Abstract
Introduction. The aim of this study was to discuss the radiation doses associated with plain radiographs, cone-beam computed tomography (CBCT), and conventional computed tomography (CT) in dentistry, with a special focus on orthodontics. Methods. A systematic search for articles was realized by MEDLINE from 1997-March 2011. Results. Twenty-seven articles met the established criteria. The data of these papers were grouped in a table and discussed. Conclusions. Increases in kV, mA, exposure time, and field of view (FOV) increase the radiation dose. The dose for CT is greater than other modalities. When the full-mouth series (FMX) is performed with round collimation, the orthodontic radiographs transmit higher dose than most of the large FOV CBCT, but it can be reduced if used rectangular collimation, showing lower effective dose than large FOV CBCT. Despite the image quality, the CBCT does not replace the FMX. In addition to the radiation dose, image quality and diagnostic needs should be strongly taken into account.Entities:
Year: 2012 PMID: 22548064 PMCID: PMC3324904 DOI: 10.1155/2012/813768
Source DB: PubMed Journal: Int J Dent ISSN: 1687-8728
Tissue-weighting factors for calculation of effective radiation dose.
| Tissue | ICRP 1990 | ICRP 2005 | ICRP 2007 |
|---|---|---|---|
|
| 0.12 | 0.12 | 0.12 |
| Breast | 0.05 | 0.12 | 0.12 |
| Colon | 0.12 | 0.12 | 0.12 |
| Lung | 0.12 | 0.12 | 0.12 |
| Stomach | 0.12 | 0.12 | 0.12 |
| Gonads | 0.20 | 0.05 | 0.08 |
|
| 0.05 | 0.05 | 0.04 |
| Bladder | 0.05 | 0.05 | 0.04 |
| Liver | 0.05 | 0.05 | 0.04 |
|
| 0.05 | 0.05 | 0.04 |
|
| 0.01 | 0.01 | 0.01 |
|
| RT | 0.01 | 0.01 |
|
| 0.01 | 0.01 | 0.01 |
|
| Not included | 0.01 | 0.01 |
| Kidney | RT | 0.01 | RT |
|
| 0.05a | 0.10b | 0.12c |
RT: Remainder tissues; aadrenals/brain/upper large intestine/small intestine/kidney/muscle/pancreas/spleen/thymus/uterus
bAdipose tissue/adrenals/connective tissue/extrathoracic airways/gallbladder/heart wall/lymphatic nodes/muscle/pancreas/prostate/spleen/thymus/uterus/cervix
cAdrenals/extrathoracic region/gallbladder/heart/prostate/kidneys/small intestine/lymphatic nodes/oral mucosa/muscle/pancreas/spleen/thymus/uterus/cervix (text in boldface represents tissues used for calculation of maxillofacial dose).
Figure 1Flow chart of the search process.
Effective doses. (ExcGland or IncGland: salivary glands excluded or included; Mx: Maxilla; Md: Mandible).
