Constrained cups appear incapable of meeting the demands of revision THA.

Philip C Noble1, Salim K Durrani, Molly M Usrey, Kenneth B Mathis, Nikolaos V Bardakos.   

Abstract

BACKGROUND: Failure rates of constrained cups for treating recurrent dislocation in revision THA range from 40% to 100%. Although constrained liners are intended to stabilize the hip by mechanically preventing dislocation, the resulting loss of range of motion may lead to impingement and, ultimately, implant failure. QUESTIONS/PURPOSES: We therefore documented the mechanisms of failure of constrained acetabular cups in revision THA and determined the type and severity of damage (wear, fracture, and impingement) that occurs in situ.
METHODS: We retrieved 57 constrained components of four different designs at revision THA and examined for the presence of rim impingement, oxidation, cracks within the liner, backside wear, pitting, scratching, abrasion, burnishing, and the presence of embedded particles. Articular wear was calculated from the volume of the concave articular bearing surface, which was measured using the fluid displacement method.
RESULTS: Failure of the locking ring was responsible for 51% of failures, whereas 28% of revisions were the result of acetabular cup loosening, 6% backside wear, and 22% infection. Impingement damage of the rim of the polyethylene liner was seen in all retrievals with moderate or severe damage in 54%. The average volumetric wear rate of the articular surface was 95 mm(3)/year.
CONCLUSIONS: Failure of the locking liner ring and loosening of the acetabular cup are the primary causes of mechanical failure with constrained liners; polyethylene is an inadequate material for restricting motion of the hip to prevent instability. The durability of these devices is unlikely to improve unless the mechanical demands are modified through increased range of motion leading to less frequent rim impingement.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22179979      PMCID: PMC3369098          DOI: 10.1007/s11999-011-2212-4

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  50 in total

1.  Cracking and impingement in ultra-high-molecular-weight polyethylene acetabular liners.

Authors:  Michael V Birman; Philip C Noble; Michael A Conditt; Steve Li; Kenneth B Mathis
Journal:  J Arthroplasty       Date:  2005-10       Impact factor: 4.757

2.  Impingement in total hip arthroplasty a study of retrieved acetabular components.

Authors:  Won Yong Shon; Todd Baldini; Margaret G Peterson; Timothy M Wright; Eduardo A Salvati
Journal:  J Arthroplasty       Date:  2005-06       Impact factor: 4.757

3.  Constrained acetabular liners: mechanisms of failure.

Authors:  Andrew G Yun; Douglas Padgett; Paul Pellicci; Lawrence D Dorr
Journal:  J Arthroplasty       Date:  2005-06       Impact factor: 4.757

4.  Prevalence of impingement in explanted Charnley acetabular components.

Authors:  R M Hall; P Siney; A Unsworth; B M Wroblewski
Journal:  J Orthop Sci       Date:  1998       Impact factor: 1.601

Review 5.  Wear in total hip and knee replacements.

Authors:  T P Schmalzried; J J Callaghan
Journal:  J Bone Joint Surg Am       Date:  1999-01       Impact factor: 5.284

6.  Difficult complications after hip joint replacement. Dislocation.

Authors:  B F Morrey
Journal:  Clin Orthop Relat Res       Date:  1997-11       Impact factor: 4.176

7.  Failure of a constrained acetabular prosthesis of a total hip arthroplasty. A report of four cases.

Authors:  B P Kaper; P M Bernini
Journal:  J Bone Joint Surg Am       Date:  1998-04       Impact factor: 5.284

8.  Wear of polyethylene acetabular components in total hip arthroplasty. An analysis of one hundred and twenty-eight components retrieved at autopsy or revision operations.

Authors:  M Jasty; D D Goetz; C R Bragdon; K R Lee; A E Hanson; J R Elder; W H Harris
Journal:  J Bone Joint Surg Am       Date:  1997-03       Impact factor: 5.284

9.  Salvage of a recurrently dislocating total hip prosthesis with use of a constrained acetabular component. A retrospective analysis of fifty-six cases.

Authors:  D D Goetz; W N Capello; J J Callaghan; T D Brown; R C Johnston
Journal:  J Bone Joint Surg Am       Date:  1998-04       Impact factor: 5.284

10.  Constrained acetabular components.

Authors:  M J Anderson; W R Murray; H B Skinner
Journal:  J Arthroplasty       Date:  1994-02       Impact factor: 4.757

View more
  17 in total

1.  Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls.

Authors:  Mitchell C Weiser; Darwin D Chen
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

2.  [Larger heads compensate for an increased risk of THA dislocation in high-risk patients].

Authors:  D Zajonz; H Philipp; S Schleifenbaum; R Möbius; N Hammer; R Grunert; T Prietzel
Journal:  Orthopade       Date:  2015-05       Impact factor: 1.087

3.  Dual mobility cups in revision total hip arthroplasty.

Authors:  Anthony Viste; Romain Desmarchelier; Michel-Henri Fessy
Journal:  Int Orthop       Date:  2016-12-08       Impact factor: 3.075

Review 4.  [Hip dislocation after revision arthroplasty : Risk assessment and treatment strategies].

Authors:  P M Prodinger; J Schauwecker; H Mühlhofer; N Harrasser; F Pohlig; C Suren; R von Eisenhart-Rothe
Journal:  Orthopade       Date:  2017-02       Impact factor: 1.087

5.  The dual mobility cup in muscular skeletal oncology: rationale and indications.

Authors:  Carmine Zoccali; Dario Attala; Alessandra Scotto di Uccio; Barbara Rossi; Gennaro Scotto; Roberto Biagini
Journal:  Int Orthop       Date:  2017-01-31       Impact factor: 3.075

6.  A dual-mobility cup reduces risk of dislocation in isolated acetabular revisions.

Authors:  Roberto Civinini; Christian Carulli; Fabrizio Matassi; Lorenzo Nistri; Massimo Innocenti
Journal:  Clin Orthop Relat Res       Date:  2012-06-15       Impact factor: 4.176

Review 7.  Dislocation following total hip replacement.

Authors:  Jens Dargel; Johannes Oppermann; Gert-Peter Brüggemann; Peer Eysel
Journal:  Dtsch Arztebl Int       Date:  2014-12-22       Impact factor: 5.594

Review 8.  Total hip arthroplasty instability in Italy.

Authors:  Francesco Falez; Matteo Papalia; Fabio Favetti; Gabriele Panegrossi; Filippo Casella; Gianluca Mazzotta
Journal:  Int Orthop       Date:  2016-12-20       Impact factor: 3.075

9.  Acetabular liner with focal constraint to prevent dislocation after THA.

Authors:  Jacob T Munro; Mihai H Vioreanu; Bassam A Masri; Clive P Duncan
Journal:  Clin Orthop Relat Res       Date:  2013-12       Impact factor: 4.176

10.  High complication rate after revision of large-head metal-on-metal total hip arthroplasty.

Authors:  Jacob T Munro; Bassam A Masri; Clive P Duncan; Donald S Garbuz
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.