Christian G Wollmann1, Karin Thudt, Bernd Kaiser, Erich Salomonowitz, Harald Mayr, Sebastian Globits.
Abstract
BACKGROUND: No published data exist about the safety of diagnostic magnetic resonance (MR) of the heart performed in a larger series of patients implanted with MR conditional pacemakers (PM). The purpose of our study is to analyse safety and potential alterations of electrical lead parameters in patients implanted with the EnRhythm/Advisa MRI SureScan PM with 5086MRI leads (Medtronic Inc.) during and after MR of the heart at 1.5 Tesla.Entities:
Mesh:
Year: 2014 PMID: 24886167 PMCID: PMC4026122 DOI: 10.1186/1532-429X-16-30
Source DB: PubMed Journal: J Cardiovasc Magn Reson ISSN: 1097-6647 Impact factor: 5.364
Patient Demographics
| 36 | 100 | |
| 12 | 33 | |
| 69 ± 13 [73] | ||
| | | |
| Higher degree AV block | 19 | 53 |
| Sick sinus syndrome | 15 | 42 |
| Other indication | 2 | 6 |
| 13 | 36 | |
| 30 | 83 | |
| 9 | 25 | |
| 14 | 39 | |
| 5 | 14 | |
| | | |
| EnRhythm MRI Sure Scan EMDR01 | 25 | 69 |
| Advisa MRI Sure Scan A3DR01 | 11 | 31 |
| | ||
| AP (%; [median]) | 40 ± 37 [23] | |
| VP (%; [median]) | 26 ± 39 [2] | |
| AF-Burden (%; [median]) | 17 ± 34 [0] | |
| 6 | 18 | |
| | | |
| V00 | 11 | 31 |
| D00 | 25 | 69 |
| | | |
| apical | 23 | 64 |
| septal | 13 | 36 |
| 35 | 97 | |
| 5.3 ± 3.8 [4.5] | ||
Measurements before and after MR of the heart
| Time from MR (months) | - | - | - | 2.8 ± 0.7 [3.0] | 16.0 ± 3.2 [15.2] | - |
| Time from implantation (months) | 5.3 ± 3.8 [4.5] | - | - | 7.9 ± 3.7 [7.6] | 21.2 ± 5.7 [20.7] | - |
| RA Sensing (mV) | 2.4 ± 1.3 [2.4] | 2.4 ± 1.5 [2.3] | 2.3 ± 1.5 [2.0] | 2.4 ± 1.3 [2.3] | 2.4 ± 1.5 [2.2] | n.s. |
| RA PCT (V@0,4ms) | 0.69 ± 0.22 [0.56] | 0.63 ± 0.19 [0.50]† | 0.62 ± 0.19 [0.50] | 0.69 ± 0.22 [0.50] | 0.64 ± 0.22 [0.50] | n.s. |
| RA Pimp (Ohms) | 511 ± 90 [508] | 512 ± 100 [504] | 503 ± 90 [478] | 504 ± 66 [508] | 490 ± 72 [484] | n.s. |
| RV Sensing (mV) | 9.8 ± 5.3 [7.5] | 10.9 ± 5.0 [10.9] | 10.3 ± 5.3 [8.9]† | 10.7 ± 4.8 [10.3] | 9.9 ± 5.1 [6.8] | n.s. |
| RV PCT (V@0,4ms) | 0.84 ± 0.27 [1.00] | 0.82 ± 0.27 [1.00] | 0.84 ± 0.24 [1.00] | 0.85 ± 0.23 [1.00] | 0.90 ± 0.67 [1.00] | n.s. |
| RV Pimp (Ohms) | 508 ± 75 [504] | 516 ± 47 [512]† | 515 ± 62 [500] | 509 ± 84 [494] | 482 ± 58 [488]† | n.s. |
| Battery voltage (V) | 3.02 ± 0.03 [3.01] | - | 3.01 ± 0.03 [3.00]† | 3.0 ± 0.02 [3.00]† | 3.00 ± 0.01 [3.00]† | 0.024 |
| | | | | | ||
| RA sensing | | 5 ± 49 [0] | 4 ± 47 [0] | 8 ± 48 [0] | 8 ± 43 [5] | n.s. |
| RA PCT | | -8 ± 17 [0] | -6 ± 28 [0] | 5 ± 40 [0] | -2 ± 38 [0] | n.s. |
| RA Pimp | | 1 ± 11 [0] | -1 ± 11 [0] | -0 ± 13 [0] | -3 ± 11 [-6] | n.s. |
| RV sensing | | 9 ± 26 [0] | 9 ± 20 [4] | 17 ± 39 [8] | -3 ± 27 [-6] | 0.046 |
| RV PCT | | -0 ± 33 [0] | 7 ± 38 [0] | 12 ± 49 [0] | 3 ± 39 [0] | n.s. |
| RV PImp | | 3 ± 7 [4] | 2 ± 7 [2] | 1 ± 14 [0] | -4 ± 10 [-4] | 0.015 |
| Battery voltage (V) | - | -0.3 ± 0.3 [-0.2] | -0.4 ± 0.6 [-0.3] | -0.6 ± 0.9 [-0.3] | n.s. | |
FU…follow-up; PCT…pacing capture threshold; Pimp…pacing impedance.
*Oneway ANOVA.
† p < 0.05 [Wilcoxon signed Ranks Test].
Pacemaker statistics
| - | 2.8 ± 0.7 [3.0] | 16.0 ± 3.2 [15.2] | - | |
| 40 ± 37 [23] | 42 ± 36 [25] | 45 ± 36 [40] | n.s. | |
| 26 ± 39 [2] | 26 ± 38 [3] | 30 ± 38 [6] | n.s. | |
| 17 ± 34 [0] | 15 ± 33 [0] | 20 ± 37 [0] | n.s. |
FU…follow-up.
*Oneway ANOVA.
Figure 1Device related imaging artifacts. A: Vertical long axis view showing subtle susceptibility artefacts (short arrow) around the ferromagnetic right ventricular lead (long arrow). In addition, there are prominent off-resonance stripe artifacts in the left ventricle (small arrows) with no impairment of endocardial border detection. B: Short axis view with the typical signal void caused by the RV lead (short arrow) in the inferior wall not interfering with the endocardial border. Note the large signal void and circular stripe artifacts caused by the generator (long arrow) with no interference with the heart.