| Exams/equipment/adjustment provided | Effective Dose ( | |||
|---|---|---|---|---|
| ICRP 60-1990 | ICRP 2005 | ICRP 103-2007 | ||
| ExcGland | IncGland | |||
| PANORAMIC RADIOGRAPHS | ||||
|
| ||||
| PM2002CCProlinePlanmeca/70 kVp/7 mA/18 s [ | 3.8 | |||
| VeraviewepocsMorita 77 kV/5 mA/8.1 s [ | 5.2 | |||
| OrthophosSiemens/62 kV/16 mA/14.1 s [ | 9 | 16.4 | ||
| PM2002CCProLinePlanmeca/64 kV/6 mA/15 s [ | 4 | 9 | ||
| PromaxPlanmeca/66 kV/6 mA/16 s [ | 17 | 26 | ||
| PM2002CCProlinePlanmeca/73 kV/5 mA/15 s [ | 10 | |||
| Digital/PM2002CCProline2000Planmeca/66 kV/4 mA/18 s [ | 8 | 12 | ||
| Digital/PM2002CCProline2000Planmeca/66 kV/8 mA/18 s [ | 23 | 38 | ||
| Digital/CranexExcelSoredex/65 kV/6 mA/19 s [ | 4.5 | 12.3 | ||
| Digital/Verawiewepocs5DMorita/70 kV/4 mA/8.2 s [ | 2.5 | 5.5 | ||
| Digital/ECProlinePlanmeca/64 kV/7 mA/18.3 s [ | 5.7 | 14.9 | ||
| Digital/Orthoralix9200DDEGendex/74 kV/4 mA/12 s [ | 2.4 | 4.7 | ||
| Digital/ProMaxPlanmeca/Adult [ | 20 | 23 | ||
| Digital/ProMaxPlanmeca/68 kV/13 mA/16 s [ | 7.1 | 24.3 | ||
| Digital/OrthophosXGSirona/64 kV/8 mA/14.1 s [ | 4.3 | 14.2 | ||
| Digital/OrthophosPlusDSSirona/66 kVp/16 mA/14.1 s [ | 6.2 | 22 | ||
| Digital/VeraviewepocsMorita/67 kV/5 mA/8.1 s [ | 2.7 | |||
| Digital/Veraviewepocs3DMorita/70 kV/5 mA/7.4 s [ | 2.9 | |||
| Digital/CranexTomeSoredex/70 kV/4 mA/15 s [ | 3.3 | 8.1 | ||
|
| ||||
| LATERAL CEPHALOMETRIC RADIOGRAPHS | ||||
|
| ||||
| OrthophosCSiemens/77 kV/14 mA/0.5 s [ | 2.3 | |||
| PM2002CCProLinePlanmeca/70 kV/12 mA/0.9 s [ | 2 | 3 | ||
| CranexTomeSoredex/70 kVp/10 mA/0.4 s [ | 3 | 3.7 | ||
| CranexTomeSoredex/Collimation/70 kVp/10 mA/0.4 s [ | 1.6 | 2.2 | ||
| PM2002CCProlinePlanmeca/80 kV/12 mA/0.5 s [ | 5 | |||
| Digital/OrthophosDSCephSiemens/73 kV/15 mA/15.8 s [ | 1.1 | |||
| Digital/ProLineCephCMPlanmeca/Collimation/70 kVp/10 mA/23 s [ | 1.7 | 3.4 | ||
| Digital/CranexTomeSoredex/Collimation/70 kVp/4 mAs [ | 1.6 | 2.2 | ||
| Digital/InterayVarian/77 kVp/6.5 mAs [ | 3.7 | 5.6 | ||
|
| ||||
| PA CEPHALOMETRIC RADIOGRAPHS | ||||
|
| ||||
| Digital/InterayVarian/75 kVp/11 mAs [ | 3.9 | 5.1 | ||
|
| ||||
| INTRAORAL RADIOGRAPHS | ||||
| IntraPlanmeca/FullMouthRadiographs/70 kV/8 mA/Digital or F-speed film/RectangularCollimation [ | 12.2 | 34.9 | ||
