Investigation of early surface delamination observed in retrieved heat-pressed tibial inserts.

R D Bloebaum1, K Nelson, L D Dorr, A A Hofmann, D J Lyman.   

Abstract

The objective was to examine possible reasons for delamination observed in tibial inserts of the porous-coated anatomic (PCA) knee replacement. To date, 33 PCA inserts have been forwarded to the authors' labs. Of these 33, 52% showed severe delamination within four years of implantation. Visual, structural, and mechanical analyses were conducted and data compared on the heat-pressed PCA type and the common machined inserts. Twenty inserts of the two different types were examined. Visual data using polarized light microscopy showed the presence of a surface layer separated from the middle region of the heat-pressed inserts by a line of demarcation 250-580 microns beneath the articulating surface. This anomaly was not observed in machined inserts. Structural analysis showed the new heat-pressed inserts had increased crystallinity in the surface layer when compared to new machined inserts. The retrieved heat-pressed inserts showed increased crystallinity in the surface and middle regions. There was a slight increase in surface crystallinity in the retrieved machined inserts. Microhardness data showed that there was an increased hardness associated with the crystallinity seen on the heat-pressed inserts. Orthopedic surgeons should be aware of early delamination and surface failure in heat-pressed inserts.

Entities:  

Mesh:

Substances:

Year:  1991        PMID: 1864029

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  4 in total

1.  Surface degradation features and microstructural properties of ultra-high molecular weight polyethylene (UHMWPe).

Authors:  G B Cornwall; C M Hansson; A J Bowe; J T Bryant
Journal:  J Mater Sci Mater Med       Date:  1997-05       Impact factor: 3.896

2.  Highly crosslinked polyethylene is safe for use in total knee arthroplasty.

Authors:  Jeffrey T Hodrick; Erik P Severson; Deborah S McAlister; Brian Dahl; Aaron A Hofmann
Journal:  Clin Orthop Relat Res       Date:  2008-09-10       Impact factor: 4.176

3.  A pictographic atlas for classifying damage modes on polyethylene bearings.

Authors:  Melinda Harman; Luca Cristofolini; Paolo Erani; Susanna Stea; Marco Viceconti
Journal:  J Mater Sci Mater Med       Date:  2011-04-02       Impact factor: 3.896

4.  Patient and surgery related factors associated with fatigue type polyethylene wear on 49 PCA and DURACON retrievals at autopsy and revision.

Authors:  Markus Rohrbach; Martin Lüem; Peter E Ochsner
Journal:  J Orthop Surg Res       Date:  2008-02-22       Impact factor: 2.359

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.