A new rebound tonometer for home monitoring of intraocular pressure.

André Rosentreter1, Kerstin S Jablonski, Anne C Mellein, Stergiani Gaki, Arno Hueber, Thomas S Dietlein.   

Abstract

BACKGROUND: To compare intraocular pressure (IOP) measurements obtained by the Icare ONE rebound tonometer (RTONE) and the Goldmann applanation tonometer (GAT) in healthy persons and glaucoma patients in a prospective study, and to investigate the influence of central corneal thickness (CCT).
METHODS: Measurements on 126 right eyes were obtained by three equally skilled ophthalmologists with each of the above-mentioned tonometers. In addition, patients measured their own IOP with the RTONE (RTONE(p)). The means and standard deviation for all tonometers were compared. Agreement between the tonometers was calculated using the Bland-Altman method.
RESULTS: A total of 95 (75.3%) patients were able to perform correct self-tonometry. Mean IOPs obtained were 17.1 ± 5.9 mmHg (RTONE performed by ophthalmologist: RTONE (o)), 17.3 ± 5.6 mmHg (RTONE(p)) and 16.5 ± 5.1 mmHg (GAT). Correlation analysis indicated a good correlation between IOP readings obtained using RTONE(o) and RTONE(p) (ρ = 0.916; p < 0.001) and RTONE(o) and GAT (ρ = 0.901; p < 0.001). Bland-Altman analysis revealed a mean difference (bias) between RTONE(o) and RTONE(p), between RTONE(o) and GAT, and between RTONE(p) and GAT of -0.2, 0.6, and 0.8 mmHg, respectively, with 95% limits of agreement of -5.0 to 4.5, -4.4 to 5.6, and -4.6 to 6.1 mmHg, respectively. The difference between RTONE(o) and GAT significantly increased with increasing CCT (ρ = 0.004), with a 10% increase in CCT resulting in a 1.8% increase in the difference.
CONCLUSIONS: Measurements obtained with the RTONE, either by an ophthalmologist or by the patient, showed an excellent correlation with those provided by applanation tonometry. RTONE generally tends to overestimate IOP compared to GAT readings and displays a dependence on CCT. This study was registered with the DRKS (German Clinical Trials Register; www.germanctr.de ; DRKS00000478).

Entities:  

Mesh:

Year:  2011        PMID: 21842435     DOI: 10.1007/s00417-011-1785-7

Source DB:  PubMed          Journal:  Graefes Arch Clin Exp Ophthalmol        ISSN: 0721-832X            Impact factor:   3.117


  33 in total

1.  Comparison of the iCare rebound tonometer and the Goldmann applanation tonometer over a wide IOP range.

Authors:  S Munkwitz; A Elkarmouty; E M Hoffmann; N Pfeiffer; H Thieme
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2008-01-12       Impact factor: 3.117

2.  Wireless contact lens sensor for intraocular pressure monitoring: assessment on enucleated pig eyes.

Authors:  Matteo Leonardi; Elie M Pitchon; Arnaud Bertsch; Philippe Renaud; André Mermoud
Journal:  Acta Ophthalmol       Date:  2008-11-12       Impact factor: 3.761

3.  Standardizing the measurement of intraocular pressure for clinical research. Guidelines from the Eye Care Technology Forum.

Authors:  M A Kass
Journal:  Ophthalmology       Date:  1996-01       Impact factor: 12.079

4.  Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device.

Authors:  Sanjay Asrani; Ashmita Chatterjee; David K Wallace; Cecilia Santiago-Turla; Sandra Stinnett
Journal:  J Glaucoma       Date:  2011-02       Impact factor: 2.503

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients.