| IntraPlanmeca/FullMouthRadiographs/70 kV/8 mA/Digital or F-speed film/RoundCollimation [ | 58.4 | 170.7 | ||
| IntraPlanmeca/FullMouthRadiographs/RoundCollimation/Adult [ | 115 | 129 | ||
| IntraPlanmeca/Bitewing(04)/70 kV/8 mA/Digital or F-speed film/RectangularCollimation [ | 1 | 5 | ||
| SiemensHeliodent70Dentotime/OcclusalMx [ | 7 | |||
|
| ||||
| LARGE FOV CONE BEAM CT | ||||
|
| ||||
| Classic i-CAT/FOV22 cm/120 kV/3–8 mA [ | 92.8 | 182.1 | ||
| Classic i-CAT/FOV22 cm/120 kV/5.7 mA [ | 134.8 | 193.4 | ||
| Classic i-CAT/FOV22 cm/120 kV/3–8 mA/2 × 20 s [ | 82 | |||
| Next Generation i-CAT/FOV23 cm/120 kV/5 mA/19 mAs/8.9 s [ | 37 | 74 | ||
| NewTom3G/FOV19 cm/110 kV/1.5 mA/8.09 mAs/36 s [ | 44.7 | 58.9 | 68 | |
| NewTom3G/FOV19 cm/110 kV/<15 mA [ | 30 | |||
| NewTom9000/FOV23 cm/110 kV/5.4 mA [ | 56.2 | |||
| CBMercuRay/FOV19 cm/100 kV/10 mA/100 mAs/10 s [ | 476.6 | 557.6 | 569 | |
| CBMercuRay/FOV19 cm/120 kV/15 mA/150 mAs/10 s [ | 846.9 | 1025.4 | 1073 | |
| CBMercuRay/FOV19 cm/100 kV/15 mA [ | 415 | 479 | ||
| CBMercuRay/FOV19 cm/120 kV/15 mA [ | 656 | 761 | ||
| CBMercuRay/FOV19 cm/100 kV/10 mA [ | 264 | 306 | ||
| CBMercuRay/FOV19 cm/100 kV/5 mA [ | 153 | 177 | ||
| CBMercuRay/FOV19 cm/100 kV/2 mA [ | 75 | 86 | ||
| Iluma/FOV19 cm/120 kV/1 mA/20 mAs/20 s [ | 50 | 98 | ||
| Iluma/FOV19 cm/120 kV/3.8 mA/152 mAs/40 s [ | 252 | 498 | ||
| Kodak9500/FOV18 cm/80 kV/86.4 mAs [ | 52 | 93 | ||
| Kodak9500/FOV18 cm/85 kV/108 mAs [ | 92 | 163 | ||
| Kodak9500/FOV18 cm/90 kV/108 mAs [ | 148 | 260 | ||
| Kodak9500/FOV18 cm/90 kV/108 mAs [ | 136 | |||
| SkyView/FOV17 cm/90 kV/51 mAs [ | 87 | |||
|
| ||||
| MEDIUM FOV CONE BEAM CT | ||||
|
| ||||
| Classic i-CAT/FOV13 cm/120 kV/3–8 mA [ | 39.5 | 110.5 | ||
| Classic i-CAT/FOV13 cm/120 kV/5.7 mA [ | 68.7 | 104.5 | ||
| Classic i-CAT/FOV13 cm/120 kV/23.87 mA [ | 61.1 | |||
| Classic i-CAT/FOV13 cm/120 kV/3–8 mA/10 s [ | 48 | |||
| Classic i-CAT/FOV13 cm/120 kV/3–8 mA/40 s [ | 77 | |||
| Classic i-CAT/FOV13 cm/120 kV/5 mA/19 mAs/20 s [ | 29 | 69 | ||
| Next Generation i-CAT/FOV13 cm/120 kV/5 mA/19 mAs/8.9 s [ | 36 | 87 | ||
| Next Generation i-CAT/FOV13 cm/120 kV/18.5 mAs [ | 83 | |||
| NewTom9000/FOV13 cm/110 kV/3.2 mA [ | 36.9 | 77.9 | ||
| NewTom9000/FOV13 cm/110 kV/3.5 mA/18 s [ | 50.3 | |||
| NewTom9000/FOV13 cm/110 kV/3.4 mA/17 s [ | 35 | 64 | ||
