Evaluating the clinical accuracy of GlucoMen®Day: a novel microdialysis-based continuous glucose monitor.

Francesco Valgimigli1, Fausto Lucarelli, Cosimo Scuffi, Sara Morandi, Iolanda Sposato.   

Abstract

BACKGROUND: The objective of this work was to determine the clinical accuracy of GlucoMen®Day, a new microdialysis-based continuous glucose monitoring system (CGMS) from A. Menarini Diagnostics (Florence, Italy). Accuracy evaluation was performed using continuous glucose-error grid analysis (CG-EGA), as recommended by the Performance Metrics for Continuous Interstitial Glucose Monitoring; Approved Guideline (POCT05-A).
METHODS: Two independent clinical trials were carried out on patients with types 1 and 2 diabetes mellitus, the glycemic levels of whom were monitored in an in-home setting for 100-hour periods. A new multiparametric algorithm was developed and used to compensate in real-time the GlucoMen®Day signal. The time lag between continuous glucose monitoring (CGM) and reference data was first estimated using the Poincaré plot method. The entire set of CGM/reference data pairs was then evaluated following the CG-EGA criteria, which allowed an estimation of the combined point and rate accuracy stratified by glycemic ranges.
RESULTS: With an estimated time lag of 11 minutes, the linear regression analysis of the CGM/reference glucose values yielded r = 0.92. The mean absolute error (MAE) was 11.4 mg/dl. The calculated mean absolute rate deviation (MARD) was 0.63 mg/dl/min. The data points falling within the A+B zones of CG-EGA were 100% in hypoglycemia, 95.7% in euglycemia, and 95.2% in hyperglycemia.
CONCLUSIONS: The GlucoMen®Day system provided reliable, real-time measurement of subcutaneous glucose levels in patients with diabetes for up to 100 hours. The device showed the ability to follow rapid glycemic excursions and detect severe hypoglycemic events accurately. Its accuracy parameters fitted the criteria of the state-of-the-art consensus guideline for CGMS, with highly consistent results from two independent studies.
© 2010 Diabetes Technology Society.

Entities:  

Mesh:

Substances:

Year:  2010        PMID: 20920438      PMCID: PMC2956803          DOI: 10.1177/193229681000400517

Source DB:  PubMed          Journal:  J Diabetes Sci Technol        ISSN: 1932-2968


  27 in total

1.  A microdialysis technique for continuous subcutaneous glucose monitoring in diabetic patients (part 1).

Authors:  A Poscia; M Mascini; D Moscone; M Luzzana; G Caramenti; P Cremonesi; F Valgimigli; C Bongiovanni; M Varalli
Journal:  Biosens Bioelectron       Date:  2003-07       Impact factor: 10.618

2.  A microdialysis technique for continuous subcutaneous glucose monitoring in diabetic patients (part 2).

Authors:  M Varalli; G Marelli; A Maran; S Bistoni; M Luzzana; P Cremonesi; G Caramenti; F Valgimigli; Alessandro Poscia
Journal:  Biosens Bioelectron       Date:  2003-07       Impact factor: 10.618

3.  Continuous glucose monitoring and type 1 diabetes.

Authors:  Martina Montagnana; Giuseppe Lippi; Gian Cesare Guidi
Journal:  N Engl J Med       Date:  2009-01-08       Impact factor: 91.245

4.  Intensive insulin therapy in the intensive care unit: assessment by continuous glucose monitoring.

Authors:  Christophe De Block; Begoña Manuel-Y-Keenoy; Luc Van Gaal; Peter Rogiers
Journal:  Diabetes Care       Date:  2006-08       Impact factor: 19.112

5.  A review of the biocompatibility of implantable devices: current challenges to overcome foreign body response.

Authors:  Yoshinori Onuki; Upkar Bhardwaj; Fotios Papadimitrakopoulos; Diane J Burgess
Journal:  J Diabetes Sci Technol       Date:  2008-11

6.  Continuous subcutaneous glucose monitoring in children with type 1 diabetes.

Authors:  H P Chase; L M Kim; S L Owen; T A MacKenzie; G J Klingensmith; R Murtfeldt; S K Garg
Journal:  Pediatrics       Date:  2001-02       Impact factor: 7.124

7.  Evaluating clinical accuracy of systems for self-monitoring of blood glucose.

Authors:  W L Clarke; D Cox; L A Gonder-Frederick; W Carter; S L Pohl
Journal:  Diabetes Care       Date:  1987 Sep-Oct       Impact factor: 19.112

8.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

9.  Toward continuous glucose monitoring with planar modified biosensors and microdialysis. Study of temperature, oxygen dependence and in vivo experiment.

