Literature DB >> 14646752

Surface damage on open box posterior-stabilized polyethylene tibial inserts.

Mauricio Silva1, Christina A Kabbash, John V Tiberi, Sang Hyun Park, Donald T Reilly, Ormonde M Mahoney, Thomas P Schmalzried.   

Abstract

Twenty retrieved Scorpio posterior-stabilized implants were available for analysis. The mean implantation time was 22 months (range, 2 days-42 months). Favorable types and amounts of surface damage were seen on the tibiofemoral and backsides of these modular PE liners that had been packaged in an inert environment and then sterilized by gamma irradiation. Delamination represented only 0.1% of the total surface damage. Off-axis loading (varus malalignment) was associated with tibial component loosening but there was no evidence of peripheral damage of PE caused by edge loading. With this open-box design, hyperextension marks on the anterior aspect of the posterior-stabilized post from femoral component impingement occurred in 11 of 20 cases and was related to sagittal component positioning: excess tibial slope or increased tibial slope combined with a flexed femoral component. Unique coarse abrasions occurred in 16 of 20 cases and were the result of cement extrusion into the open box, especially with varus malalignment. These observations provide guidance for optimizing the surgical technique and the design of posterior-stabilized total knee components.

Entities:  

Mesh:

Substances:

Year:  2003        PMID: 14646752     DOI: 10.1097/01.blo.0000092998.90435.98

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  5 in total

1.  The John Insall Award: no functional advantage of a mobile bearing posterior stabilized TKA.

Authors:  Ormonde M Mahoney; Tracy L Kinsey; Theresa J D'Errico; Jianhua Shen
Journal:  Clin Orthop Relat Res       Date:  2012-01       Impact factor: 4.176

2.  A pictographic atlas for classifying damage modes on polyethylene bearings.

Authors:  Melinda Harman; Luca Cristofolini; Paolo Erani; Susanna Stea; Marco Viceconti
Journal:  J Mater Sci Mater Med       Date:  2011-04-02       Impact factor: 3.896

3.  5- to 9-year survivorship of single-radius, posterior-stabilized TKA.

Authors:  Ormonde M Mahoney; Tracy L Kinsey
Journal:  Clin Orthop Relat Res       Date:  2008-01-10       Impact factor: 4.176

4.  Does tibial design modification improve implant stability for total knee arthroplasty? An experimental cadaver study.

Authors:  Sebastian Jaeger; Marvin Eissler; Martin Schwarze; Mareike Schonhoff; J Philippe Kretzer; Rudi G Bitsch
Journal:  Bone Joint Res       Date:  2022-04       Impact factor: 4.410

5.  Damages of the tibial post in constrained total knee prostheses in the early postoperative course - a scanning electron microscopic study of polyethylene inlays.

Authors:  Adrian Skwara; Carsten O Tibesku; Rudolf Reichelt; Susanne Fuchs-Winkelmann
Journal:  BMC Musculoskelet Disord       Date:  2008-06-11       Impact factor: 2.362

  5 in total