Literature DB >> 20436362

Evaluation of the ICare rebound tonometer as a home intraocular pressure monitoring device.

Sanjay Asrani1, Ashmita Chatterjee, David K Wallace, Cecilia Santiago-Turla, Sandra Stinnett.   

Abstract

PURPOSE: (a) To investigate whether the ICare rebound tonometer can provide accurate measurements of intraocular pressure (IOP) in the hands of an inexperienced user compared with ICare measurements and Goldmann tonometry by a trained technician and (b) to assess the intrauser reproducibility of IOP measurements and the learning curve among patients using the ICare rebound tonometer.
METHODS: A trained technician used the ICare rebound tonometer to measure the IOP of the right eye of 100 glaucoma patients. The technician then instructed each patient on use of the ICare tonometer. Each patient then measured his/her own pressure using the ICare tonometer. Finally, a different technician, who was masked to both of the earlier readings, measured IOP by Goldmann applanation tonometry. Thirty patients repeated the ICare measurement 3 times (once every 5 min) 20 minutes after the initial IOP measurement.
RESULTS: Of the 100 patients, 82 of patient ICare and the technician ICare readings were within 3 mm Hg of each other, and 75 of the patient ICare and Goldmann applanation tonometry measurements were within 3 mm Hg of each other. Intraclass correlations between self-administered ICare measurements 1 and 2, 1 and 3, and 2 and 3 were 0.69, 0.71 and 0.81, respectively.
CONCLUSION: In this study, the ICare rebound tonometer was accurate and reliable in the hands of patients. Patients can easily learn to self-administer this test, possibly allowing for home monitoring of IOP.

Entities:  

Mesh:

Year:  2011        PMID: 20436362     DOI: 10.1097/IJG.0b013e3181d78811

Source DB:  PubMed          Journal:  J Glaucoma        ISSN: 1057-0829            Impact factor:   2.503


  13 in total

1.  Evaluation of the Icare-ONE rebound tonometer as a self-measuring intraocular pressure device in normal subjects.

Authors:  Ioannis Halkiadakis; Aimilianos Stratos; George Stergiopoulos; Eleni Patsea; Sotiris Skouriotis; Panagiotis Mitropoulos; Dimitrios Papaconstantinou; Gerasimos Georgopoulos
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-08       Impact factor: 3.117

2.  Age variations in intraocular pressure in a cohort of healthy Austrian school children.

Authors:  W A Dusek; B K Pierscionek; J F McClelland
Journal:  Eye (Lond)       Date:  2012-03-23       Impact factor: 3.775

3.  [Application of rebound self-tonometry for measurements in a supine position].

Authors:  Afsaneh Amani; André Rosentreter; Thomas Dietlein; Robert Hoerster
Journal:  Ophthalmologe       Date:  2019-08       Impact factor: 1.059

4.  A new rebound tonometer for home monitoring of intraocular pressure.

Authors:  André Rosentreter; Kerstin S Jablonski; Anne C Mellein; Stergiani Gaki; Arno Hueber; Thomas S Dietlein
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2011-08-14       Impact factor: 3.117

5.  A comparison of Icare PRO and Perkins tonometers in anesthetized children.

Authors:  Massimiliano Serafino; Edoardo Villani; Andrea Lembo; Giovanni Rabbiolo; Claudia Specchia; Rupal H Trivedi; Paolo Nucci
Journal:  Int Ophthalmol       Date:  2019-07-16       Impact factor: 2.031

6.  Twenty-four-hour efficacy of preservative-free tafluprost for open-angle glaucoma patients, assessed by home intraocular pressure (Icare-ONE) and blood-pressure monitoring.

Authors:  Soon Young Cho; Yong Yeon Kim; Chungkwon Yoo; Tae-Eun Lee
Journal:  Jpn J Ophthalmol       Date:  2015-09-28       Impact factor: 2.447

Review 7.  An overview of home tonometry and telemetry for intraocular pressure monitoring in humans.

Authors:  Edward Yung; Valerie Trubnik; L Jay Katz
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-05-29       Impact factor: 3.117

8.  COMPARISON OF IOP READINGS USING REBOUND I CARE TONOMETER AND PERKINS APPLANATION TONOMETER IN AN AFRICAN POPULATION.

Authors:  D S Ademola-Popoola; A F Odi; T M Akande
Journal:  J West Afr Coll Surg       Date:  2014 Jan-Mar

9.  [Ease of handling of first and second generation rebound tonometers].

Authors:  N Mihailovic; J Termühlen; M Alnawaiseh; N Eter; T S Dietlein; A Rosentreter
Journal:  Ophthalmologe       Date:  2016-04       Impact factor: 1.059

10.  Short-term effect of standard automated perimetry testing on intraocular pressure in patients with open-angle glaucoma.

Authors:  Chang Mok Lee; Young Cheol Yoo
Journal:  ISRN Ophthalmol       Date:  2013-02-14