Literature DB >> 13679353

Dosimetry of two extraoral direct digital imaging devices: NewTom cone beam CT and Orthophos Plus DS panoramic unit.

J B Ludlow1, L E Davies-Ludlow, S L Brooks.   

Abstract

OBJECTIVES: This study provides effective dose measurements for two extraoral direct digital imaging devices, the NewTom 9000 cone beam CT (CBCT) unit and the Orthophos Plus DS panoramic unit.
METHODS: Thermoluminescent dosemeters were placed at 20 sites throughout the layers of the head and neck of a tissue-equivalent RANDO phantom. Variations in phantom orientation and beam collimation were used to create three different CBCT examination techniques: a combined maxillary and mandibular scan (Max/Man), a maxillary scan and a mandibular scan. Ten exposures for each technique were used to ensure a reliable measure of radiation from the dosemeters. Average tissue-absorbed dose, weighted equivalent dose and effective dose were calculated for each major anatomical site. Effective doses of individual organs were summed with salivary gland exposures (E(SAL)) and without salivary gland exposures (E(ICRP60)) to calculate two measures of whole-body effective dose.
RESULTS: The effective doses for CBCT were: Max/Man scan, E(ICRP60)=36.3 micro Sv, E(SAL)=77.9 micro Sv; maxillary scan, E(ICRP60)=19.9 micro Sv, E(SAL)=41.5 micro Sv; and mandibular scan, E(ICRP60)=34.7 micro Sv, E(SAL)=74.7 micro Sv. Effective doses for the panoramic examination were E(ICRP60)=6.2 micro Sv and E(SAL)=22.0 micro Sv.
CONCLUSION: When viewed in the context of potential diagnostic yield, the E(ICRP60) of 36.3 micro Sv for the NewTom compares favourably with published effective doses for conventional CT (314 micro Sv) and film tomography (2-9 micro Sv per image). CBCT examinations resulted in doses that were 3-7 (E(ICRP60)) and 2-4 (E(SAL)) times the panoramic doses observed in this study.

Entities:  

Mesh:

Year:  2003        PMID: 13679353     DOI: 10.1259/dmfr/26310390

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  97 in total

1.  Necessity of 3D visualization for the removal of lower wisdom teeth: required sample size to prove non-inferiority of panoramic radiography compared to CBCT.

Authors:  Felix Roeder; Daniel Wachtlin; Ralf Schulze
Journal:  Clin Oral Investig       Date:  2011-04-26       Impact factor: 3.573

2.  Characterization of optically stimulated luminescence dosemeters to measure organ doses in diagnostic radiology.

Authors:  A Endo; T Katoh; I Kobayashi; R Joshi; J Sur; T Okano
Journal:  Dentomaxillofac Radiol       Date:  2011-11-24       Impact factor: 2.419

3.  Effective doses from cone beam CT investigation of the jaws.

Authors:  J Davies; B Johnson; Na Drage
Journal:  Dentomaxillofac Radiol       Date:  2012-01       Impact factor: 2.419

4.  Comparative study between conventional and cone beam CT-synthesized half and total skull cephalograms.

Authors:  G S Liedke; E L Delamare; M B Vizzotto; H L D da Silveira; J R Prietsch; V Dutra; H E D da Silveira
Journal:  Dentomaxillofac Radiol       Date:  2012-02       Impact factor: 2.419

5.  Reproducibility of three-dimensional cephalometric landmarks in cone-beam and low-dose computed tomography.

Authors:  R Olszewski; L Frison; M Wisniewski; J M Denis; S Vynckier; G Cosnard; F Zech; H Reychler
Journal:  Clin Oral Investig       Date:  2012-02-17       Impact factor: 3.573

6.  Digital method for quantification of circumferential periodontal bone level using cone beam CT.

Authors:  Jonathan Fleiner; Christian Hannig; Dirk Schulze; Andres Stricker; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2012-03-21       Impact factor: 3.573

7.  Accuracy of templates for navigated implantation made by rapid prototyping with DICOM datasets of cone beam computer tomography (CBCT).

Authors:  Jochen Weitz; Herbert Deppe; Sebastian Stopp; Tim Lueth; Steffen Mueller; Bettina Hohlweg-Majert
Journal:  Clin Oral Investig       Date:  2010-09-21       Impact factor: 3.573

Review 8.  Modern dental imaging: a review of the current technology and clinical applications in dental practice.

Authors:  Bart Vandenberghe; Reinhilde Jacobs; Hilde Bosmans
Journal:  Eur Radiol       Date:  2010-06-11       Impact factor: 5.315

9.  Radiation exposure to foetus and breasts from dental X-ray examinations: effect of lead shields.

Authors:  Anna Kelaranta; Marja Ekholm; Paula Toroi; Mika Kortesniemi
Journal:  Dentomaxillofac Radiol       Date:  2015-08-27       Impact factor: 2.419

10.  Assessment of phantom dosimetry and image quality of i-CAT FLX cone-beam computed tomography.

Authors:  John B Ludlow; Cameron Walker
Journal:  Am J Orthod Dentofacial Orthop       Date:  2013-12       Impact factor: 2.650