Literature DB >> 21513536

Is the rotation of the femoral head a potential initiation for cutting out? A theoretical and experimental approach.

Andreas Lenich1, Samuel Bachmeier, Lukas Prantl, Michael Nerlich, Jochen Hammer, Edgar Mayr, Amir Andreas Al-Munajjed, Bernd Füchtmeier.   

Abstract

BACKGROUND: Since cut-out still remains one of the major clinical challenges in the field of osteoporotic proximal femur fractures, remarkable developments have been made in improving treatment concepts. However, the mechanics of these complications have not been fully understood.We hypothesize using the experimental data and a theoretical model that a previous rotation of the femoral head due to de-central implant positioning can initiate a cut-out.
METHODS: In this investigation we analysed our experimental data using two common screws (DHS/Gamma 3) and helical blades (PFN A/TFN) for the fixation of femur fractures in a simple theoretical model applying typical gait pattern on de-central positioned implants. In previous tests during a forced implant rotation by a biomechanical testing machine in a human femoral head the two screws showed failure symptoms (2-6Nm) at the same magnitude as torques acting in the hip during daily activities with de-central implant positioning, while the helical blades showed a better stability (10-20Nm).To calculate the torque of the head around the implant only the force and the leverarm is needed (N [Nm] = F [N] * × [m]). The force F is a product of the mass M [kg] multiplied by the acceleration g [m/s2]. The leverarm is the distance between the center of the head of femur and the implant center on a horizontal line.
RESULTS: Using 50% of 75 kg body weight a torque of 0.37Nm for the 1 mm decentralized position and 1.1Nm for the 3 mm decentralized position of the implant was calculated. At 250% BW, appropriate to a normal step, torques of 1.8Nm (1 mm) and 5.5Nm (3 mm) have been calculated.Comparing of the experimental and theoretical results shows that both screws fail in the same magnitude as torques occur in a more than 3 mm de-central positioned implant.
CONCLUSION: We conclude the center-center position in the head of femur of any kind of lag screw or blade is to be achieved to minimize rotation of the femoral head and to prevent further mechanical complications.

Entities:  

Mesh:

Year:  2011        PMID: 21513536      PMCID: PMC3108935          DOI: 10.1186/1471-2474-12-79

Source DB:  PubMed          Journal:  BMC Musculoskelet Disord        ISSN: 1471-2474            Impact factor:   2.362


  14 in total

1.  A ten-year analysis of intertrochanteric fractures of the femur.

Authors:  M CLEVELAND; D M BOSWORTH; F R THOMPSON; H J WILSON; T ISHIZUKA
Journal:  J Bone Joint Surg Am       Date:  1959-12       Impact factor: 5.284

2.  Anisotropy of the fatigue behaviour of cancellous bone.

Authors:  S Dendorfer; H J Maier; D Taylor; J Hammer
Journal:  J Biomech       Date:  2007-11-19       Impact factor: 2.712

3.  First clinical and biomechanical results of the Trochanteric Fixation Nail (TFN).

Authors:  A Lenich; J Fierlbeck; A Al-Munajjed; S Dendorfer; R Mai; B Füchtmeier; E Mayr; J Hammer
Journal:  Technol Health Care       Date:  2006       Impact factor: 1.285

4.  [Epidemiology of trochanteric femoral fractures over 2 decades (1972-1988)].

Authors:  A Lustenberger; R Ganz
Journal:  Unfallchirurg       Date:  1995-05       Impact factor: 1.000

5.  The AO/ASIF proximal femoral nail (PFN) for the treatment of unstable trochanteric femoral fracture.

Authors:  G Al-yassari; R J Langstaff; J W M Jones; M Al-Lami
Journal:  Injury       Date:  2002-06       Impact factor: 2.586

Review 6.  Hip joint loading during walking and running, measured in two patients.

Authors:  G Bergmann; F Graichen; A Rohlmann
Journal:  J Biomech       Date:  1993-08       Impact factor: 2.712

7.  ["Cutting out" in pertrochanteric fractures--problem of osteoporosis?].

