Literature DB >> 23423623

Acetabular liner with focal constraint to prevent dislocation after THA.

Jacob T Munro1, Mihai H Vioreanu, Bassam A Masri, Clive P Duncan.   

Abstract

BACKGROUND: Dislocation continues to commonly cause failure after primary and revision total hip arthroplasty (THA). Fully constrained liners intended to prevent dislocation are nonetheless associated with a substantial incidence of failure by redislocation, mechanical failure, aseptic loosening, or a combination. Constrained liners with cutouts of the elevated rims can theoretically increase range of movement and therefore decrease the risk dislocation, but it is unclear if they do so in practice and whether they are associated with early wear or loosening. QUESTIONS/PURPOSES: We therefore determined (1) occurrence or recurrence of dislocation and (2) rate of complications associated with constrained implants with cutouts; and (3) assessed for early cup loosening.
METHODS: We retrospectively reviewed the records of 81 patients at high risk for dislocation who had 82 constrained liners inserted for primary (n = 10) or revision (n = 72) THA between 2008 and 2010. From the records we extracted demographic and implant data and instances of recurrent dislocation, implant failure, osteolysis, loosening, or construct failure. The minimum followup was 24 months (mean, 34 months; range, 24-49 months).
RESULTS: Three liners failed as a result of further dislocation (3%). Three deep infections occurred. One patient had progressive loosening at the shell-bone interface.
CONCLUSIONS: Our observations suggest this liner is associated with a relatively low risk of dislocation in patients at high risk for dislocation and those with recurrent dislocation.

Entities:  

Mesh:

Year:  2013        PMID: 23423623      PMCID: PMC3825877          DOI: 10.1007/s11999-013-2858-1

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  36 in total

1.  Use of constrained acetabular components for hip instability: an average 10-year follow-up study.

Authors:  Barron R B Bremner; Devon D Goetz; John J Callaghan; William N Capello; Richard C Johnston
Journal:  J Arthroplasty       Date:  2003-10       Impact factor: 4.757

2.  The Frank Stinchfield Award: Dislocation in revision THA: do large heads (36 and 40 mm) result in reduced dislocation rates in a randomized clinical trial?

Authors:  Donald S Garbuz; Bassam A Masri; Clive P Duncan; Nelson V Greidanus; Eric R Bohm; Martin J Petrak; Craig J Della Valle; Allan E Gross
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

3.  Constrained cups appear incapable of meeting the demands of revision THA.

Authors:  Philip C Noble; Salim K Durrani; Molly M Usrey; Kenneth B Mathis; Nikolaos V Bardakos
Journal:  Clin Orthop Relat Res       Date:  2012-07       Impact factor: 4.176

4.  Outcome and complications of constrained acetabular components.

Authors:  Cao Yang; Stuart B Goodman
Journal:  Orthopedics       Date:  2009-02       Impact factor: 1.390

5.  Modes of failure of Osteonics constrained tripolar implants: a retrospective analysis of forty-three failed implants.

Authors:  Olivier Guyen; David G Lewallen; Miguel E Cabanela
Journal:  J Bone Joint Surg Am       Date:  2008-07       Impact factor: 5.284

6.  Revision for recurrent instability: what are the predictors of failure?

Authors:  Aaron H Carter; Eoin C Sheehan; S M Javad Mortazavi; James J Purtill; Peter F Sharkey; Javad Parvizi
Journal:  J Arthroplasty       Date:  2011-05-08       Impact factor: 4.757

7.  Postoperative total hip prosthetic femoral head dislocations. Incidence, etiologic factors, and management.

Authors:  R S Turner
Journal:  Clin Orthop Relat Res       Date:  1994-04       Impact factor: 4.176

8.  Constrained liner in neurologic or cognitively impaired patients undergoing primary THA.

Authors:  Philippe Hernigou; Paolo Filippini; Charles-Henri Flouzat-Lachaniette; Sobrinho Uirassu Batista; Alexandre Poignard
Journal:  Clin Orthop Relat Res       Date:  2010-12       Impact factor: 4.176

9.  An analysis of the risk of hip dislocation with a contemporary total joint registry.

Authors:  Monti Khatod; Thomas Barber; Elizabeth Paxton; Robert Namba; Donald Fithian
Journal:  Clin Orthop Relat Res       Date:  2006-06       Impact factor: 4.176

10.  Total hip arthroplasty dislocation rate following isolated cup revision using Hueter's direct anterior approach on a fracture table.

Authors:  A Cogan; S Klouche; P Mamoudy; E Sariali
Journal:  Orthop Traumatol Surg Res       Date:  2011-07-23       Impact factor: 2.256

View more
  3 in total

1.  Can Dislocation of a Constrained Liner Be Salvaged With Dual-mobility Constructs in Revision THA?

Authors:  Brian P Chalmers; Graham D Pallante; Michael J Taunton; Rafael J Sierra; Robert T Trousdale
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

2.  The focally constrained liner is a reasonable option for revision of unstable total hip arthroplasty.

Authors:  Tatu J Mäkinen; Simcha G Fichman; Wael A Rahman; Tomas Amenabar; Oleg Safir; Allan E Gross; Paul R T Kuzyk
Journal:  Int Orthop       Date:  2015-12-23       Impact factor: 3.075

3.  Constrained captive acetabular cup for recurrent dislocation of hemiarthroplasty in elderly: A case series.

Authors:  Aysha Rajeev; Paul Banaszkiewicz
Journal:  Int J Surg Case Rep       Date:  2016-04-22
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.