Literature DB >> 20953784

Pulsed lavage improves fixation strength of cemented tibial components.

Ulf J Schlegel1, Jan Siewe, Karl S Delank, Peer Eysel, Klaus Püschel, Michael M Morlock, Anne Gebert de Uhlenbrock.   

Abstract

Pulsatile lavage is purported to improve radiographic survival in cemented total knee arthroplasty (TKA). Similarly, a potential improvement of fixation strength of the tibial tray has been assumed based on the increased cement penetration. In this study, the influence of pulsed lavage on fixation strength of the tibial component and bone cement penetration was evaluated in six pairs of cadaveric specimens. Following surgical preparation, the tibial surface was irrigated using pulsatile lavage on one side of a pair, while on the other side syringe lavage was applied. All tibial components were implanted using the same cementing technique. Cement penetration and bone mineral density was assessed based on computed tomography data. Fixation strength of the tibial trays was determined by a pull-out test with a material testing machine. Median pull-out forces and cement penetration were significantly (p = 0.031) improved in the pulsed lavage group as compared to the syringe lavage group. Enhanced fixation strength is suggested as being a key to improved survival of the implant. Consequently, pulsatile lavage should be considered as a mandatory preparation step when cementing tibial components in TKA.

Mesh:

Substances:

Year:  2010        PMID: 20953784      PMCID: PMC3167426          DOI: 10.1007/s00264-010-1137-y

Source DB:  PubMed          Journal:  Int Orthop        ISSN: 0341-2695            Impact factor:   3.075


  31 in total

1.  Modular fixed-bearing total knee arthroplasty with retention of the posterior cruciate ligament. A study of patients followed for a minimum of fifteen years.

Authors:  Michael C Dixon; Richard R Brown; Dominik Parsch; Richard D Scott
Journal:  J Bone Joint Surg Am       Date:  2005-03       Impact factor: 5.284

2.  Mechanical bond strength of the cement-tibial component interface in total knee arthroplasty.

Authors:  Gavin T Pittman; Christopher L Peters; Jerod L Hines; Kent N Bachus
Journal:  J Arthroplasty       Date:  2006-09       Impact factor: 4.757

3.  Radiolucent lines below the tibial component of a total knee replacement (TKR)--a comparison between single-and two-stage cementation techniques.

Authors:  Abhijit Ranjan Guha; Ujjwal Kanti Debnath; Niall McKenzie Graham
Journal:  Int Orthop       Date:  2007-03-16       Impact factor: 3.075

4.  Radiolucency and migration after Oxford unicompartmental knee arthroplasty.

Authors:  Paul Rea; Andrew Short; Hemant Pandit; Andrew J Price; Peter Kyberd; David J Beard; Harinderjit S Gill; David W Murray
Journal:  Orthopedics       Date:  2007-05       Impact factor: 1.390

5.  An experimental comparison of different devices for pulsatile high-pressure lavage and their relevance to cement intrusion into cancellous bone.

Authors:  Thomas Kalteis; Dominik Pförringer; Thomas Herold; Martin Handel; Tobias Renkawitz; Wolfgang Plitz
Journal:  Arch Orthop Trauma Surg       Date:  2007-09-01       Impact factor: 3.067

6.  The effect of cement gun and cement syringe use on the tibial cement mantle in total knee arthroplasty.

Authors:  Michael J Lutz; Paul F Pincus; Sarah L Whitehouse; Brett R Halliday
Journal:  J Arthroplasty       Date:  2008-04-02       Impact factor: 4.757

7.  The Coventry Award. Modular tibial insert micromotion. A concern with contemporary knee implants.

Authors:  N L Parks; G A Engh; L D Topoleski; J Emperado
Journal:  Clin Orthop Relat Res       Date:  1998-11       Impact factor: 4.176

8.  Tensile strength of the cement-bone interface depends on the amount of bone interdigitated with PMMA cement.

Authors:  K A Mann; D C Ayers; F W Werner; R J Nicoletta; M D Fortino
Journal:  J Biomech       Date:  1997-04       Impact factor: 2.712

