moritz Wildgruber; Marcus Settles; Hans-Henning Eckstein; Alexander Zimmermann; Christian Reeps; Ernst J Rummeny; Armin M Huber
Journal:  Eur Radiol       Date:  2015-11-11       Impact factor: 5.315

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.