Authors:  Paolo Brusini; Maria Letizia Salvetat; Marco Zeppieri; Claudia Tosoni; Lucia Parisi
Journal:  J Glaucoma       Date:  2006-06       Impact factor: 2.503

7.  The effect of corneal thickness on applanation tonometry.

Authors:  M M Whitacre; R A Stein; K Hassanein
Journal:  Am J Ophthalmol       Date:  1993-05-15       Impact factor: 5.258

8.  Effect of corneal thickness on dynamic contour, rebound, and goldmann tonometry.

Authors:  Jose M Martinez-de-la-Casa; Julian Garcia-Feijoo; Eva Vico; Ana Fernandez-Vidal; Jose M Benitez del Castillo; Mohamed Wasfi; J Garcia-Sanchez
Journal:  Ophthalmology       Date:  2006-09-25       Impact factor: 12.079

9.  Method for the noninvasive measurement of intraocular pressure in mice.

Authors:  John Danias; Antti I Kontiola; Theodoros Filippopoulos; Thom Mittag
Journal:  Invest Ophthalmol Vis Sci       Date:  2003-03       Impact factor: 4.799

10.  The influence of corneal properties on rebound tonometry.

Authors:  Wan-sang Chui; Andrew Lam; Davie Chen; Roger Chiu
Journal:  Ophthalmology       Date:  2007-06-15       Impact factor: 12.079

View more
  13 in total

1.  [Rebound tonometry after vitreoretinal surgery].

Authors:  A Rosentreter; R Hoerster; T Schick; N Eter; T S Dietlein; S Fauser
Journal:  Ophthalmologe       Date:  2015-11       Impact factor: 1.059

2.  [Evaluation of the self-tonometer Icare ONE in comparison to Goldmann applanation tonometry].

Authors:  V Witte; Ä Glass; R Beck; R Guthoff
Journal:  Ophthalmologe       Date:  2012-10       Impact factor: 1.059

3.  [Application of rebound self-tonometry for measurements in a supine position].

Authors:  Afsaneh Amani; André Rosentreter; Thomas Dietlein; Robert Hoerster
Journal:  Ophthalmologe       Date:  2019-08       Impact factor: 1.059

4.  Measurement of Intraocular Pressure by Patients With Glaucoma.

Authors:  Savva Pronin; Lyndsay Brown; Roly Megaw; Andrew J Tatham
Journal:  JAMA Ophthalmol       Date:  2017-10-01       Impact factor: 7.389

5.  Comparison of intraocular pressure measurements with different contact tonometers in young healthy persons.

Authors:  Saulius Galgauskas; Rasa Strupaite; Ernesta Strelkauskaite; Rimvydas Asoklis
Journal:  Int J Ophthalmol       Date:  2016-01-18       Impact factor: 1.779

6.  Twenty-four-hour efficacy of preservative-free tafluprost for open-angle glaucoma patients, assessed by home intraocular pressure (Icare-ONE) and blood-pressure monitoring.

Authors:  Soon Young Cho; Yong Yeon Kim; Chungkwon Yoo; Tae-Eun Lee
Journal:  Jpn J Ophthalmol       Date:  2015-09-28       Impact factor: 2.447

Review 7.  An overview of home tonometry and telemetry for intraocular pressure monitoring in humans.

Authors:  Edward Yung; Valerie Trubnik; L Jay Katz
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-05-29       Impact factor: 3.117

8.  [Ease of handling of first and second generation rebound tonometers].

Authors:  N Mihailovic; J Termühlen; M Alnawaiseh; N Eter; T S Dietlein; A Rosentreter
Journal:  Ophthalmologe       Date:  2016-04       Impact factor: 1.059

Review 9.  Systematic Review of Current Devices for 24-h Intraocular Pressure Monitoring.

Authors:  Sabita M Ittoop; Jeffrey R SooHoo; Leonard K Seibold; Kaweh Mansouri; Malik Y Kahook
Journal:  Adv Ther       Date:  2016-08-16       Impact factor: 3.845

Review 10.  Noninvasive intraocular pressure monitoring: current insights.

Authors:  Stefan De Smedt
Journal:  Clin Ophthalmol       Date:  2015-07-30
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.