Literature DB >> 33717882

A unique cause of Gamma 3 cut-out: A case report and literature review.

Ioannis Papaioannou1, Thomas Repantis1, Panagiotis Korovessis1.   

Abstract

Gamma 3 nail is a wide spread intramedullary device for fixation of per trochanteric fractures. Cut out of the lag screw is the most common complication of this implant. We present a 62-year-old female patient, who underwent a total hip arthroplasty following cut out of a Gamma 3 nail in the femoral neck. The cause of the cut out in our case is actually unique. Our intraoperative findings accompanied with the radiographic evaluation argue that the malposition of the set screw was the cause of failure, due to the rotational instability of femoral head-lag screw unit. We present this case with detailed description, highlighting the proper use of this specific nail and appose a brief literature review.
© 2020 Delhi Orthopedic Association. All rights reserved.

Entities:  

Keywords:  Cut out failure; Gamma 3 nail; Head-screw rotation; Intertrochanteric fracture; Set screw malposition

Year:  2020        PMID: 33717882      PMCID: PMC7920013          DOI: 10.1016/j.jcot.2020.08.006

Source DB:  PubMed          Journal:  J Clin Orthop Trauma        ISSN: 0976-5662


  8 in total

1.  Is the rotation of the femoral head a potential initiation for cutting out? A theoretical and experimental approach.

Authors:  Andreas Lenich; Samuel Bachmeier; Lukas Prantl; Michael Nerlich; Jochen Hammer; Edgar Mayr; Amir Andreas Al-Munajjed; Bernd Füchtmeier
Journal:  BMC Musculoskelet Disord       Date:  2011-04-22       Impact factor: 2.362

2.  Lag screws for hip fracture fixation: Evaluation of migration resistance under simulated walking.

Authors:  Larry W Ehmke; Daniel C Fitzpatrick; James C Krieg; Steven M Madey; Michael Bottlang
Journal:  J Orthop Res       Date:  2005-07-01       Impact factor: 3.494

Review 3.  Is rotation the mode of failure in pertrochanteric fractures fixed with nails? Theoretical approach and illustrative cases.

Authors:  C Kokoroghiannis; D Vasilakos; K Zisis; G Dimitriou; E Pappa; D Evangelopoulos
Journal:  Eur J Orthop Surg Traumatol       Date:  2019-09-20

4.  Risk factors for cut-out failure of Gamma3 nails in treating unstable intertrochanteric fractures: An analysis of 176 patients.

Authors:  Shang-Wen Tsai; Chien-Fu Jeff Lin; Yun-Hsuan Tzeng; Chun-Cheng Lin; Ching-Kuei Huang; Ming-Chau Chang; Chao-Ching Chiang
Journal:  J Chin Med Assoc       Date:  2017-09       Impact factor: 2.743

5.  Inadequate 'three-point' proximal fixation predicts failure of the Gamma nail.

Authors:  S G F Abram; T C B Pollard; A J M D Andrade
Journal:  Bone Joint J       Date:  2013-06       Impact factor: 5.082

6.  Is the lag screw sliding effective in the intramedullary nailing in A1 and A2 AO-OTA intertrochanteric fractures? A prospective study of Sliding and None-sliding lag screw in Gamma-III nail.

Authors:  Yi Zhu; Severin Meili; Changqing Zhang; Congfeng Luo; Bing-fang Zeng
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2012-09-01       Impact factor: 2.953

7.  Critical factors in cut-out complication after Gamma Nail treatment of proximal femoral fractures.

Authors:  Alicja J Bojan; Claudia Beimel; Gilbert Taglang; David Collin; Carl Ekholm; Anders Jönsson
Journal:  BMC Musculoskelet Disord       Date:  2013-01-02       Impact factor: 2.362

8.  Proximal Femoral Nail Unlocked versus Locked (ProFNUL): a protocol for a multicentre, parallel-armed randomised controlled trial for the effect of femoral nail mode of lag screw locking and screw configuration in the treatment of intertrochanteric femur fractures.

Authors:  Arjun Sivakumar; Dominic Thewlis; Andreas Ladurner; Suzanne Edwards; Mark Rickman
Journal:  BMJ Open       Date:  2020-02-10       Impact factor: 2.692

  8 in total