Polyethylene insert damage in unicondylar knee replacement: a comparison of in vivo function and in vitro simulation.

M Harman1, S Affatato, M Spinelli, M Zavalloni, S Stea, A Toni.   

Abstract

Modification of knee joint wear simulation methods has included 'anatomic attachment' of unicondylar knee replacements (UKR) onto synthetic femurs with material properties and morphology similar to human femurs. The present study assesses the effect of such modification by comparing the damage patterns on UKR polyethylene inserts after in vitro simulation using standard and modified simulation methods with those on inserts retrieved after in vivo function. Three groups of UKR inserts were evaluated after retrieval (Explant Group, n = 17) or after knee joint wear simulation with the components attached to standard metal blocks (Standard Group, n = 6) or synthetic femurs (Anatomic Group, n = 6). All UKR had similar non-conforming articular surfaces. Articular damage patterns (mode, frequency, and area) were quantified using digital image photogrammetry. Although some common damage modes were noted, knee joint wear simulation with standard or 'anatomic' attachment did not generate damage pattern sizes similar to the explanted UKR. A focal damage pattern consistent with contact between the metal femoral articular surface and the polyethylene inserts was evident on all inserts, but only the Explant Group had evidence of dispersed damage dominated by abrasive modes. Synthetic femurs added complexity to the wear simulation without generating wear patterns substantially more similar to those observed on retrieved inserts.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20839650     DOI: 10.1243/09544119JEIM699

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  3 in total

1.  A pictographic atlas for classifying damage modes on polyethylene bearings.

Authors:  Melinda Harman; Luca Cristofolini; Paolo Erani; Susanna Stea; Marco Viceconti
Journal:  J Mater Sci Mater Med       Date:  2011-04-02       Impact factor: 3.896

2.  Experimental testing of total knee replacements with UHMW-PE inserts: impact of severe wear test conditions.

Authors:  Carmen Zietz; Joern Reinders; Jens Schwiesau; Alexander Paulus; Jan Philippe Kretzer; Thomas Grupp; Sandra Utzschneider; Rainer Bader
Journal:  J Mater Sci Mater Med       Date:  2015-02-26       Impact factor: 3.896

3.  Analysis of Carbon Fiber Reinforced PEEK Hinge Mechanism Articulation Components in a Rotating Hinge Knee Design: A Comparison of In Vitro and Retrieval Findings.

Authors:  Ronja A Schierjott; Alexander Giurea; Hans-Joachim Neuhaus; Jens Schwiesau; Andreas M Pfaff; Sandra Utzschneider; Gianluca Tozzi; Thomas M Grupp
Journal:  Biomed Res Int       Date:  2016-12-22       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.