| NewTom9000/FOV13 cm/110 kV/3.4 mA/17 s/Thyroid Protector [ | 23 | 52 | ||
| NewTom3G/FOV15 cm/110 kV/<15 mA [ | 57 | |||
| NewTom5Gi/FOV15 cm/110 kV/8.8 mAs [ | 194 | |||
| CBMercuRay/FOV15 cm/120 kV/15 mA/120/mAs/10 s [ | 288.9 | 435.5 | 560 | |
| CBMercuRay/FOV15 cm/100 kV/15 mA [ | 354 | 402 | ||
| CBMercuRay/FOV15 cm/120 kV/15 mA [ | 601 | 680 | ||
| Galileos/FOV15 cm/85 kV/5 mA/21 mAs/14 s [ | 28 | 70 | ||
| Galileos/FOV15 cm/85 kV/7 mA/42 mAs/14 s [ | 52 | 128 | ||
| GalileosComfort/FOV15 cm/85 kV/28 mAs [ | 84 | |||
| Kodak9500/FOV15 cm/80 kV/86.4 mAs [ | 39 | 76 | ||
| Kodak9500/FOV15 cm/85 kV/108 mAs [ | 49 | 98 | ||
| Kodak9500/FOV15 cm/90 kv/108 mAs [ | 76 | 166 | ||
| IlumaElite/FOV14 cm/120 kV/76 mAs [ | 368 | |||
| Scanora3D/FOV13.5 cm/85 kV/48 mAs [ | 68 | |||
|
| ||||
| SMALL FOV CONE BEAM CT | ||||
|
| ||||
| Classic i-CAT/FOV6 cmMx/120 kV/3–8 mA [ | 9.7 | 36.5 | ||
| Classic i-CAT/FOV6 cmMx/120 kV/3–8 mA/HighResolution [ | 18.5 | 68.3 | ||
| Classic i-CAT/FOV6 cmMx/120 kV/3–8 mA/20 s [ | 45 | |||
| Classic i-CAT/FOV6 cmMx/120 kV/3–8 mA/40 s [ | 77 | |||
| Classic i-CAT/FOV6 cmMd/120 kV/3–8 mA [ | 23.9 | 75.3 | ||
| Classic i-CAT/FOV6 cmMd/120 kV/3–8 mA/HighResolution [ | 47.2 | 148.5 | ||
| Classic i-CAT/FOV6 cmMd/120 kV/3–8 mA/20 s [ | 34 | |||
| Classic i-CAT/FOV6 cmMd/120 kV/3–8 mA/40 s [ | 64 | |||
| Classic i-CAT/FOV8 cm/120 kV/3–8 mA/40 s [ | 37 | |||
| Next Generation i-CAT/FOV6 cmMd/120 kV/18.5 mAs [ | 45 | |||
| NewTom9000/FOVMx [ | 19.9 | 41.5 | ||
| NewTom9000/FOVMd [ | 34.7 | 74.7 | ||
| NewTom5G/FOV10 cm/110 kV/10.4 mAs [ | 83 | |||
| NewTom5Gi/FOV8 cm/110 kV/43 mAs [ | 265 | |||
| CBMercuRay/FOV10 cmMx/120 kV/15 mA/150 mAs/10 s [ | 168.4 | 283.3 | 407 | |
| CBMercuRay/FOV10 cm/100 kV/15 mA [ | 328 | 369 | ||
| CBMercuRay/FOV10 cm/120 kV/15 mA [ | 535 | 603 | ||
| CBMercuRay/FOV10 cm/120 kV/15 mA [ | 451.8 | 510.5 | ||
| Promax3D/FOV8 cm/84 kVp/12 mA/6 s [ | 269 | 674 | ||
| Promax3D/FOV8 cm/84 kV/12 mA/72 mAs/18 s [ | 151 | 488 | ||
| Promax3D/FOV8 cm/84 kV/16 mA/96 mAs/18 s [ | 203 | 652 | ||
| Promax3D/FOV8 cm/84 kV/8 mA/12 s/NormalResolution [ | 102 | |||
| Promax3D/FOV8 cm/84 kV/10 mA/12 s/NormalResolution [ | 169 | |||
| Promax3D/FOV8 cm/84 kV/12 mA/12 s/NormalResolution [ | 216 | |||
