| Literature DB >> 21993594 |
Werner Jung1, Vlada Zvereva, Bajram Hajredini, Sebastian Jäckle.
Abstract
Due of its superior soft tissue imaging capabilities, magnetic resonance imaging (MRI) has become the imaging modality of choice in many clinical situations, as illustrated by the tremendous growth in the number of MRIs performed over the past 2 decades. In parallel, the number of patients who require pacemakers or implantable cardiac defibrillators is increasing as indications for these devices broaden and the population ages. Taken together, these phenomena present an important clinical issue, as MR scans are generally contraindicated-except in urgent situations-in patients who have implanted cardiovascular devices. Potentially deleterious interactions between the magnetic fields and radio frequency (RF) energy produced by MR equipment and implantable devices have been identified, including inhibition of pacing, asynchronous/high-rate pacing, lead tip heating, and loss of capture. New devices that incorporate technologies to improve MR safety in patients with pacemakers have recently received approval in Europe and are under evaluation in the United States. Initial data from these devices suggest that these devices are safe in the MRI environment.Entities:
Mesh:
Year: 2011 PMID: 21993594 PMCID: PMC3224227 DOI: 10.1007/s10840-011-9610-0
Source DB: PubMed Journal: J Interv Card Electrophysiol ISSN: 1383-875X Impact factor: 1.900
Fig. 1Total MRI procedure volume in millions, 2000–2007, hospital and nonhospital sites [2]. Reprinted with permission from AuntMinnie.com
Potential effects of magnetic resonance imaging on pacemakers [10]
| Potential effects of MRI on pacemakers | |
|---|---|
| (1) | Static magnetic field |
| Mechanical forces on ferromagnetic components | |
| Unpredictable magnetic sensor activation, reed-switch closure | |
| Changes in electrocardiograms | |
| (2) | Modulated radio frequency field |
| Heating of cardiac tissue adjacent to lead electrodes | |
| Possible induction of life-threatening arrhythmias (very rare) | |
| Pacemaker reprogramming or reset | |
| RF interactions with the device (over- and undersensing) | |
| (3) | Gradient magnetic field |
| Possible induction of life-threatening arrhythmias (unlikely in bipolar mode) | |
| Induced voltages on leads cause over- and undersensing | |
| (4) | Combined field effects |
| Alteration of device function because of electromagnetic interference | |
| Mechanical forces (vibration) | |
| Electronic reset of device | |
| Damage to pacemaker/implantable cardiac defibrillator and/or leads | |
Potential variables affecting the magnitude of risks in pacemaker/ICD patients undergoing MRI
| (1) | Pacemaker design |
| Ferromagnetic content | |
| Switch design (reed vs. Hall) | |
| Availability of dedicated modes for use during MRI | |
| (2) | Lead design and length |
| Lead geometry | |
| Lead input filtering capacitance | |
| (3) | Blood flow at lead/tissue interface |
| (4) | Presence of abandoned leads |
| (5) | MRI scan duration |
| (6) | Strength of RF field |
| (7) | Isocenter of scan |
| (8) | Type of imaging sequence |
| (9) | Patient and device/lead position within scanner |
| (10) | Type and extent of patient monitoring |
Fig. 2Potential for inhibition of pacing [26]
New technologies incorporated in MR-conditional pacing systems
| Change | Potential benefit |
|---|---|
| Lead input filtering capacitance | Minimize energy induced on leads/discharged at tip |
| Reed switch replaced by Hall sensor | Control switch behavior |
| Internal power supply circuit protection | Prevents energy induced on telemetry antenna from disrupting internal power supplies |
| Reduction in ferromagnetic components | Decreases susceptibility to magnetic attraction |
| Dedicated modes for use during MRI | Suspension of diagnostic data collection and atrial arrhythmia therapy |
| Lead geometry changes | Reduce lead heating |
| Radiographic markers | Allows identification of MRI-conditional devices |