Authors:  Francesco Ricci; Felice Caprio; Alessandro Poscia; Francesco Valgimigli; Dimitri Messeri; Elena Lepori; Giorgio Dall'Oglio; Giuseppe Palleschi; Danila Moscone
Journal:  Biosens Bioelectron       Date:  2006-09-26       Impact factor: 10.618

10.  Pre- and postoperative accuracy and safety of a real-time continuous glucose monitoring system in cardiac surgical patients: a randomized pilot study.

Authors:  Susan J Logtenberg; Nanne Kleefstra; Ferdinand T Snellen; Klaas H Groenier; Robbert J Slingerland; Arno P Nierich; Henk J Bilo
Journal:  Diabetes Technol Ther       Date:  2009-01       Impact factor: 6.118

View more
  17 in total

1.  Analysis of GlucoMen®Day: a novel microdialysis-based continuous glucose monitor.

Authors:  Thomas Kubiak
Journal:  J Diabetes Sci Technol       Date:  2010-09-01

2.  Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel.

Authors:  Guido Freckmann; Stefan Pleus; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2013-07-01

Review 3.  Intensive insulin therapy in critically ill hospitalized patients: making it safe and effective.

Authors:  David C Klonoff
Journal:  J Diabetes Sci Technol       Date:  2011-05-01

4.  Minimizing the impact of time lag variability on accuracy evaluation of continuous glucose monitoring systems.

Authors:  Cosimo Scuffi; Fausto Lucarelli; Francesco Valgimigli
Journal:  J Diabetes Sci Technol       Date:  2012-11-01

5.  Periodic extraction of interstitial fluid from the site of subcutaneous insulin infusion for the measurement of glucose: a novel single-port technique for the treatment of type 1 diabetes patients.

Authors:  Werner Regittnig; Stefan Lindpointner; Stefan Korsatko; Dina Tutkur; Manfred Bodenlenz; Thomas R Pieber
Journal:  Diabetes Technol Ther       Date:  2012-11-05       Impact factor: 6.118

6.  Use of microdialysis-based continuous glucose monitoring to drive real-time semi-closed-loop insulin infusion.

Authors:  Guido Freckmann; Nina Jendrike; Stefan Pleus; Harvey Buck; Steven Bousamra; Paul Galley; Ajay Thukral; Robin Wagner; Stefan Weinert; Cornelia Haug
Journal:  J Diabetes Sci Technol       Date:  2014-09-09

Review 7.  Continuous Glucose Monitoring, Future Products, and Update on Worldwide Artificial Pancreas Projects.

Authors:  Jort Kropff; J Hans DeVries
Journal:  Diabetes Technol Ther       Date:  2016-02       Impact factor: 6.118

8.  Interstitium versus Blood Equilibrium in Glucose Concentration and its Impact on Subcutaneous Continuous Glucose Monitoring Systems.

Authors:  Cosimo Scuffi
Journal:  Eur Endocrinol       Date:  2014-02-28

9.  GlucoMen Day continuous glucose monitoring system: a screening for enzymatic and electrochemical interferents.

Authors:  Fausto Lucarelli; Francesco Ricci; Felice Caprio; Francesco Valgimigli; Cosimo Scuffi; Danila Moscone; Giuseppe Palleschi
Journal:  J Diabetes Sci Technol       Date:  2012-09-01

10.  Performance evaluation of a continuous glucose monitoring system under conditions similar to daily life.

Authors:  Stefan Pleus; Christina Schmid; Manuela Link; Eva Zschornack; Hans-Martin Klötzer; Cornelia Haug; Guido Freckmann
Journal:  J Diabetes Sci Technol       Date:  2013-07-01
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.