A manufacturer's role in reducing the dose of cone beam computed tomography examinations: effect of beam filtration.

J B Ludlow1.   

Abstract

OBJECTIVES: The dosimetry of the Kodak 9500 cone beam CT (CBCT) unit (Carestream Health, Rochester, NY) was measured before and after installation of copper filtration.
METHODS: Dosimetry of a pre-production Kodak 9500 CBCT unit was compared with a current production unit with 0.4 mm of added filtration and increased kVp. Thermoluminescent dosimeter 100 chips were placed at 24 locations in a RANDO (radiation analogue dosimetry) head phantom (Nuclear Associates, Hicksville, NY). Small, medium and large adult default exposure settings were used in separate dosimeter runs for large and medium field of view (FOV) examinations with both units. Equivalent dose and effective dose were calculated using International Commission on Radiological Protection (ICRP) 1990 and 2007 tissue weights.
RESULTS: Estimations of risk using 2007 ICRP calculations increased by an average of 77% for large FOV scans and 125% for the medium FOV scans in comparison with 1990 calculations. With added filtration, effective dose for medium FOV examinations for default settings were: small adult 76 µSv, medium adult 98 µSv, and large adult 166 µSv. Effective doses for large FOV examinations were: small adult 93 µSv, medium adult 163 µSv, and large adult 260 µSv. Effective dose was reduced by an average of 43% in examinations made with increased filtration and adjusted kVp.
CONCLUSION: The manufacturer's installation of additional filtration with the adjustment of kVp in the Kodak 9500 CBCT unit resulted in significant patient dose reductions for examinations at all adult default settings.

Entities:  

Mesh:

Year:  2011        PMID: 21239575      PMCID: PMC3520301          DOI: 10.1259/dmfr/31708191

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  11 in total

1.  Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit.

Authors:  J B Ludlow; L E Davies-Ludlow; S L Brooks
Journal:  Dentomaxillofac Radiol       Date:  2003-07       Impact factor: 2.419

2.  1990 Recommendations of the International Commission on Radiological Protection.

Authors: 
Journal:  Ann ICRP       Date:  1991

3.  Dosimetry of 3 CBCT devices for oral and maxillofacial radiology: CB Mercuray, NewTom 3G and i-CAT.

Authors:  J B Ludlow; L E Davies-Ludlow; S L Brooks; W B Howerton
Journal:  Dentomaxillofac Radiol       Date:  2006-07       Impact factor: 2.419

4.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

Review 5.  Cone-beam computerized tomography (CBCT) imaging of the oral and maxillofacial region: a systematic review of the literature.

Authors:  W De Vos; J Casselman; G R J Swennen
Journal:  Int J Oral Maxillofac Surg       Date:  2009-05-21       Impact factor: 2.789

6.  Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology.

Authors:  John B Ludlow; Marija Ivanovic
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-05-27

7.  Radiobiologic risk estimation from dental radiology. Part I. Absorbed doses to critical organs.

Authors:  T E Underhill; I Chilvarquer; K Kimura; R P Langlais; W D McDavid; J W Preece; G Barnwell
Journal:  Oral Surg Oral Med Oral Pathol       Date:  1988-07

8.  Absorbed bone marrow dose in certain dental radiographic techniques.

Authors:  S C White; T C Rose
Journal:  J Am Dent Assoc       Date:  1979-04       Impact factor: 3.634

9.  Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer.

Authors:  Rebecca Smith-Bindman; Jafi Lipson; Ralph Marcus; Kwang-Pyo Kim; Mahadevappa Mahesh; Robert Gould; Amy Berrington de González; Diana L Miglioretti
Journal:  Arch Intern Med       Date:  2009-12-14

10.  Patient risk related to common dental radiographic examinations: the impact of 2007 International Commission on Radiological Protection recommendations regarding dose calculation.

Authors:  John B Ludlow; Laura E Davies-Ludlow; Stuart C White
Journal:  J Am Dent Assoc       Date:  2008-09       Impact factor: 3.634

View more
  13 in total

1.  Radiation dose of cone-beam computed tomography compared to conventional radiographs in orthodontics.

Authors:  Luca Signorelli; Raphael Patcas; Timo Peltomäki; Marc Schätzle
Journal:  J Orofac Orthop       Date:  2016-01-08       Impact factor: 1.938

2.  Cone beam CT: a current overview of devices.

Authors:  A Nemtoi; C Czink; D Haba; A Gahleitner
Journal:  Dentomaxillofac Radiol       Date:  2013       Impact factor: 2.419

Review 3.  Effective dose of cone beam CT (CBCT) of the facial skeleton: a systematic review.

Authors:  A Al-Okshi; C Lindh; H Salé; M Gunnarsson; M Rohlin
Journal:  Br J Radiol       Date:  2015-01       Impact factor: 3.039

4.  Automatic fusion of lateral cephalograms and digital volume tomography data-perspective for combining two modalities in the future.

Authors:  S Sauppe; C Abkai; J Hourfar; B Ludwig; J Ulrici; E Hell
Journal:  Dentomaxillofac Radiol       Date:  2015-07-24       Impact factor: 2.419

5.  Effective dose span of ten different cone beam CT devices.

Authors:  D Rottke; S Patzelt; P Poxleitner; D Schulze
Journal:  Dentomaxillofac Radiol       Date:  2013-04-12       Impact factor: 2.419

6.  Effectiveness of thyroid gland shielding in dental CBCT using a paediatric anthropomorphic phantom.

Authors:  A Hidalgo; J Davies; K Horner; C Theodorakou
Journal:  Dentomaxillofac Radiol       Date:  2014-11-20       Impact factor: 2.419

Review 7.  Cone beam computed tomography in paediatric dentistry: overview of recent literature.

Authors:  J K M Aps
Journal:  Eur Arch Paediatr Dent       Date:  2013-06

Review 8.  Effective dose of dental CBCT-a meta analysis of published data and additional data for nine CBCT units.

Authors:  J B Ludlow; R Timothy; C Walker; R Hunter; E Benavides; D B Samuelson; M J Scheske
Journal:  Dentomaxillofac Radiol       Date:  2015       Impact factor: 2.419

9.  Cone-beam computed tomography and radiographs in dentistry: aspects related to radiation dose.

Authors:  Diego Coelho Lorenzoni; Ana Maria Bolognese; Daniela Gamba Garib; Fabio Ribeiro Guedes; Eduardo Franzotti Sant'anna
Journal:  Int J Dent       Date:  2012-04-04

10.  Digital orthodontic radiographic set versus cone-beam computed tomography: an evaluation of the effective dose.

Authors:  Lillian Atsumi Simabuguro Chinem; Beatriz de Souza Vilella; Cláudia Lúcia de Pinho Maurício; Lucia Viviana Canevaro; Luiz Fernando Deluiz; Oswaldo de Vasconcellos Vilella
Journal:  Dental Press J Orthod       Date:  2016 Jul-Aug
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.