Authors:  F Bonnaire; A Weber; O Bösl; C Eckhardt; K Schwieger; B Linke
Journal:  Unfallchirurg       Date:  2007-05       Impact factor: 1.000

8.  Biomechanical characterisation of osteosyntheses for proximal femur fractures: helical blade versus screw.

Authors:  Amir A Al-Munajjed; Joachim Hammer; Edgar Mayr; Michael Nerlich; Andreas Lenich
Journal:  Stud Health Technol Inform       Date:  2008

9.  The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip.

Authors:  M R Baumgaertner; S L Curtin; D M Lindskog; J M Keggi
Journal:  J Bone Joint Surg Am       Date:  1995-07       Impact factor: 5.284

10.  Is a helical shaped implant a superior alternative to the Dynamic Hip Screw for unstable femoral neck fractures? A biomechanical investigation.

Authors:  Markus Windolf; Volker Braunstein; Christof Dutoit; Karsten Schwieger
Journal:  Clin Biomech (Bristol, Avon)       Date:  2008-10-31       Impact factor: 2.063

View more
  20 in total

Review 1.  [Pertrochanteric femoral fractures in the elderly].

Authors:  G H Sandmann; P Biberthaler
Journal:  Unfallchirurg       Date:  2015-05       Impact factor: 1.000

2.  A unique cause of Gamma 3 cut-out: A case report and literature review.

Authors:  Ioannis Papaioannou; Thomas Repantis; Panagiotis Korovessis
Journal:  J Clin Orthop Trauma       Date:  2020-08-13

Review 3.  Is rotation the mode of failure in pertrochanteric fractures fixed with nails? Theoretical approach and illustrative cases.

Authors:  C Kokoroghiannis; D Vasilakos; K Zisis; G Dimitriou; E Pappa; D Evangelopoulos
Journal:  Eur J Orthop Surg Traumatol       Date:  2019-09-20

Review 4.  Intertrochanteric fractures: a review of fixation methods.

Authors:  Senthil Nathan Sambandam; Jayadev Chandrasekharan; Varatharaj Mounasamy; Cyril Mauffrey
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-03-30

Review 5.  [Hip fractures in the elderly : Osteosynthesis versus joint replacement].

Authors:  M Knobe; C H Siebert
Journal:  Orthopade       Date:  2014-04       Impact factor: 1.087

6.  Screw-blade fixation systems in Pauwels three femoral neck fractures: a biomechanical evaluation.

Authors:  Matthias Knobe; Simon Altgassen; Klaus-Jürgen Maier; Gertraud Gradl-Dietsch; Chris Kaczmarek; Sven Nebelung; Kajetan Klos; Bong-Sung Kim; Boyko Gueorguiev; Klemens Horst; Benjamin Buecking
Journal:  Int Orthop       Date:  2017-08-06       Impact factor: 3.075

7.  [The rotationally stable screw-anchor with trochanteric stabilizing plate (RoSA/TSP) : First results in unstable trochanteric femur fractures].

Authors:  K-J Maier; B Bücking; K Horst; H Andruszkow; F Hildebrand; M Knobe
Journal:  Unfallchirurg       Date:  2017-12       Impact factor: 1.000

8.  Short vs. long intramedullary nail systems in trochanteric fractures: A randomized prospective single center study.

Authors:  Șerban Dragosloveanu; Christiana D M Dragosloveanu; Dragoș C Cotor; Cristian I Stoica
Journal:  Exp Ther Med       Date:  2021-12-01       Impact factor: 2.447

9.  Outcome of the dynamic helical hip screw system for intertrochanteric hip fractures in the elderly patients.

Authors:  Natasha T O'Malley; Andrew-Paul Deeb; Karilee W Bingham; Stephen L Kates
Journal:  Geriatr Orthop Surg Rehabil       Date:  2012-06

10.  Screw versus helical proximal femoral nail in the treatment of unstable trochanteric fractures in the elderly.

Authors:  Col Narinder Kumar; Maj P K Srivastava
Journal:  J Clin Orthop Trauma       Date:  2018-07-23
View more