Evaluation of a computed radiography system for megavoltage photon beam dosimetry.

Arthur J Olch1.   

Abstract

Computed radiography (CR) systems have been gaining adoption as digital replacements for film for diagnostic and therapy imaging. As a result, film processors are being removed from service, leaving a void for the medical physicists who use film and processors for two-dimensional mega-voltage beam dosimetry. This is the first report to evaluate the ability of a commercial CR reader and storage phosphor plate system to accurately quantitate absolute dose and dose distributions from a 6 MV photon beam. There are potential advantages and disadvantages of current CR systems compared to film systems. CR systems inherently produce a linear dose-response over several logs of dose. However, the barium in the storage phosphor has a higher atomic number than the silver in film, resulting in significant energy sensitivity. The purpose of this work is to fully characterize the impact of these and other features of this CR system relevant to dosimetry. The tests performed and reported on in this study include uniformity of readout across a uniform field, geometrical accuracy, intra- and interday reproducibility, signal decay with time and with light exposure, dose-to-signal calibration, high dose effects, obliquity effects, perpendicular and parallel calibration results, field size and depth of measurement effects and the use of lead filters to minimize them, and intensity modulated radiation therapy quality assurance test results compared to that for film. Practical techniques are provided to optimize the accuracy of the system as a dosimetric replacement for film.

Entities:  

Mesh:

Year:  2005        PMID: 16266113     DOI: 10.1118/1.2012787

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  10 in total

1.  Quasi-IMAT study with conventional equipment to show high plan quality with a single gantry arc.

Authors:  Judith Alvarez Moret; Oliver Kölbl; Ludwig Bogner
Journal:  Strahlenther Onkol       Date:  2009-02-18       Impact factor: 3.621

2.  The role of oxygen in the photostimulation luminescence process of europium doped potassium chloride.

Authors:  Zhiyan Xiao; Thomas R Mazur; Joseph P Driewer; H Harold Li
Journal:  Radiat Meas       Date:  2015-05-01       Impact factor: 1.898

3.  Radiation hardness of the storage phosphor europium doped potassium chloride for radiation therapy dosimetry.

Authors:  Joseph P Driewer; Haijian Chen; Andres Osvet; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2011-08       Impact factor: 4.071

4.  Temporal signal stability of KCl:Eu(2+) storage phosphor dosimeters.

Authors:  Zhiyan Xiao; Rachael Hansel; Haijian Chen; Dongsu Du; Deshan Yang; H Harold Li
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

5.  X ray storage performance of KCl:Eu2+ with high cumulated dose.

Authors:  Rachael A Hansel; Zhiyan Xiao; Lei Zhang; H Harold Li
Journal:  Nucl Instrum Methods Phys Res B       Date:  2014-05-01       Impact factor: 1.377

6.  Two-dimensional high spatial-resolution dosimeter using europium doped potassium chloride: a feasibility study.

Authors:  H Harold Li; Joseph P Driewer; Zhaohui Han; Daniel A Low; Deshan Yang; Zhiyan Xiao
Journal:  Phys Med Biol       Date:  2014-03-20       Impact factor: 3.609

7.  Performance of KCl:Eu2+ storage phosphor dosimeters for low-dose measurements.

Authors:  H Harold Li; Zhiyan Xiao; Rachael Hansel; Nels Knutson; Deshan Yang
Journal:  Phys Med Biol       Date:  2013-06-05       Impact factor: 3.609

8.  Quantitative megavoltage radiation therapy dosimetry using the storage phosphor KCl: Eu2+.

Authors:  Zhaohui Han; Joseph P Driewer; Yuanshui Zheng; Daniel A Low; H Harold Li
Journal:  Med Phys       Date:  2009-08       Impact factor: 4.071

9.  Recent developments of optically stimulated luminescence materials and techniques for radiation dosimetry and clinical applications.

Authors:  A S Pradhan; J I Lee; J L Kim
Journal:  J Med Phys       Date:  2008-07

10.  Detection of electron beam energy variations using a computed radiography system.

Authors:  Yang C Cai; Yuanyuan Ge; Damian Bernard; Julius Turian; James C H Chu
Journal:  J Appl Clin Med Phys       Date:  2009-10-15       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.