9.  Follow-up of 11-16 years after modular fixed-bearing TKA.

Authors:  D Parsch; M Krüger; M T Moser; F Geiger
Journal:  Int Orthop       Date:  2008-04-05       Impact factor: 3.075

10.  Midterm results of all-polyethylene tibial components in primary total knee arthroplasty.

Authors:  David F Dalury; Donald L Pomeroy; Ricardo A Gonzales; Thomas A Gruen; Mary Jo Adams; Janene A Empson
Journal:  J Arthroplasty       Date:  2008-06-11       Impact factor: 4.757

View more
  17 in total

1.  Loss of cement-bone interlock in retrieved tibial components from total knee arthroplasties.

Authors:  Mark A Miller; Jacklyn R Goodheart; Timothy H Izant; Clare M Rimnac; Richard J Cleary; Kenneth A Mann
Journal:  Clin Orthop Relat Res       Date:  2013-08-24       Impact factor: 4.176

2.  Comparison of different cement application techniques for tibial component fixation in TKA.

Authors:  Ulf J Schlegel; Nicholas E Bishop; Klaus Püschel; Michael M Morlock; Katrin Nagel
Journal:  Int Orthop       Date:  2014-08-01       Impact factor: 3.075

3.  An in vitro comparison of tibial tray cementation using gun pressurization or pulsed lavage.

Authors:  Ulf J Schlegel; Klaus Püschel; Michael M Morlock; Katrin Nagel
Journal:  Int Orthop       Date:  2014-03-12       Impact factor: 3.075

4.  Pulsatile Lavage During Cementation of Total Knee Arthroplasty - Is Fixation Impaired? A Cadaver Study.

Authors:  Claudio Glowalla; Max Ertl; Ulrich Lenze; Igor Lazic; Rainer Burgkart; Jan J Lang; Rudiger VON Eisenhart-Rothe; Florian Pohlig
Journal:  In Vivo       Date:  2022 Mar-Apr       Impact factor: 2.155

5.  Cementation technique for elbow arthroplasty; an international survey.

Authors:  Tariq A Kwaees; Rohit Singhal; Denise Eygendaal; Charalambos P Charalambous
Journal:  J Orthop       Date:  2019-07-01

6.  Mechanical properties of a cemented porous implant interface.

Authors:  Nicholas A Beckmann; Rudi G Bitsch; Joern B Seeger; Matthias Cm Klotz; Jan Philippe Kretzer; Sebastian Jaeger
Journal:  Acta Orthop       Date:  2014-05-05       Impact factor: 3.717

7.  Effect of tibial tray design on cement morphology in total knee arthroplasty.

Authors:  Ulf J Schlegel; Klaus Püschel; Michael M Morlock; Katrin Nagel
Journal:  J Orthop Surg Res       Date:  2014-11-29       Impact factor: 2.359

8.  Tibial cleaning method for cemented total knee arthroplasty: An experimental study.

Authors:  Peter Helwig; Lukas Konstantinidis; Anja Hirschmüller; Verena Miltenberger; Kerstin Kuminack; Norbert P Südkamp; Oliver Hauschild
Journal:  Indian J Orthop       Date:  2013-01       Impact factor: 1.251

9.  Micromotion at the tibial plateau in primary and revision total knee arthroplasty: fixed versus rotating platform designs.

Authors:  S R Small; R D Rogge; R A Malinzak; E M Reyes; P L Cook; K A Farley; M A Ritter
Journal:  Bone Joint Res       Date:  2016-04       Impact factor: 5.853

10.  Biomechanical effects of morphological variations of the cortical wall at the bone-cement interface.

Authors:  Chun-Lin Zhang; Guo-Qi Shen; Kun-Peng Zhu; Dong-Xu Liu
Journal:  J Orthop Surg Res       Date:  2016-07-01       Impact factor: 2.359

View more