| Promax3D/FOV8 cm/84 kV/14 mA/12 s/NormalResolution [ | 272 | |||
| Promax3D/FOV8 cm/84 kV/16 mA/12 s/NormalResolution [ | 298 | |||
| Promax3D/FOV8 cm/84 kV/8 mA/2.8 s/LowDose [ | 30 | |||
| Promax3D/FOV8 cm/84 kV/16 mA/12 s/HighDose [ | 306 | |||
| Promax3D/FOV8 cm/84 kV/8 mA/8.3 s/LowDose [ | 87 | |||
| Promax3D/FOV8 cm/84 kV/169 mAs/HighDose [ | 122 | |||
| Promax3D/FOV8 cm/84 kV/19.9 mAs/LowDose [ | 28 | |||
| PreXion3D/FOV8.1 cm/90 kV/4 mA/76 mAs/19 s [ | 66 | 189 | ||
| PreXion3D/FOV8.1 cm/90 kV/4 mA/148 mAs/37 s [ | 154 | 388 | ||
| 3D Accuitomo 170/FOV5 cmMx/90 kV/87.5 mAs [ | 54 | |||
| Kodak9500/FOV8 cm/90 kV/108 mAs [ | 92 | |||
| PicassoTrio HighDose/FOV7 cm/85 kV/127 mAs [ | 123 | |||
| PicassoTrio LowDose/FOV7 cm/85 kV/91 mAs [ | 81 | |||
| Scanora 3D/FOV7.5 cmMx/85 kV/30 mAs [ | 46 | |||
| Scanora 3D/FOV7.5 cmMd/85 kV/30 mAs [ | 47 | |||
| Scanora 3D/FOV7.5 cmMxMd/85 kV/30 mAs [ | 45 | |||
| Veraviewpocs3D/FOV8 cm/70 kV/51 mAs [ | 73 | |||
|
| ||||
| CONVENTIONAL CT | ||||
|
| ||||
| SomatomVolumeZoom4/Scan22.6 cmFullHead/120 kV/ 90 mA/44.12 s/Slice0.75 mm [ | 1110 | |||
| SomatomSensation16/Scan22.5 cmFullHeadl/120 kV/90 mA/29.48 s/slice0.75 mm [ | 995 | |||
| Mx8000IDTPhilips/Scan22.5 cmFullHead/120 kV/140 mA/29.6 s/Slice0.75 mm [ | 1160 | |||
| Somatom64/Scan12 cm/120 kV/90 mA [ | 453 | 860 | ||
| Somatom64CareDose4D/Scan12 cm/120 kV/90 mA [ | 285 | 534 | ||
| SomatomPlusVolumeZoom4/ScanMx+Md/Slice1.25 mm/21.25 s/120 kVp/150 mA [ | 2110 | |||
| SomatomSensation/Scan10 cm/120 kV/90 mA [ | 429.7 | |||
| ExcelTwin/Scan9.6 cm/120 kV/300 mAs/Slice2 mm/2sporslice [ | 314 | 924 | ||
| HiSpeedQX/i/Scan7.7 cmMx+Md/120 kV/100 mA [ | 595.6 | 768.9 | ||
| SomatomVolumeZoom4/Scan7.2 cmMd/120 kV/90 mA/15.16 s/Slice0.75 mm [ | 494 | |||
| SomatomSensation16/Scan6.3 cmMd/120 kV/90 mA/7.87 s/Slice0.75 m [ | 474 | |||
| Mx8000IDTPhilips/Scan6 cmMd/120 kV/140 mA/7.89 s/Slice0.75 mm [ | 541 | |||
| SomatomPlus4VolumeZoom Scan5.2 cmMd/120 kV/100 mAs [ | 250 | |||
| ElscintExcel2400/ScanMd/120 kVp/315 mAs [ | 2426 | 3324 | ||
| SomatomPlus4VolumeZoom/ScanMd/Slice1.25 mm/12.64 s/120 kVp/150 mA [ | 1320 | |||
| SomatomPlusVolumeZoom4/ScanMx/Slice1.25 mm/9.47 s/ 120 kVp/150 mA [ | 1400 | |||
| ElscintExcel2400/ScanMx/120 kVp/315 mAs [ | 1031 | 1202 | ||